Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans

PURPOSE: The present study aims to design and fabricate a novel, versatile, and cost-effective Polymethyl Methacrylate (PMMA) head phantom for the dosimetric pretreatment verification of radiotherapy (RT) treatment plans. MATERIALS AND METHODS: The head phantom designing involves slice-wise modeling...

Descripción completa

Detalles Bibliográficos
Autores principales: Shaiju, V. S., Kumar, Rajesh, Phani, Debjani, Rajasekhar, K. V., Zacharia, George, Bhasi, Saju, Nair, Raghuram K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7416869/
https://www.ncbi.nlm.nih.gov/pubmed/32831488
http://dx.doi.org/10.4103/jmp.JMP_21_20
_version_ 1783569377266761728
author Shaiju, V. S.
Kumar, Rajesh
Phani, Debjani
Rajasekhar, K. V.
Zacharia, George
Bhasi, Saju
Nair, Raghuram K.
author_facet Shaiju, V. S.
Kumar, Rajesh
Phani, Debjani
Rajasekhar, K. V.
Zacharia, George
Bhasi, Saju
Nair, Raghuram K.
author_sort Shaiju, V. S.
collection PubMed
description PURPOSE: The present study aims to design and fabricate a novel, versatile, and cost-effective Polymethyl Methacrylate (PMMA) head phantom for the dosimetric pretreatment verification of radiotherapy (RT) treatment plans. MATERIALS AND METHODS: The head phantom designing involves slice-wise modeling of an adult head using PMMA. The phantom has provisions to hold detectors such as ionization chambers of different sizes, Gafchromic films, gel dosimeter, and optically stimulated luminescence dosimeter. For the point dose verification purpose, 15 volumetric modulated arc therapy patient plans were selected, and doses were measured using a CC13 ionization chamber. The percentage gamma passing rate was calculated for acceptance criteria 3%/3 mm and 2%/2 mm using OmniPro I’mRT film QA software, and Gafchromic EBT3 films were used for 2D planar dose verification. RESULTS: Treatment planning system calculated, and the measured point doses showed a percentage deviation ranged from 0.26 to 1.92. The planar dose fluence measurements, for set acceptance criteria of 3%/3 mm and 2%/2 mm, percentages of points having gamma value <1 were in the range of 99.17 ± 0.25 to 99.88 ± 0.15 and 93.16 ± 0.38 to 98.89 ± 0.23, respectively. Measured dose verification indices were within the acceptable limit. CONCLUSIONS: The dosimetric study reveals that head phantom can be used for routine pretreatment verification for the cranial RT, especially for stereotactic radiosurgery/RT as a part of patient-specific quality assurance. The presently fabricated and validated phantom is novel, versatile, and cost-effective, and many institutes can afford it.
format Online
Article
Text
id pubmed-7416869
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-74168692020-08-20 Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans Shaiju, V. S. Kumar, Rajesh Phani, Debjani Rajasekhar, K. V. Zacharia, George Bhasi, Saju Nair, Raghuram K. J Med Phys Original Article PURPOSE: The present study aims to design and fabricate a novel, versatile, and cost-effective Polymethyl Methacrylate (PMMA) head phantom for the dosimetric pretreatment verification of radiotherapy (RT) treatment plans. MATERIALS AND METHODS: The head phantom designing involves slice-wise modeling of an adult head using PMMA. The phantom has provisions to hold detectors such as ionization chambers of different sizes, Gafchromic films, gel dosimeter, and optically stimulated luminescence dosimeter. For the point dose verification purpose, 15 volumetric modulated arc therapy patient plans were selected, and doses were measured using a CC13 ionization chamber. The percentage gamma passing rate was calculated for acceptance criteria 3%/3 mm and 2%/2 mm using OmniPro I’mRT film QA software, and Gafchromic EBT3 films were used for 2D planar dose verification. RESULTS: Treatment planning system calculated, and the measured point doses showed a percentage deviation ranged from 0.26 to 1.92. The planar dose fluence measurements, for set acceptance criteria of 3%/3 mm and 2%/2 mm, percentages of points having gamma value <1 were in the range of 99.17 ± 0.25 to 99.88 ± 0.15 and 93.16 ± 0.38 to 98.89 ± 0.23, respectively. Measured dose verification indices were within the acceptable limit. CONCLUSIONS: The dosimetric study reveals that head phantom can be used for routine pretreatment verification for the cranial RT, especially for stereotactic radiosurgery/RT as a part of patient-specific quality assurance. The presently fabricated and validated phantom is novel, versatile, and cost-effective, and many institutes can afford it. Wolters Kluwer - Medknow 2020 2020-07-20 /pmc/articles/PMC7416869/ /pubmed/32831488 http://dx.doi.org/10.4103/jmp.JMP_21_20 Text en Copyright: © 2020 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Shaiju, V. S.
Kumar, Rajesh
Phani, Debjani
Rajasekhar, K. V.
Zacharia, George
Bhasi, Saju
Nair, Raghuram K.
Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans
title Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans
title_full Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans
title_fullStr Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans
title_full_unstemmed Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans
title_short Design, Fabrication, and Validation of a Polymethyl Methacrylate Head Phantom for Dosimetric Verification of Cranial Radiotherapy Treatment Plans
title_sort design, fabrication, and validation of a polymethyl methacrylate head phantom for dosimetric verification of cranial radiotherapy treatment plans
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7416869/
https://www.ncbi.nlm.nih.gov/pubmed/32831488
http://dx.doi.org/10.4103/jmp.JMP_21_20
work_keys_str_mv AT shaijuvs designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans
AT kumarrajesh designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans
AT phanidebjani designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans
AT rajasekharkv designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans
AT zachariageorge designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans
AT bhasisaju designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans
AT nairraghuramk designfabricationandvalidationofapolymethylmethacrylateheadphantomfordosimetricverificationofcranialradiotherapytreatmentplans