D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections

The D-dimer test is a component of the modified scoring criteria for periprosthetic joint infection (PJI). The performance of the D-dimer test varies greatly among laboratories because of the lack of standardization. Laboratories may use different assays and will produce widely varying results for t...

Descripción completa

Detalles Bibliográficos
Autores principales: Pearson, Lauren N., Moser, Karen A., Schmidt, Robert L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7451938/
https://www.ncbi.nlm.nih.gov/pubmed/32875020
http://dx.doi.org/10.1016/j.artd.2020.07.014
_version_ 1783575077401395200
author Pearson, Lauren N.
Moser, Karen A.
Schmidt, Robert L.
author_facet Pearson, Lauren N.
Moser, Karen A.
Schmidt, Robert L.
author_sort Pearson, Lauren N.
collection PubMed
description The D-dimer test is a component of the modified scoring criteria for periprosthetic joint infection (PJI). The performance of the D-dimer test varies greatly among laboratories because of the lack of standardization. Laboratories may use different assays and will produce widely varying results for the same sample. This study used published proficiency testing data from 3903 laboratories to demonstrate the variability in D-dimer results and estimate the misclassification rate of patients using the proposed cutoff for the test as a component of PJI criteria. Given the variability in D-dimer results, a clinically significant percentage of patients are likely to be misclassified. The data illustrate that a universal cutoff for this marker in the context of assessment for PJI is not appropriate. Each site must conduct a study to determine an appropriate cutoff for their unique testing platform.
format Online
Article
Text
id pubmed-7451938
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-74519382020-08-31 D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections Pearson, Lauren N. Moser, Karen A. Schmidt, Robert L. Arthroplast Today Brief Communication The D-dimer test is a component of the modified scoring criteria for periprosthetic joint infection (PJI). The performance of the D-dimer test varies greatly among laboratories because of the lack of standardization. Laboratories may use different assays and will produce widely varying results for the same sample. This study used published proficiency testing data from 3903 laboratories to demonstrate the variability in D-dimer results and estimate the misclassification rate of patients using the proposed cutoff for the test as a component of PJI criteria. Given the variability in D-dimer results, a clinically significant percentage of patients are likely to be misclassified. The data illustrate that a universal cutoff for this marker in the context of assessment for PJI is not appropriate. Each site must conduct a study to determine an appropriate cutoff for their unique testing platform. Elsevier 2020-08-23 /pmc/articles/PMC7451938/ /pubmed/32875020 http://dx.doi.org/10.1016/j.artd.2020.07.014 Text en © 2020 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Brief Communication
Pearson, Lauren N.
Moser, Karen A.
Schmidt, Robert L.
D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections
title D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections
title_full D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections
title_fullStr D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections
title_full_unstemmed D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections
title_short D-Dimer Varies Widely Across Instrument Platforms and is Not a Reliable Indicator of Periprosthetic Joint Infections
title_sort d-dimer varies widely across instrument platforms and is not a reliable indicator of periprosthetic joint infections
topic Brief Communication
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7451938/
https://www.ncbi.nlm.nih.gov/pubmed/32875020
http://dx.doi.org/10.1016/j.artd.2020.07.014
work_keys_str_mv AT pearsonlaurenn ddimervarieswidelyacrossinstrumentplatformsandisnotareliableindicatorofperiprostheticjointinfections
AT moserkarena ddimervarieswidelyacrossinstrumentplatformsandisnotareliableindicatorofperiprostheticjointinfections
AT schmidtrobertl ddimervarieswidelyacrossinstrumentplatformsandisnotareliableindicatorofperiprostheticjointinfections