Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter

OBJECTIVE: To compare the radiation dose and the objective and subjective image quality of 80 kVp and 80/150 kVp with tin filter (80/Sn150 kVp) computed tomography (CT) in oncology patients. METHODS: One-hundred-and-forty-five consecutive oncology patients who underwent third-generation dual-source...

Descripción completa

Detalles Bibliográficos
Autores principales: Choi, Seung Joon, Ahn, Su Joa, Park, So Hyun, Park, Seong Ho, Pak, Seong Yong, Choi, Jae Won, Shim, Young Sup, Jeong, Yu Mi, Kim, Bohyun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7470424/
https://www.ncbi.nlm.nih.gov/pubmed/32881876
http://dx.doi.org/10.1371/journal.pone.0231431
_version_ 1783578585991217152
author Choi, Seung Joon
Ahn, Su Joa
Park, So Hyun
Park, Seong Ho
Pak, Seong Yong
Choi, Jae Won
Shim, Young Sup
Jeong, Yu Mi
Kim, Bohyun
author_facet Choi, Seung Joon
Ahn, Su Joa
Park, So Hyun
Park, Seong Ho
Pak, Seong Yong
Choi, Jae Won
Shim, Young Sup
Jeong, Yu Mi
Kim, Bohyun
author_sort Choi, Seung Joon
collection PubMed
description OBJECTIVE: To compare the radiation dose and the objective and subjective image quality of 80 kVp and 80/150 kVp with tin filter (80/Sn150 kVp) computed tomography (CT) in oncology patients. METHODS: One-hundred-and-forty-five consecutive oncology patients who underwent third-generation dual-source dual-energy CT of the abdomen for evaluation of malignant visceral, peritoneal, extraperitoneal, and bone tumor were retrospectively recruited. Two radiologists independently reviewed each observation in 80 kVp CT and 80/Sn150 kVp CT. Modified line-density profile of the tumor and contrast-to-noise ratio (CNR) were measured. Diagnostic confidence, lesion conspicuity, and subjective image quality were calculated and compared between image sets. The effective dose and size-specific dose estimate (SSDE) were calculated in the image sets. RESULTS: Modified line-density profile analysis revealed higher attenuation differences between the tumor and normal tissue in 80 kVp CT than in 80/Sn150 kVp CT (127 vs. 107, P = 0.05). The 80 kVp CT showed increased CNR in the liver (8.0 vs. 7.6) and the aorta (18.9 vs. 16.3) than the 80/Sn150 kVp CT. The 80 kVp CT yielded higher enhancement of organs (4.9 ± 0.2 vs. 4.7 ± 0.4, P<0.001) and lesion conspicuity (4.9 ± 0.3 vs. 4.8 ± 0.5, P = 0.035) than the 80/Sn150 kVp CT; overall image quality and confidence index were comparable. The effective dose was reduced by 45.2% with 80 kVp CT (2.3 mSv ± 0.9) compared to 80/Sn150 kVp CT (4.1 mSv ± 1.5). The SSDE was 7.4 ± 3.8 mGy on 80/Sn150 kVp CT and 4.1 ± 2.2 mGy on 80 kVp CT. CONCLUSIONS: The 80 kVp CT reduced the radiation dose by 45.2% in oncology patients while showing comparable or superior image quality to that of 80/Sn150 kVp CT for abdominal tumor evaluation.
format Online
Article
Text
id pubmed-7470424
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-74704242020-09-11 Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter Choi, Seung Joon Ahn, Su Joa Park, So Hyun Park, Seong Ho Pak, Seong Yong Choi, Jae Won Shim, Young Sup Jeong, Yu Mi Kim, Bohyun PLoS One Research Article OBJECTIVE: To compare the radiation dose and the objective and subjective image quality of 80 kVp and 80/150 kVp with tin filter (80/Sn150 kVp) computed tomography (CT) in oncology patients. METHODS: One-hundred-and-forty-five consecutive oncology patients who underwent third-generation dual-source dual-energy CT of the abdomen for evaluation of malignant visceral, peritoneal, extraperitoneal, and bone tumor were retrospectively recruited. Two radiologists independently reviewed each observation in 80 kVp CT and 80/Sn150 kVp CT. Modified line-density profile of the tumor and contrast-to-noise ratio (CNR) were measured. Diagnostic confidence, lesion conspicuity, and subjective image quality were calculated and compared between image sets. The effective dose and size-specific dose estimate (SSDE) were calculated in the image sets. RESULTS: Modified line-density profile analysis revealed higher attenuation differences between the tumor and normal tissue in 80 kVp CT than in 80/Sn150 kVp CT (127 vs. 107, P = 0.05). The 80 kVp CT showed increased CNR in the liver (8.0 vs. 7.6) and the aorta (18.9 vs. 16.3) than the 80/Sn150 kVp CT. The 80 kVp CT yielded higher enhancement of organs (4.9 ± 0.2 vs. 4.7 ± 0.4, P<0.001) and lesion conspicuity (4.9 ± 0.3 vs. 4.8 ± 0.5, P = 0.035) than the 80/Sn150 kVp CT; overall image quality and confidence index were comparable. The effective dose was reduced by 45.2% with 80 kVp CT (2.3 mSv ± 0.9) compared to 80/Sn150 kVp CT (4.1 mSv ± 1.5). The SSDE was 7.4 ± 3.8 mGy on 80/Sn150 kVp CT and 4.1 ± 2.2 mGy on 80 kVp CT. CONCLUSIONS: The 80 kVp CT reduced the radiation dose by 45.2% in oncology patients while showing comparable or superior image quality to that of 80/Sn150 kVp CT for abdominal tumor evaluation. Public Library of Science 2020-09-03 /pmc/articles/PMC7470424/ /pubmed/32881876 http://dx.doi.org/10.1371/journal.pone.0231431 Text en © 2020 Choi et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Choi, Seung Joon
Ahn, Su Joa
Park, So Hyun
Park, Seong Ho
Pak, Seong Yong
Choi, Jae Won
Shim, Young Sup
Jeong, Yu Mi
Kim, Bohyun
Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter
title Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter
title_full Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter
title_fullStr Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter
title_full_unstemmed Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter
title_short Dual-source abdominopelvic computed tomography: Comparison of image quality and radiation dose of 80 kVp and 80/150 kVp with tin filter
title_sort dual-source abdominopelvic computed tomography: comparison of image quality and radiation dose of 80 kvp and 80/150 kvp with tin filter
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7470424/
https://www.ncbi.nlm.nih.gov/pubmed/32881876
http://dx.doi.org/10.1371/journal.pone.0231431
work_keys_str_mv AT choiseungjoon dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT ahnsujoa dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT parksohyun dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT parkseongho dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT pakseongyong dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT choijaewon dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT shimyoungsup dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT jeongyumi dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter
AT kimbohyun dualsourceabdominopelviccomputedtomographycomparisonofimagequalityandradiationdoseof80kvpand80150kvpwithtinfilter