Performance evaluation of three i‐SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer

BACKGROUND: Capillary blood is the most commonly used sample for point‐of‐care (POC) glucometers. However, in critically ill patients, the glucose levels measured from capillary blood may not be reliable. Thus, we aimed to evaluate and compare the accuracy of glucose levels measured with POC glucome...

Descripción completa

Detalles Bibliográficos
Autores principales: Kim, Ha Nui, Moon, Kyung Chul, Yoon, Soo‐Young
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7439437/
https://www.ncbi.nlm.nih.gov/pubmed/32430994
http://dx.doi.org/10.1002/jcla.23356
_version_ 1783572982004711424
author Kim, Ha Nui
Moon, Kyung Chul
Yoon, Soo‐Young
author_facet Kim, Ha Nui
Moon, Kyung Chul
Yoon, Soo‐Young
author_sort Kim, Ha Nui
collection PubMed
description BACKGROUND: Capillary blood is the most commonly used sample for point‐of‐care (POC) glucometers. However, in critically ill patients, the glucose levels measured from capillary blood may not be reliable. Thus, we aimed to evaluate and compare the accuracy of glucose levels measured with POC glucometers and the YSI 2300 glucose analyzer using leftover arterial blood samples. METHODS: In total, 100 leftover heparinized arterial blood samples were used to evaluate the performance of three i‐SENS glucometers (BAROzen H Expert plus, CareSens PRO, and CareSens H Beat) and the ACCU‐CHEK(®) Inform II glucometer. The reference value was obtained using the YSI 2300 glucose analyzer. The results were analyzed based on International Organization for Standardization 15197:2013 guidelines. RESULTS: More than 95% of results obtained using POC glucometers were within ±15 mg/dL of the reference value for glucose concentrations <100 mg/dL and within ±15% of the reference value for glucose concentrations ≥100 mg/dL. In the consensus error grid analysis, more than 99% of results were found to be within zones A and B. An excellent correlation was found between the values obtained using POC glucometers and the YSI 2300 glucose analyzer (R (2) > .99). CONCLUSION: The i‐SENS glucometers showed stable and accurate results when leftover arterial blood samples were used. Therefore, POC glucometers could be useful in critical care settings, such as intensive care units, where arterial samples are routinely used.
format Online
Article
Text
id pubmed-7439437
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-74394372020-08-21 Performance evaluation of three i‐SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer Kim, Ha Nui Moon, Kyung Chul Yoon, Soo‐Young J Clin Lab Anal Research Articles BACKGROUND: Capillary blood is the most commonly used sample for point‐of‐care (POC) glucometers. However, in critically ill patients, the glucose levels measured from capillary blood may not be reliable. Thus, we aimed to evaluate and compare the accuracy of glucose levels measured with POC glucometers and the YSI 2300 glucose analyzer using leftover arterial blood samples. METHODS: In total, 100 leftover heparinized arterial blood samples were used to evaluate the performance of three i‐SENS glucometers (BAROzen H Expert plus, CareSens PRO, and CareSens H Beat) and the ACCU‐CHEK(®) Inform II glucometer. The reference value was obtained using the YSI 2300 glucose analyzer. The results were analyzed based on International Organization for Standardization 15197:2013 guidelines. RESULTS: More than 95% of results obtained using POC glucometers were within ±15 mg/dL of the reference value for glucose concentrations <100 mg/dL and within ±15% of the reference value for glucose concentrations ≥100 mg/dL. In the consensus error grid analysis, more than 99% of results were found to be within zones A and B. An excellent correlation was found between the values obtained using POC glucometers and the YSI 2300 glucose analyzer (R (2) > .99). CONCLUSION: The i‐SENS glucometers showed stable and accurate results when leftover arterial blood samples were used. Therefore, POC glucometers could be useful in critical care settings, such as intensive care units, where arterial samples are routinely used. John Wiley and Sons Inc. 2020-05-19 /pmc/articles/PMC7439437/ /pubmed/32430994 http://dx.doi.org/10.1002/jcla.23356 Text en © 2020 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Articles
Kim, Ha Nui
Moon, Kyung Chul
Yoon, Soo‐Young
Performance evaluation of three i‐SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer
title Performance evaluation of three i‐SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer
title_full Performance evaluation of three i‐SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer
title_fullStr Performance evaluation of three i‐SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer
title_full_unstemmed Performance evaluation of three i‐SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer
title_short Performance evaluation of three i‐SENS glucometers using arterial blood samples compared with the YSI 2300 Glucose Analyzer
title_sort performance evaluation of three i‐sens glucometers using arterial blood samples compared with the ysi 2300 glucose analyzer
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7439437/
https://www.ncbi.nlm.nih.gov/pubmed/32430994
http://dx.doi.org/10.1002/jcla.23356
work_keys_str_mv AT kimhanui performanceevaluationofthreeisensglucometersusingarterialbloodsamplescomparedwiththeysi2300glucoseanalyzer
AT moonkyungchul performanceevaluationofthreeisensglucometersusingarterialbloodsamplescomparedwiththeysi2300glucoseanalyzer
AT yoonsooyoung performanceevaluationofthreeisensglucometersusingarterialbloodsamplescomparedwiththeysi2300glucoseanalyzer