Vascular Stents Coated with Electrospun Drug-Eluting Material: Functioning in Rabbit Iliac Artery
A stenting procedure aimed at blood flow restoration in stenosed arteries significantly improves the efficiency of vascular surgery. However, the current challenge is to prevent neointimal growth, which reduces the vessel lumen, in the stented segments in the long run. We tested in vivo drug-eluting...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7465440/ https://www.ncbi.nlm.nih.gov/pubmed/32759856 http://dx.doi.org/10.3390/polym12081741 |
_version_ | 1783577588317290496 |
---|---|
author | Kuznetsov, Konstantin A. Murashov, Ivan S. Chernonosova, Vera S. Chelobanov, Boris P. Stepanova, Alena O. Sergeevichev, David S. Karpenko, Andrey A. Laktionov, Pavel P. |
author_facet | Kuznetsov, Konstantin A. Murashov, Ivan S. Chernonosova, Vera S. Chelobanov, Boris P. Stepanova, Alena O. Sergeevichev, David S. Karpenko, Andrey A. Laktionov, Pavel P. |
author_sort | Kuznetsov, Konstantin A. |
collection | PubMed |
description | A stenting procedure aimed at blood flow restoration in stenosed arteries significantly improves the efficiency of vascular surgery. However, the current challenge is to prevent neointimal growth, which reduces the vessel lumen, in the stented segments in the long run. We tested in vivo drug-eluting coating applied by electrospinning to metal vascular stents to inhibit the overgrowth of neointimal cells via both the drug release and mechanical support of the vascular wall. The blend of polycaprolactone with human serum albumin and paclitaxel was used for stent coating by electrospinning. The drug-eluting stents (DESs) were placed using a balloon catheter to the rabbit common iliac artery for 1, 3, and 6 months. The blood flow rate was ultrasonically determined in vivo. After explantation, the stented arterial segment was visually and histologically examined. Any undesirable biological responses (rejection or hemodynamically significant stenosis) were unobservable in the experimental groups. DESs were less traumatic and induced weaker neointimal growth; over six months, the blood flow increased by 37% versus bare-metal stents, where it increased by at least double the rate. Thus, electrospun-coated DESs demonstrate considerable advantages over the bare-metal variants. |
format | Online Article Text |
id | pubmed-7465440 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-74654402020-09-04 Vascular Stents Coated with Electrospun Drug-Eluting Material: Functioning in Rabbit Iliac Artery Kuznetsov, Konstantin A. Murashov, Ivan S. Chernonosova, Vera S. Chelobanov, Boris P. Stepanova, Alena O. Sergeevichev, David S. Karpenko, Andrey A. Laktionov, Pavel P. Polymers (Basel) Article A stenting procedure aimed at blood flow restoration in stenosed arteries significantly improves the efficiency of vascular surgery. However, the current challenge is to prevent neointimal growth, which reduces the vessel lumen, in the stented segments in the long run. We tested in vivo drug-eluting coating applied by electrospinning to metal vascular stents to inhibit the overgrowth of neointimal cells via both the drug release and mechanical support of the vascular wall. The blend of polycaprolactone with human serum albumin and paclitaxel was used for stent coating by electrospinning. The drug-eluting stents (DESs) were placed using a balloon catheter to the rabbit common iliac artery for 1, 3, and 6 months. The blood flow rate was ultrasonically determined in vivo. After explantation, the stented arterial segment was visually and histologically examined. Any undesirable biological responses (rejection or hemodynamically significant stenosis) were unobservable in the experimental groups. DESs were less traumatic and induced weaker neointimal growth; over six months, the blood flow increased by 37% versus bare-metal stents, where it increased by at least double the rate. Thus, electrospun-coated DESs demonstrate considerable advantages over the bare-metal variants. MDPI 2020-08-04 /pmc/articles/PMC7465440/ /pubmed/32759856 http://dx.doi.org/10.3390/polym12081741 Text en © 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Kuznetsov, Konstantin A. Murashov, Ivan S. Chernonosova, Vera S. Chelobanov, Boris P. Stepanova, Alena O. Sergeevichev, David S. Karpenko, Andrey A. Laktionov, Pavel P. Vascular Stents Coated with Electrospun Drug-Eluting Material: Functioning in Rabbit Iliac Artery |
title | Vascular Stents Coated with Electrospun Drug-Eluting Material: Functioning in Rabbit Iliac Artery |
title_full | Vascular Stents Coated with Electrospun Drug-Eluting Material: Functioning in Rabbit Iliac Artery |
title_fullStr | Vascular Stents Coated with Electrospun Drug-Eluting Material: Functioning in Rabbit Iliac Artery |
title_full_unstemmed | Vascular Stents Coated with Electrospun Drug-Eluting Material: Functioning in Rabbit Iliac Artery |
title_short | Vascular Stents Coated with Electrospun Drug-Eluting Material: Functioning in Rabbit Iliac Artery |
title_sort | vascular stents coated with electrospun drug-eluting material: functioning in rabbit iliac artery |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7465440/ https://www.ncbi.nlm.nih.gov/pubmed/32759856 http://dx.doi.org/10.3390/polym12081741 |
work_keys_str_mv | AT kuznetsovkonstantina vascularstentscoatedwithelectrospundrugelutingmaterialfunctioninginrabbitiliacartery AT murashovivans vascularstentscoatedwithelectrospundrugelutingmaterialfunctioninginrabbitiliacartery AT chernonosovaveras vascularstentscoatedwithelectrospundrugelutingmaterialfunctioninginrabbitiliacartery AT chelobanovborisp vascularstentscoatedwithelectrospundrugelutingmaterialfunctioninginrabbitiliacartery AT stepanovaalenao vascularstentscoatedwithelectrospundrugelutingmaterialfunctioninginrabbitiliacartery AT sergeevichevdavids vascularstentscoatedwithelectrospundrugelutingmaterialfunctioninginrabbitiliacartery AT karpenkoandreya vascularstentscoatedwithelectrospundrugelutingmaterialfunctioninginrabbitiliacartery AT laktionovpavelp vascularstentscoatedwithelectrospundrugelutingmaterialfunctioninginrabbitiliacartery |