Optimization and validation of patient‐based real‐time quality control procedure using moving average and average of normals with multi‐rules for TT3, TT4, FT3, FT3, and TSH on three analyzers

BACKGROUND: We have designed a patient‐based real‐time quality control (PBRTQC) procedure to detect analytical shifts and review analytical trends of measurement procedures. METHODS: All the nine months' patient results of total thyroxine (TT4), total triiodothyronine (TT3), free thyroxine (FT4...

Descripción completa

Detalles Bibliográficos
Autores principales: Song, Chao, Zhou, Jun, Xia, Jun, Ye, Deli, Chen, Qian, Li, Weixing
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7439426/
https://www.ncbi.nlm.nih.gov/pubmed/32363618
http://dx.doi.org/10.1002/jcla.23314