Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans
PURPOSE: The present study aims to design and fabricate a novel, versatile, and cost-effective Polymethyl Methacrylate (PMMA) head phantom for the dosimetric pretreatment verification of radiotherapy (RT) treatment plans. MATERIALS AND METHODS: The head phantom designing involves slice-wise modeling...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7416869/ https://www.ncbi.nlm.nih.gov/pubmed/32831488 http://dx.doi.org/10.4103/jmp.JMP_21_20 |
_version_ | 1783569377266761728 |
---|---|
author | Shaiju, V. S. Kumar, Rajesh Phani, Debjani Rajasekhar, K. V. Zacharia, George Bhasi, Saju Nair, Raghuram K. |
author_facet | Shaiju, V. S. Kumar, Rajesh Phani, Debjani Rajasekhar, K. V. Zacharia, George Bhasi, Saju Nair, Raghuram K. |
author_sort | Shaiju, V. S. |
collection | PubMed |
description | PURPOSE: The present study aims to design and fabricate a novel, versatile, and cost-effective Polymethyl Methacrylate (PMMA) head phantom for the dosimetric pretreatment verification of radiotherapy (RT) treatment plans. MATERIALS AND METHODS: The head phantom designing involves slice-wise modeling of an adult head using PMMA. The phantom has provisions to hold detectors such as ionization chambers of different sizes, Gafchromic films, gel dosimeter, and optically stimulated luminescence dosimeter. For the point dose verification purpose, 15 volumetric modulated arc therapy patient plans were selected, and doses were measured using a CC13 ionization chamber. The percentage gamma passing rate was calculated for acceptance criteria 3%/3 mm and 2%/2 mm using OmniPro I’mRT film QA software, and Gafchromic EBT3 films were used for 2D planar dose verification. RESULTS: Treatment planning system calculated, and the measured point doses showed a percentage deviation ranged from 0.26 to 1.92. The planar dose fluence measurements, for set acceptance criteria of 3%/3 mm and 2%/2 mm, percentages of points having gamma value <1 were in the range of 99.17 ± 0.25 to 99.88 ± 0.15 and 93.16 ± 0.38 to 98.89 ± 0.23, respectively. Measured dose verification indices were within the acceptable limit. CONCLUSIONS: The dosimetric study reveals that head phantom can be used for routine pretreatment verification for the cranial RT, especially for stereotactic radiosurgery/RT as a part of patient-specific quality assurance. The presently fabricated and validated phantom is novel, versatile, and cost-effective, and many institutes can afford it. |
format | Online Article Text |
id | pubmed-7416869 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-74168692020-08-20 Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans Shaiju, V. S. Kumar, Rajesh Phani, Debjani Rajasekhar, K. V. Zacharia, George Bhasi, Saju Nair, Raghuram K. J Med Phys Original Article PURPOSE: The present study aims to design and fabricate a novel, versatile, and cost-effective Polymethyl Methacrylate (PMMA) head phantom for the dosimetric pretreatment verification of radiotherapy (RT) treatment plans. MATERIALS AND METHODS: The head phantom designing involves slice-wise modeling of an adult head using PMMA. The phantom has provisions to hold detectors such as ionization chambers of different sizes, Gafchromic films, gel dosimeter, and optically stimulated luminescence dosimeter. For the point dose verification purpose, 15 volumetric modulated arc therapy patient plans were selected, and doses were measured using a CC13 ionization chamber. The percentage gamma passing rate was calculated for acceptance criteria 3%/3 mm and 2%/2 mm using OmniPro I’mRT film QA software, and Gafchromic EBT3 films were used for 2D planar dose verification. RESULTS: Treatment planning system calculated, and the measured point doses showed a percentage deviation ranged from 0.26 to 1.92. The planar dose fluence measurements, for set acceptance criteria of 3%/3 mm and 2%/2 mm, percentages of points having gamma value <1 were in the range of 99.17 ± 0.25 to 99.88 ± 0.15 and 93.16 ± 0.38 to 98.89 ± 0.23, respectively. Measured dose verification indices were within the acceptable limit. CONCLUSIONS: The dosimetric study reveals that head phantom can be used for routine pretreatment verification for the cranial RT, especially for stereotactic radiosurgery/RT as a part of patient-specific quality assurance. The presently fabricated and validated phantom is novel, versatile, and cost-effective, and many institutes can afford it. Wolters Kluwer - Medknow 2020 2020-07-20 /pmc/articles/PMC7416869/ /pubmed/32831488 http://dx.doi.org/10.4103/jmp.JMP_21_20 Text en Copyright: © 2020 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Shaiju, V. S. Kumar, Rajesh Phani, Debjani Rajasekhar, K. V. Zacharia, George Bhasi, Saju Nair, Raghuram K. Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans |
title | Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans |
title_full | Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans |
title_fullStr | Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans |
title_full_unstemmed | Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans |
title_short | Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans |
title_sort | design, fabrication, and validation of a polymethyl methacrylate head phantom for dosimetric verification of cranial radiotherapy treatment plans |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7416869/ https://www.ncbi.nlm.nih.gov/pubmed/32831488 http://dx.doi.org/10.4103/jmp.JMP_21_20 |
work_keys_str_mv | AT shaijuvs designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans AT kumarrajesh designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans AT phanidebjani designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans AT rajasekharkv designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans AT zachariageorge designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans AT bhasisaju designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans AT nairraghuramk designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans |