Cargando…

Prediction of Chromatographic Elution Order of Analytical Mixtures Based on Quantitative Structure-Retention Relationships and Multi-Objective Optimization

Prediction of the retention time from the molecular structure using quantitative structure-retention relationships is a powerful tool for the development of methods in reversed-phase HPLC. However, its fundamental limitation lies in the fact that low error in the prediction of the retention time doe...

Descripción completa

Detalles Bibliográficos
Autores principales: Žuvela, Petar, Liu, J. Jay, Wong, Ming Wah, Bączek, Tomasz
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7411958/
https://www.ncbi.nlm.nih.gov/pubmed/32640765
http://dx.doi.org/10.3390/molecules25133085
Descripción
Sumario:Prediction of the retention time from the molecular structure using quantitative structure-retention relationships is a powerful tool for the development of methods in reversed-phase HPLC. However, its fundamental limitation lies in the fact that low error in the prediction of the retention time does not necessarily guarantee a prediction of the elution order. Here, we propose a new method for the prediction of the elution order from quantitative structure-retention relationships using multi-objective optimization. Two case studies were evaluated: (i) separation of organic molecules in a Supelcosil LC-18 column, and (ii) separation of peptides in seven columns under varying conditions. Results have shown that, when compared to predictions based on the conventional model, the relative root mean square error of the elution order decreases by 48.84%, while the relative root mean square error of the retention time increases by 4.22% on average across both case studies. The predictive ability in terms of both retention time and elution order and the corresponding applicability domains were defined. The models were deemed stable and robust with few to no structural outliers.