Repeatability and Reproducibility of Total Retinal Blood Flow Measurements Using Bi-Directional Doppler OCT

PURPOSE: To investigate the repeatability and reproducibility of total retinal blood flow measurements using a custom-built dual-beam bidirectional Doppler optical coherence tomography (OCT) system in healthy subjects. METHODS: Repeatability and reproducibility were analyzed in 10 and 34 healthy sub...

Descripción completa

Detalles Bibliográficos
Autores principales: Szegedi, Stephan, Hommer, Nikolaus, Kallab, Martin, Puchner, Stefan, Schmidl, Doreen, Werkmeister, René M., Garhöfer, Gerhard, Schmetterer, Leopold
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7414639/
https://www.ncbi.nlm.nih.gov/pubmed/32832239
http://dx.doi.org/10.1167/tvst.9.7.34
_version_ 1783569009092853760
author Szegedi, Stephan
Hommer, Nikolaus
Kallab, Martin
Puchner, Stefan
Schmidl, Doreen
Werkmeister, René M.
Garhöfer, Gerhard
Schmetterer, Leopold
author_facet Szegedi, Stephan
Hommer, Nikolaus
Kallab, Martin
Puchner, Stefan
Schmidl, Doreen
Werkmeister, René M.
Garhöfer, Gerhard
Schmetterer, Leopold
author_sort Szegedi, Stephan
collection PubMed
description PURPOSE: To investigate the repeatability and reproducibility of total retinal blood flow measurements using a custom-built dual-beam bidirectional Doppler optical coherence tomography (OCT) system in healthy subjects. METHODS: Repeatability and reproducibility were analyzed in 10 and 34 healthy subjects, respectively. For repeatability, measurements were taken twice within 30 minutes, for reproducibility, twice within two to five weeks. Two analysis approaches were compared for calculation of absolute blood velocities: a previously published approach resulting in values for total arterial (Q(A,abs)) and total venous blood flow (Q(V,abs)) and a novel approach taking into account that there is a fixed relation between the phase shift in the two OCT channels (Q(A,new,) Q(V,new)). Repeatability and reproducibility were quantified using intraclass correlation coefficients (ICCs). RESULTS: For Q(A,abs) and Q(V,abs), ICC values between 0.78 and 0.84 were obtained. Q(A,new) and Q(V,new) values revealed better repeatability and reproducibility as compared to the convential appoach. Repeatability ICCs for Q(A,new) and Q(V,new) were between 0.91 and 0.93, and reproducibility ICCs were between 0.87 and 0.91 indicating excellent reproducibility. Good agreement was observed between total retinal blood flow values as measured from retinal arteries and retinal veins. CONCLUSIONS: Measurement of total retinal blood flow using dual-beam Doppler OCT shows excellent reproducibility, which can further be improved by using a novel algorithm for calculating blood velocities in retinal vessels. TRANSLATIONAL RELEVANCE: Our data indicate that dual-beam Doppler OCT can be used for longitudinal studies. Hence, quantitative retinal blood flow may be established as a biomarker for progression vascular eye diseases.
format Online
Article
Text
id pubmed-7414639
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher The Association for Research in Vision and Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-74146392020-08-21 Repeatability and Reproducibility of Total Retinal Blood Flow Measurements Using Bi-Directional Doppler OCT Szegedi, Stephan Hommer, Nikolaus Kallab, Martin Puchner, Stefan Schmidl, Doreen Werkmeister, René M. Garhöfer, Gerhard Schmetterer, Leopold Transl Vis Sci Technol Article PURPOSE: To investigate the repeatability and reproducibility of total retinal blood flow measurements using a custom-built dual-beam bidirectional Doppler optical coherence tomography (OCT) system in healthy subjects. METHODS: Repeatability and reproducibility were analyzed in 10 and 34 healthy subjects, respectively. For repeatability, measurements were taken twice within 30 minutes, for reproducibility, twice within two to five weeks. Two analysis approaches were compared for calculation of absolute blood velocities: a previously published approach resulting in values for total arterial (Q(A,abs)) and total venous blood flow (Q(V,abs)) and a novel approach taking into account that there is a fixed relation between the phase shift in the two OCT channels (Q(A,new,) Q(V,new)). Repeatability and reproducibility were quantified using intraclass correlation coefficients (ICCs). RESULTS: For Q(A,abs) and Q(V,abs), ICC values between 0.78 and 0.84 were obtained. Q(A,new) and Q(V,new) values revealed better repeatability and reproducibility as compared to the convential appoach. Repeatability ICCs for Q(A,new) and Q(V,new) were between 0.91 and 0.93, and reproducibility ICCs were between 0.87 and 0.91 indicating excellent reproducibility. Good agreement was observed between total retinal blood flow values as measured from retinal arteries and retinal veins. CONCLUSIONS: Measurement of total retinal blood flow using dual-beam Doppler OCT shows excellent reproducibility, which can further be improved by using a novel algorithm for calculating blood velocities in retinal vessels. TRANSLATIONAL RELEVANCE: Our data indicate that dual-beam Doppler OCT can be used for longitudinal studies. Hence, quantitative retinal blood flow may be established as a biomarker for progression vascular eye diseases. The Association for Research in Vision and Ophthalmology 2020-06-25 /pmc/articles/PMC7414639/ /pubmed/32832239 http://dx.doi.org/10.1167/tvst.9.7.34 Text en Copyright 2020 The Authors http://creativecommons.org/licenses/by/4.0/ This work is licensed under a Creative Commons Attribution 4.0 International License.
spellingShingle Article
Szegedi, Stephan
Hommer, Nikolaus
Kallab, Martin
Puchner, Stefan
Schmidl, Doreen
Werkmeister, René M.
Garhöfer, Gerhard
Schmetterer, Leopold
Repeatability and Reproducibility of Total Retinal Blood Flow Measurements Using Bi-Directional Doppler OCT
title Repeatability and Reproducibility of Total Retinal Blood Flow Measurements Using Bi-Directional Doppler OCT
title_full Repeatability and Reproducibility of Total Retinal Blood Flow Measurements Using Bi-Directional Doppler OCT
title_fullStr Repeatability and Reproducibility of Total Retinal Blood Flow Measurements Using Bi-Directional Doppler OCT
title_full_unstemmed Repeatability and Reproducibility of Total Retinal Blood Flow Measurements Using Bi-Directional Doppler OCT
title_short Repeatability and Reproducibility of Total Retinal Blood Flow Measurements Using Bi-Directional Doppler OCT
title_sort repeatability and reproducibility of total retinal blood flow measurements using bi-directional doppler oct
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7414639/
https://www.ncbi.nlm.nih.gov/pubmed/32832239
http://dx.doi.org/10.1167/tvst.9.7.34
work_keys_str_mv AT szegedistephan repeatabilityandreproducibilityoftotalretinalbloodflowmeasurementsusingbidirectionaldoppleroct
AT hommernikolaus repeatabilityandreproducibilityoftotalretinalbloodflowmeasurementsusingbidirectionaldoppleroct
AT kallabmartin repeatabilityandreproducibilityoftotalretinalbloodflowmeasurementsusingbidirectionaldoppleroct
AT puchnerstefan repeatabilityandreproducibilityoftotalretinalbloodflowmeasurementsusingbidirectionaldoppleroct
AT schmidldoreen repeatabilityandreproducibilityoftotalretinalbloodflowmeasurementsusingbidirectionaldoppleroct
AT werkmeisterrenem repeatabilityandreproducibilityoftotalretinalbloodflowmeasurementsusingbidirectionaldoppleroct
AT garhofergerhard repeatabilityandreproducibilityoftotalretinalbloodflowmeasurementsusingbidirectionaldoppleroct
AT schmettererleopold repeatabilityandreproducibilityoftotalretinalbloodflowmeasurementsusingbidirectionaldoppleroct