Cargando…

The Comaneci Device for Entering an Acute-Angled Small Feeder during Cerebral Arteriovenous Malformation Embolization: A Novel Technique to Support Microcatheterization

In this report, we describe the use of the Comaneci device to support microcatheterization in a small branch arising from a parent artery during embolization. In 2 cases, arteriovenous malformations presented with intracranial hemorrhage. A microcatheter was navigated into a small feeder while the C...

Descripción completa

Detalles Bibliográficos
Autores principales: Wessels, Lars, de Vries, Joost, Boogaarts, Hieronymus D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society of Interventional Neuroradiology 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7412652/
https://www.ncbi.nlm.nih.gov/pubmed/32570303
http://dx.doi.org/10.5469/neuroint.2020.00122
Descripción
Sumario:In this report, we describe the use of the Comaneci device to support microcatheterization in a small branch arising from a parent artery during embolization. In 2 cases, arteriovenous malformations presented with intracranial hemorrhage. A microcatheter was navigated into a small feeder while the Comaneci device was deployed just distal to the feeder with an acute angle from the parent artery. Our technical note represents an alternative option of catheterization in cases with difficult access to small feeders originating from higher flow arteries at a sharp angle while maintaining continuous flow in the parent artery.