Does the addition of vitamin E to conventional UHMWPE improve the wear performance of hip acetabular cups? Micro-Raman characterization of differently processed polyethylene acetabular cups worn on a hip joint simulator
In knee replacements, vitamin E-doped ultra-high molecular weight polyethylene (UHMWPE) shows a better wear behavior than standard UHMWPE. Therefore, different sets of polyethylene (PE) acetabular cups, i.e. standard UHMWPE and cross-linked polyethylene irradiated with 50 kGy and 75 kGy, were compar...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Associação Brasileira de Divulgação Científica
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7437989/ https://www.ncbi.nlm.nih.gov/pubmed/32813853 http://dx.doi.org/10.1590/1414-431X20209930 |
_version_ | 1783572724787970048 |
---|---|
author | Di Foggia, M. Affatato, S. Taddei, P. |
author_facet | Di Foggia, M. Affatato, S. Taddei, P. |
author_sort | Di Foggia, M. |
collection | PubMed |
description | In knee replacements, vitamin E-doped ultra-high molecular weight polyethylene (UHMWPE) shows a better wear behavior than standard UHMWPE. Therefore, different sets of polyethylene (PE) acetabular cups, i.e. standard UHMWPE and cross-linked polyethylene irradiated with 50 kGy and 75 kGy, were compared, at a molecular level, with vitamin E-doped UHMWPE to evaluate their wear performance after being tested on a hip joint simulator for five million cycles. Unworn control and worn acetabular cups were analyzed by micro-Raman spectroscopy to gain insight into the effects of wear on the microstructure and phase composition of PE. Macroscopic wear was evaluated through mass loss measurements. The data showed that the samples could be divided into two groups: 1) standard and vitamin E-doped cups (mass loss of about 100 mg) and 2) the cross-linked cups (mass loss of about 30-40 mg). Micro-Raman spectroscopy disclosed different wear mechanisms in the four sets of acetabular cups, which were related to surface topography data. The vitamin E-doped samples did not show a better wear behavior than the cross-linked ones in terms of either mass loss or morphology changes. However, they showed lower variation at the morphological level (lower changes in phase composition) than the UHMWPE cups, thus confirming a certain protecting role of vitamin E against microstructural changes induced by wear testing. |
format | Online Article Text |
id | pubmed-7437989 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Associação Brasileira de Divulgação Científica |
record_format | MEDLINE/PubMed |
spelling | pubmed-74379892020-08-28 Does the addition of vitamin E to conventional UHMWPE improve the wear performance of hip acetabular cups? Micro-Raman characterization of differently processed polyethylene acetabular cups worn on a hip joint simulator Di Foggia, M. Affatato, S. Taddei, P. Braz J Med Biol Res Research Article In knee replacements, vitamin E-doped ultra-high molecular weight polyethylene (UHMWPE) shows a better wear behavior than standard UHMWPE. Therefore, different sets of polyethylene (PE) acetabular cups, i.e. standard UHMWPE and cross-linked polyethylene irradiated with 50 kGy and 75 kGy, were compared, at a molecular level, with vitamin E-doped UHMWPE to evaluate their wear performance after being tested on a hip joint simulator for five million cycles. Unworn control and worn acetabular cups were analyzed by micro-Raman spectroscopy to gain insight into the effects of wear on the microstructure and phase composition of PE. Macroscopic wear was evaluated through mass loss measurements. The data showed that the samples could be divided into two groups: 1) standard and vitamin E-doped cups (mass loss of about 100 mg) and 2) the cross-linked cups (mass loss of about 30-40 mg). Micro-Raman spectroscopy disclosed different wear mechanisms in the four sets of acetabular cups, which were related to surface topography data. The vitamin E-doped samples did not show a better wear behavior than the cross-linked ones in terms of either mass loss or morphology changes. However, they showed lower variation at the morphological level (lower changes in phase composition) than the UHMWPE cups, thus confirming a certain protecting role of vitamin E against microstructural changes induced by wear testing. Associação Brasileira de Divulgação Científica 2020-08-17 /pmc/articles/PMC7437989/ /pubmed/32813853 http://dx.doi.org/10.1590/1414-431X20209930 Text en https://creativecommons.org/licenses/by/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Di Foggia, M. Affatato, S. Taddei, P. Does the addition of vitamin E to conventional UHMWPE improve the wear performance of hip acetabular cups? Micro-Raman characterization of differently processed polyethylene acetabular cups worn on a hip joint simulator |
title | Does the addition of vitamin E to conventional UHMWPE improve the wear performance of hip acetabular cups? Micro-Raman characterization of differently processed polyethylene acetabular cups worn on a hip joint simulator |
title_full | Does the addition of vitamin E to conventional UHMWPE improve the wear performance of hip acetabular cups? Micro-Raman characterization of differently processed polyethylene acetabular cups worn on a hip joint simulator |
title_fullStr | Does the addition of vitamin E to conventional UHMWPE improve the wear performance of hip acetabular cups? Micro-Raman characterization of differently processed polyethylene acetabular cups worn on a hip joint simulator |
title_full_unstemmed | Does the addition of vitamin E to conventional UHMWPE improve the wear performance of hip acetabular cups? Micro-Raman characterization of differently processed polyethylene acetabular cups worn on a hip joint simulator |
title_short | Does the addition of vitamin E to conventional UHMWPE improve the wear performance of hip acetabular cups? Micro-Raman characterization of differently processed polyethylene acetabular cups worn on a hip joint simulator |
title_sort | does the addition of vitamin e to conventional uhmwpe improve the wear performance of hip acetabular cups? micro-raman characterization of differently processed polyethylene acetabular cups worn on a hip joint simulator |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7437989/ https://www.ncbi.nlm.nih.gov/pubmed/32813853 http://dx.doi.org/10.1590/1414-431X20209930 |
work_keys_str_mv | AT difoggiam doestheadditionofvitaminetoconventionaluhmwpeimprovethewearperformanceofhipacetabularcupsmicroramancharacterizationofdifferentlyprocessedpolyethyleneacetabularcupswornonahipjointsimulator AT affatatos doestheadditionofvitaminetoconventionaluhmwpeimprovethewearperformanceofhipacetabularcupsmicroramancharacterizationofdifferentlyprocessedpolyethyleneacetabularcupswornonahipjointsimulator AT taddeip doestheadditionofvitaminetoconventionaluhmwpeimprovethewearperformanceofhipacetabularcupsmicroramancharacterizationofdifferentlyprocessedpolyethyleneacetabularcupswornonahipjointsimulator |