Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TA...
Autores principales: | Shin, Sunghwan, Oh, Jongwon, Park, Hyung-Doo |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Society for Laboratory Medicine
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7443522/ https://www.ncbi.nlm.nih.gov/pubmed/32829587 http://dx.doi.org/10.3343/alm.2021.41.1.114 |
Ejemplares similares
-
A comparison of stability of chemical analytes in plasma from the BD Vacutainer(®) Barricor™ tube with mechanical separator versus tubes containing gel separator
por: Gawria, Ghassaan, et al.
Publicado: (2019) -
Influence of centrifugation conditions on the results of 77 routine clinical chemistry analytes using standard vacuum blood collection tubes and the new BD-Barricor tubes
por: Cadamuro, Janne, et al.
Publicado: (2017) -
Comparison of Barricor™ vs. lithium heparin tubes for selected routine biochemical analytes and evaluation of post centrifugation stability
por: Dupuy, Anne Marie, et al.
Publicado: (2018) -
The local clinical validation of a new lithium heparin tube with a barrier: BD Vacutainer® Barricor LH Plasma tube
por: Arslan, Fatma Demet, et al.
Publicado: (2017) -
Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube?
por: Mandić, Sanja, et al.
Publicado: (2020)