A New Approach for Testing Fetal Heart Rate Monitors

In this paper, a new approach for the periodical testing and the functionality evaluation of a fetal heart rate monitor device based on ultrasound principle is proposed. The design and realization of the device are presented, together with the description of its features and functioning tests. In th...

Descripción completa

Detalles Bibliográficos
Autores principales: Bibbo, Daniele, Klinkovsky, Tomas, Penhaker, Marek, Kudrna, Petr, Peter, Lukas, Augustynek, Martin, Kašík, Vladimír, Kubicek, Jan, Selamat, Ali, Cerny, Martin, Bielcik, Daniel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7436177/
https://www.ncbi.nlm.nih.gov/pubmed/32722397
http://dx.doi.org/10.3390/s20154139
_version_ 1783572493283360768
author Bibbo, Daniele
Klinkovsky, Tomas
Penhaker, Marek
Kudrna, Petr
Peter, Lukas
Augustynek, Martin
Kašík, Vladimír
Kubicek, Jan
Selamat, Ali
Cerny, Martin
Bielcik, Daniel
author_facet Bibbo, Daniele
Klinkovsky, Tomas
Penhaker, Marek
Kudrna, Petr
Peter, Lukas
Augustynek, Martin
Kašík, Vladimír
Kubicek, Jan
Selamat, Ali
Cerny, Martin
Bielcik, Daniel
author_sort Bibbo, Daniele
collection PubMed
description In this paper, a new approach for the periodical testing and the functionality evaluation of a fetal heart rate monitor device based on ultrasound principle is proposed. The design and realization of the device are presented, together with the description of its features and functioning tests. In the designed device, a relay element, driven by an electric signal that allows switching at two specific frequencies, is used to simulate the fetus and the mother’s heartbeat. The simulator was designed to be compliant with the standard requirements for accurate assessment and measurement of medical devices. The accuracy of the simulated signals was evaluated, and it resulted to be stable and reliable. The generated frequencies show an error of about 0.5% with respect to the nominal one while the accuracy of the test equipment was within ±3% of the test signal set frequency. This value complies with the technical standard for the accuracy of fetal heart rate monitor devices. Moreover, the performed tests and measurements show the correct functionality of the developed simulator. The proposed equipment and testing respect the technical requirements for medical devices. The features of the proposed device make it simple and quick in testing a fetal heart rate monitor, thus providing an efficient way to evaluate and test the correlation capabilities of commercial apparatuses.
format Online
Article
Text
id pubmed-7436177
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-74361772020-08-24 A New Approach for Testing Fetal Heart Rate Monitors Bibbo, Daniele Klinkovsky, Tomas Penhaker, Marek Kudrna, Petr Peter, Lukas Augustynek, Martin Kašík, Vladimír Kubicek, Jan Selamat, Ali Cerny, Martin Bielcik, Daniel Sensors (Basel) Article In this paper, a new approach for the periodical testing and the functionality evaluation of a fetal heart rate monitor device based on ultrasound principle is proposed. The design and realization of the device are presented, together with the description of its features and functioning tests. In the designed device, a relay element, driven by an electric signal that allows switching at two specific frequencies, is used to simulate the fetus and the mother’s heartbeat. The simulator was designed to be compliant with the standard requirements for accurate assessment and measurement of medical devices. The accuracy of the simulated signals was evaluated, and it resulted to be stable and reliable. The generated frequencies show an error of about 0.5% with respect to the nominal one while the accuracy of the test equipment was within ±3% of the test signal set frequency. This value complies with the technical standard for the accuracy of fetal heart rate monitor devices. Moreover, the performed tests and measurements show the correct functionality of the developed simulator. The proposed equipment and testing respect the technical requirements for medical devices. The features of the proposed device make it simple and quick in testing a fetal heart rate monitor, thus providing an efficient way to evaluate and test the correlation capabilities of commercial apparatuses. MDPI 2020-07-25 /pmc/articles/PMC7436177/ /pubmed/32722397 http://dx.doi.org/10.3390/s20154139 Text en © 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Bibbo, Daniele
Klinkovsky, Tomas
Penhaker, Marek
Kudrna, Petr
Peter, Lukas
Augustynek, Martin
Kašík, Vladimír
Kubicek, Jan
Selamat, Ali
Cerny, Martin
Bielcik, Daniel
A New Approach for Testing Fetal Heart Rate Monitors
title A New Approach for Testing Fetal Heart Rate Monitors
title_full A New Approach for Testing Fetal Heart Rate Monitors
title_fullStr A New Approach for Testing Fetal Heart Rate Monitors
title_full_unstemmed A New Approach for Testing Fetal Heart Rate Monitors
title_short A New Approach for Testing Fetal Heart Rate Monitors
title_sort new approach for testing fetal heart rate monitors
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7436177/
https://www.ncbi.nlm.nih.gov/pubmed/32722397
http://dx.doi.org/10.3390/s20154139
work_keys_str_mv AT bibbodaniele anewapproachfortestingfetalheartratemonitors
AT klinkovskytomas anewapproachfortestingfetalheartratemonitors
AT penhakermarek anewapproachfortestingfetalheartratemonitors
AT kudrnapetr anewapproachfortestingfetalheartratemonitors
AT peterlukas anewapproachfortestingfetalheartratemonitors
AT augustynekmartin anewapproachfortestingfetalheartratemonitors
AT kasikvladimir anewapproachfortestingfetalheartratemonitors
AT kubicekjan anewapproachfortestingfetalheartratemonitors
AT selamatali anewapproachfortestingfetalheartratemonitors
AT cernymartin anewapproachfortestingfetalheartratemonitors
AT bielcikdaniel anewapproachfortestingfetalheartratemonitors
AT bibbodaniele newapproachfortestingfetalheartratemonitors
AT klinkovskytomas newapproachfortestingfetalheartratemonitors
AT penhakermarek newapproachfortestingfetalheartratemonitors
AT kudrnapetr newapproachfortestingfetalheartratemonitors
AT peterlukas newapproachfortestingfetalheartratemonitors
AT augustynekmartin newapproachfortestingfetalheartratemonitors
AT kasikvladimir newapproachfortestingfetalheartratemonitors
AT kubicekjan newapproachfortestingfetalheartratemonitors
AT selamatali newapproachfortestingfetalheartratemonitors
AT cernymartin newapproachfortestingfetalheartratemonitors
AT bielcikdaniel newapproachfortestingfetalheartratemonitors