Efficient (23)Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum (23)Na (CRISTINA)

PURPOSE: To capture the multiquantum coherence (MQC) (23)Na signal. Different phase‐cycling options and sequences are compared in a unified theoretical layout, and a novel sequence is developed. METHODS: An open source simulation overview is provided with graphical explanations to facilitate MQC und...

Descripción completa

Detalles Bibliográficos
Autores principales: Hoesl, Michaela A. U., Schad, Lothar R., Rapacchi, Stanislas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7402015/
https://www.ncbi.nlm.nih.gov/pubmed/32463978
http://dx.doi.org/10.1002/mrm.28284
_version_ 1783566675707166720
author Hoesl, Michaela A. U.
Schad, Lothar R.
Rapacchi, Stanislas
author_facet Hoesl, Michaela A. U.
Schad, Lothar R.
Rapacchi, Stanislas
author_sort Hoesl, Michaela A. U.
collection PubMed
description PURPOSE: To capture the multiquantum coherence (MQC) (23)Na signal. Different phase‐cycling options and sequences are compared in a unified theoretical layout, and a novel sequence is developed. METHODS: An open source simulation overview is provided with graphical explanations to facilitate MQC understanding and access to techniques. Biases such as B(0 )inhomogeneity and stimulated echo signal were simulated for 4 different phase‐cycling options previously described. Considerations for efficiency and accuracy lead to the implementation of a 2D Cartesian single and triple quantum imaging of sodium (CRISTINA) sequence employing two 6‐step cycles in combination with a multi‐echo readout. CRISTINA was compared to simultaneous single‐quantum and triple‐quantum‐filtered MRI of sodium (SISTINA) under strong static magnetic gradient. CRISTINA capabilities were assessed on 8 × 60 mL, 0% to 5% agarose phantom with 50 to 154 mM (23)Na concentration at 7 T. CRISTINA was demonstrated subsequently in vivo in the brain. RESULTS: Simulation of B(0) inhomogeneity showed severe signal dropout, which can lead to erroneous MQC measurement. Stimulated echo signal was highest at the time of triple‐quantum coherences signal maximum. However, stimulated echo signal is separated by Fourier Transform as an offset and did not interfere with MQC signals. The multi‐echo readout enabled capturing both single‐quantum coherences and triple‐quantum coherences signal evolution at once. Signal combination of 2 phase‐cycles with a corresponding B(0) map was found to recover the signal optimally. Experimental results confirm and complement the simulations. CONCLUSION: Considerations for efficient MQC measurements, most importantly avoiding B(0) signal loss, led to the design of CRISTINA. CRISTINA captures triple‐quantum coherences and single‐quantum coherences signal evolution to provide complete sodium signal characterization including [Formula: see text] fast, [Formula: see text] slow, MQC amplitudes, and sodium concentration.
format Online
Article
Text
id pubmed-7402015
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-74020152020-08-06 Efficient (23)Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum (23)Na (CRISTINA) Hoesl, Michaela A. U. Schad, Lothar R. Rapacchi, Stanislas Magn Reson Med Full Papers—Imaging Methodology PURPOSE: To capture the multiquantum coherence (MQC) (23)Na signal. Different phase‐cycling options and sequences are compared in a unified theoretical layout, and a novel sequence is developed. METHODS: An open source simulation overview is provided with graphical explanations to facilitate MQC understanding and access to techniques. Biases such as B(0 )inhomogeneity and stimulated echo signal were simulated for 4 different phase‐cycling options previously described. Considerations for efficiency and accuracy lead to the implementation of a 2D Cartesian single and triple quantum imaging of sodium (CRISTINA) sequence employing two 6‐step cycles in combination with a multi‐echo readout. CRISTINA was compared to simultaneous single‐quantum and triple‐quantum‐filtered MRI of sodium (SISTINA) under strong static magnetic gradient. CRISTINA capabilities were assessed on 8 × 60 mL, 0% to 5% agarose phantom with 50 to 154 mM (23)Na concentration at 7 T. CRISTINA was demonstrated subsequently in vivo in the brain. RESULTS: Simulation of B(0) inhomogeneity showed severe signal dropout, which can lead to erroneous MQC measurement. Stimulated echo signal was highest at the time of triple‐quantum coherences signal maximum. However, stimulated echo signal is separated by Fourier Transform as an offset and did not interfere with MQC signals. The multi‐echo readout enabled capturing both single‐quantum coherences and triple‐quantum coherences signal evolution at once. Signal combination of 2 phase‐cycles with a corresponding B(0) map was found to recover the signal optimally. Experimental results confirm and complement the simulations. CONCLUSION: Considerations for efficient MQC measurements, most importantly avoiding B(0) signal loss, led to the design of CRISTINA. CRISTINA captures triple‐quantum coherences and single‐quantum coherences signal evolution to provide complete sodium signal characterization including [Formula: see text] fast, [Formula: see text] slow, MQC amplitudes, and sodium concentration. John Wiley and Sons Inc. 2020-05-28 2020-11 /pmc/articles/PMC7402015/ /pubmed/32463978 http://dx.doi.org/10.1002/mrm.28284 Text en © 2020 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society forMagnetic Resonance in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.
spellingShingle Full Papers—Imaging Methodology
Hoesl, Michaela A. U.
Schad, Lothar R.
Rapacchi, Stanislas
Efficient (23)Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum (23)Na (CRISTINA)
title Efficient (23)Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum (23)Na (CRISTINA)
title_full Efficient (23)Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum (23)Na (CRISTINA)
title_fullStr Efficient (23)Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum (23)Na (CRISTINA)
title_full_unstemmed Efficient (23)Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum (23)Na (CRISTINA)
title_short Efficient (23)Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum (23)Na (CRISTINA)
title_sort efficient (23)na triple‐quantum signal imaging on clinical scanners: cartesian imaging of single and triple‐quantum (23)na (cristina)
topic Full Papers—Imaging Methodology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7402015/
https://www.ncbi.nlm.nih.gov/pubmed/32463978
http://dx.doi.org/10.1002/mrm.28284
work_keys_str_mv AT hoeslmichaelaau efficient23natriplequantumsignalimagingonclinicalscannerscartesianimagingofsingleandtriplequantum23nacristina
AT schadlotharr efficient23natriplequantumsignalimagingonclinicalscannerscartesianimagingofsingleandtriplequantum23nacristina
AT rapacchistanislas efficient23natriplequantumsignalimagingonclinicalscannerscartesianimagingofsingleandtriplequantum23nacristina