Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube

The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TA...

Descripción completa

Detalles Bibliográficos
Autores principales: Shin, Sunghwan, Oh, Jongwon, Park, Hyung-Doo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society for Laboratory Medicine 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7443522/
https://www.ncbi.nlm.nih.gov/pubmed/32829587
http://dx.doi.org/10.3343/alm.2021.41.1.114
_version_ 1783573645373734912
author Shin, Sunghwan
Oh, Jongwon
Park, Hyung-Doo
author_facet Shin, Sunghwan
Oh, Jongwon
Park, Hyung-Doo
author_sort Shin, Sunghwan
collection PubMed
description The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TAT). Blood samples were collected from 30 healthy volunteers in a Barricor tube, serum separating tube (SST, Vacutainer SST II Tube 8.5 mL, #368972; BD), or plasma separating tube (PST, Vacutainer PST Tube 8.0 mL, #367964; BD) in random order. Next, 27 chemistry analytes, six immunochemistry analytes, and two cardiac markers were compared using Passing-Bablok regression and the Bland-Altman method. Preanalytical TAT was measured for each tube. The Barricor tube exhibited bias exceeding the desirable limit for nine and four analytes compared with the SST and PST, respectively. The Barricor tube lactate dehydrogenase value showed a bias of −10.29% and −9.86% compared with that of the SST and PST, respectively. The preanalytical TAT of Barricor tube was 8.8 minutes, which was the shortest among the three tubes. The clinical performance of the Barricor tube was equivalent to that of the SST and PST for most analytes, with an apparent advantage in preanalytical TAT. When using the Barricor tube, the reference range needs to be changed for some analytes that exceed the desirable bias limit.
format Online
Article
Text
id pubmed-7443522
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Korean Society for Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-74435222021-01-01 Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube Shin, Sunghwan Oh, Jongwon Park, Hyung-Doo Ann Lab Med Brief Communication The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TAT). Blood samples were collected from 30 healthy volunteers in a Barricor tube, serum separating tube (SST, Vacutainer SST II Tube 8.5 mL, #368972; BD), or plasma separating tube (PST, Vacutainer PST Tube 8.0 mL, #367964; BD) in random order. Next, 27 chemistry analytes, six immunochemistry analytes, and two cardiac markers were compared using Passing-Bablok regression and the Bland-Altman method. Preanalytical TAT was measured for each tube. The Barricor tube exhibited bias exceeding the desirable limit for nine and four analytes compared with the SST and PST, respectively. The Barricor tube lactate dehydrogenase value showed a bias of −10.29% and −9.86% compared with that of the SST and PST, respectively. The preanalytical TAT of Barricor tube was 8.8 minutes, which was the shortest among the three tubes. The clinical performance of the Barricor tube was equivalent to that of the SST and PST for most analytes, with an apparent advantage in preanalytical TAT. When using the Barricor tube, the reference range needs to be changed for some analytes that exceed the desirable bias limit. Korean Society for Laboratory Medicine 2021-01 2021-01-01 /pmc/articles/PMC7443522/ /pubmed/32829587 http://dx.doi.org/10.3343/alm.2021.41.1.114 Text en Copyright © Korean Society for Laboratory Medicine This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Brief Communication
Shin, Sunghwan
Oh, Jongwon
Park, Hyung-Doo
Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
title Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
title_full Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
title_fullStr Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
title_full_unstemmed Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
title_short Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
title_sort comparison of three blood collection tubes for 35 biochemical analytes: the becton dickinson barricor tube, serum separating tube, and plasma separating tube
topic Brief Communication
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7443522/
https://www.ncbi.nlm.nih.gov/pubmed/32829587
http://dx.doi.org/10.3343/alm.2021.41.1.114
work_keys_str_mv AT shinsunghwan comparisonofthreebloodcollectiontubesfor35biochemicalanalytesthebectondickinsonbarricortubeserumseparatingtubeandplasmaseparatingtube
AT ohjongwon comparisonofthreebloodcollectiontubesfor35biochemicalanalytesthebectondickinsonbarricortubeserumseparatingtubeandplasmaseparatingtube
AT parkhyungdoo comparisonofthreebloodcollectiontubesfor35biochemicalanalytesthebectondickinsonbarricortubeserumseparatingtubeandplasmaseparatingtube