Intradevice Repeatability and Interdevice Agreement of Ocular Biometric Measurements: A Comparison of Two Swept-Source Anterior Segment OCT Devices

PURPOSE: To assess the repeatability and agreement of ocular biometric parameters measured using the Tomey CASIA SS-1000 and Heidelberg ANTERION anterior segment optical coherence tomography (AS-OCT) devices. METHODS: Both eyes of subjects 18 years of age or older were scanned three times with the C...

Descripción completa

Detalles Bibliográficos
Autores principales: Pardeshi, Anmol A., Song, Abe E., Lazkani, Naim, Xie, Xiaobin, Huang, Alex, Xu, Benjamin Y.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7442878/
https://www.ncbi.nlm.nih.gov/pubmed/32879770
http://dx.doi.org/10.1167/tvst.9.9.14
_version_ 1783573522225823744
author Pardeshi, Anmol A.
Song, Abe E.
Lazkani, Naim
Xie, Xiaobin
Huang, Alex
Xu, Benjamin Y.
author_facet Pardeshi, Anmol A.
Song, Abe E.
Lazkani, Naim
Xie, Xiaobin
Huang, Alex
Xu, Benjamin Y.
author_sort Pardeshi, Anmol A.
collection PubMed
description PURPOSE: To assess the repeatability and agreement of ocular biometric parameters measured using the Tomey CASIA SS-1000 and Heidelberg ANTERION anterior segment optical coherence tomography (AS-OCT) devices. METHODS: Both eyes of subjects 18 years of age or older were scanned three times with the CASIA and ANTERION under standardized dark lighting. One AS-OCT image along the horizontal (temporal-nasal) meridian was analyzed per eye and per scan. Pupillary diameter (PD) was within 15% for all pairwise comparisons. Anterior chamber depth, lens vault, anterior chamber width, angle opening distance, trabecular iris space area, and scleral spur angle (SSA500) were measured using manufacturer-provided image analysis software. Intraclass correlation (ICC), Wilcoxon signed-rank, and Bland-Altman analyses were performed to assess intradevice repeatability and interdevice agreement of measurements. RESULTS: Thirty-two eyes of 21 subjects were analyzed. There was excellent agreement (ICC >0.98) and no significant difference (P > 0.05) in PD across all comparisons. Intradevice measurement repeatability was excellent for both the CASIA (ICC range 0.93–0.99) and ANTERION (ICC range 0.97–0.99). Interdevice measurement agreement was also excellent (ICC range 0.85–0.96). Measurements within and between devices were similar (P > 0.06) for all parameters except SSA500 (P = 0.03). Linear regression and Bland-Altman plots showed the relationship was consistent across the entire range of measurements. CONCLUSIONS: Intradevice measurement repeatability is excellent for the CASIA and ANTERION. Interdevice measurement agreement between the two devices exceeds metrics reported by previous comparison studies. TRANSLATIONAL RELEVANCE: Modern swept-source AS-OCT devices produce highly repeatable measurements of ocular biometric parameters that are nearly interchangeable across devices.
format Online
Article
Text
id pubmed-7442878
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher The Association for Research in Vision and Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-74428782020-09-01 Intradevice Repeatability and Interdevice Agreement of Ocular Biometric Measurements: A Comparison of Two Swept-Source Anterior Segment OCT Devices Pardeshi, Anmol A. Song, Abe E. Lazkani, Naim Xie, Xiaobin Huang, Alex Xu, Benjamin Y. Transl Vis Sci Technol Article PURPOSE: To assess the repeatability and agreement of ocular biometric parameters measured using the Tomey CASIA SS-1000 and Heidelberg ANTERION anterior segment optical coherence tomography (AS-OCT) devices. METHODS: Both eyes of subjects 18 years of age or older were scanned three times with the CASIA and ANTERION under standardized dark lighting. One AS-OCT image along the horizontal (temporal-nasal) meridian was analyzed per eye and per scan. Pupillary diameter (PD) was within 15% for all pairwise comparisons. Anterior chamber depth, lens vault, anterior chamber width, angle opening distance, trabecular iris space area, and scleral spur angle (SSA500) were measured using manufacturer-provided image analysis software. Intraclass correlation (ICC), Wilcoxon signed-rank, and Bland-Altman analyses were performed to assess intradevice repeatability and interdevice agreement of measurements. RESULTS: Thirty-two eyes of 21 subjects were analyzed. There was excellent agreement (ICC >0.98) and no significant difference (P > 0.05) in PD across all comparisons. Intradevice measurement repeatability was excellent for both the CASIA (ICC range 0.93–0.99) and ANTERION (ICC range 0.97–0.99). Interdevice measurement agreement was also excellent (ICC range 0.85–0.96). Measurements within and between devices were similar (P > 0.06) for all parameters except SSA500 (P = 0.03). Linear regression and Bland-Altman plots showed the relationship was consistent across the entire range of measurements. CONCLUSIONS: Intradevice measurement repeatability is excellent for the CASIA and ANTERION. Interdevice measurement agreement between the two devices exceeds metrics reported by previous comparison studies. TRANSLATIONAL RELEVANCE: Modern swept-source AS-OCT devices produce highly repeatable measurements of ocular biometric parameters that are nearly interchangeable across devices. The Association for Research in Vision and Ophthalmology 2020-08-07 /pmc/articles/PMC7442878/ /pubmed/32879770 http://dx.doi.org/10.1167/tvst.9.9.14 Text en Copyright 2020 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
spellingShingle Article
Pardeshi, Anmol A.
Song, Abe E.
Lazkani, Naim
Xie, Xiaobin
Huang, Alex
Xu, Benjamin Y.
Intradevice Repeatability and Interdevice Agreement of Ocular Biometric Measurements: A Comparison of Two Swept-Source Anterior Segment OCT Devices
title Intradevice Repeatability and Interdevice Agreement of Ocular Biometric Measurements: A Comparison of Two Swept-Source Anterior Segment OCT Devices
title_full Intradevice Repeatability and Interdevice Agreement of Ocular Biometric Measurements: A Comparison of Two Swept-Source Anterior Segment OCT Devices
title_fullStr Intradevice Repeatability and Interdevice Agreement of Ocular Biometric Measurements: A Comparison of Two Swept-Source Anterior Segment OCT Devices
title_full_unstemmed Intradevice Repeatability and Interdevice Agreement of Ocular Biometric Measurements: A Comparison of Two Swept-Source Anterior Segment OCT Devices
title_short Intradevice Repeatability and Interdevice Agreement of Ocular Biometric Measurements: A Comparison of Two Swept-Source Anterior Segment OCT Devices
title_sort intradevice repeatability and interdevice agreement of ocular biometric measurements: a comparison of two swept-source anterior segment oct devices
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7442878/
https://www.ncbi.nlm.nih.gov/pubmed/32879770
http://dx.doi.org/10.1167/tvst.9.9.14
work_keys_str_mv AT pardeshianmola intradevicerepeatabilityandinterdeviceagreementofocularbiometricmeasurementsacomparisonoftwosweptsourceanteriorsegmentoctdevices
AT songabee intradevicerepeatabilityandinterdeviceagreementofocularbiometricmeasurementsacomparisonoftwosweptsourceanteriorsegmentoctdevices
AT lazkaninaim intradevicerepeatabilityandinterdeviceagreementofocularbiometricmeasurementsacomparisonoftwosweptsourceanteriorsegmentoctdevices
AT xiexiaobin intradevicerepeatabilityandinterdeviceagreementofocularbiometricmeasurementsacomparisonoftwosweptsourceanteriorsegmentoctdevices
AT huangalex intradevicerepeatabilityandinterdeviceagreementofocularbiometricmeasurementsacomparisonoftwosweptsourceanteriorsegmentoctdevices
AT xubenjaminy intradevicerepeatabilityandinterdeviceagreementofocularbiometricmeasurementsacomparisonoftwosweptsourceanteriorsegmentoctdevices