Cargando…

From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450

BACKGROUND: The purpose of this work was to propose an approach based on noise measurement to adapt present clinical acquisition and reconstruction parameters adapted to a PMT-based system (Biograph mCT) to a SiPM-based system (Biograph Vision 450) sharing identical geometrical properties. The NEMA...

Descripción completa

Detalles Bibliográficos
Autores principales: Carlier, Thomas, Ferrer, Ludovic, Conti, Maurizio, Bodet-Milin, Caroline, Rousseau, Caroline, Bercier, Yanic, Bendriem, Bernard, Kraeber-Bodéré, Françoise
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7471223/
https://www.ncbi.nlm.nih.gov/pubmed/32880792
http://dx.doi.org/10.1186/s40658-020-00323-w
_version_ 1783578738212995072
author Carlier, Thomas
Ferrer, Ludovic
Conti, Maurizio
Bodet-Milin, Caroline
Rousseau, Caroline
Bercier, Yanic
Bendriem, Bernard
Kraeber-Bodéré, Françoise
author_facet Carlier, Thomas
Ferrer, Ludovic
Conti, Maurizio
Bodet-Milin, Caroline
Rousseau, Caroline
Bercier, Yanic
Bendriem, Bernard
Kraeber-Bodéré, Françoise
author_sort Carlier, Thomas
collection PubMed
description BACKGROUND: The purpose of this work was to propose an approach based on noise measurement to adapt present clinical acquisition and reconstruction parameters adapted to a PMT-based system (Biograph mCT) to a SiPM-based system (Biograph Vision 450) sharing identical geometrical properties. The NEMA performance (NEMA) of the recently released Biograph Vision 450 PET/CT (Vision) was also derived. METHODS: All measurements were conducted on Vision and Biograph mCT with TrueV (mCT). A full NEMA-based performance was derived for Vision only. The adaptation of acquisition and reconstruction parameters from mCT to Vision was done using the NEMA image quality phantom. The noise level reached using mCT was set as the reference value for six different numbers of net true coincidences. The noise level computed using Vision was matched to the reference noise level (within 0.01%) using a different reconstruction set-up to determine the potential reduction of count numbers for the same noise level. RESULTS: Vision sensitivity was 9.1 kcps/MBq for a timing resolution of 213 ps at 5.3 kBq/mL. The NEMA-based CR for the 10-mm sphere was better than 75% regardless the reconstruction set-up studied. The mCT reference noise properties could be achieved using Vision with a scan time reduction (STR) of 1.34 with four iterations and a 440 × 440 matrix size (or STR = 1.89 with a 220 × 220 matrix size) together with a 3D CR improvement of 53% for the 10-mm sphere (24% using 220 × 220). CONCLUSION: The Vision exhibited improved NEMA performances compared to mCT. Using the proposed approach, the time acquisition could be divided by almost two, while keeping the same noise properties as that of mCT with a marked improvement of contrast recovery.
format Online
Article
Text
id pubmed-7471223
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-74712232020-09-15 From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450 Carlier, Thomas Ferrer, Ludovic Conti, Maurizio Bodet-Milin, Caroline Rousseau, Caroline Bercier, Yanic Bendriem, Bernard Kraeber-Bodéré, Françoise EJNMMI Phys Original Research BACKGROUND: The purpose of this work was to propose an approach based on noise measurement to adapt present clinical acquisition and reconstruction parameters adapted to a PMT-based system (Biograph mCT) to a SiPM-based system (Biograph Vision 450) sharing identical geometrical properties. The NEMA performance (NEMA) of the recently released Biograph Vision 450 PET/CT (Vision) was also derived. METHODS: All measurements were conducted on Vision and Biograph mCT with TrueV (mCT). A full NEMA-based performance was derived for Vision only. The adaptation of acquisition and reconstruction parameters from mCT to Vision was done using the NEMA image quality phantom. The noise level reached using mCT was set as the reference value for six different numbers of net true coincidences. The noise level computed using Vision was matched to the reference noise level (within 0.01%) using a different reconstruction set-up to determine the potential reduction of count numbers for the same noise level. RESULTS: Vision sensitivity was 9.1 kcps/MBq for a timing resolution of 213 ps at 5.3 kBq/mL. The NEMA-based CR for the 10-mm sphere was better than 75% regardless the reconstruction set-up studied. The mCT reference noise properties could be achieved using Vision with a scan time reduction (STR) of 1.34 with four iterations and a 440 × 440 matrix size (or STR = 1.89 with a 220 × 220 matrix size) together with a 3D CR improvement of 53% for the 10-mm sphere (24% using 220 × 220). CONCLUSION: The Vision exhibited improved NEMA performances compared to mCT. Using the proposed approach, the time acquisition could be divided by almost two, while keeping the same noise properties as that of mCT with a marked improvement of contrast recovery. Springer International Publishing 2020-09-03 /pmc/articles/PMC7471223/ /pubmed/32880792 http://dx.doi.org/10.1186/s40658-020-00323-w Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Original Research
Carlier, Thomas
Ferrer, Ludovic
Conti, Maurizio
Bodet-Milin, Caroline
Rousseau, Caroline
Bercier, Yanic
Bendriem, Bernard
Kraeber-Bodéré, Françoise
From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450
title From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450
title_full From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450
title_fullStr From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450
title_full_unstemmed From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450
title_short From a PMT-based to a SiPM-based PET system: a study to define matched acquisition/reconstruction parameters and NEMA performance of the Biograph Vision 450
title_sort from a pmt-based to a sipm-based pet system: a study to define matched acquisition/reconstruction parameters and nema performance of the biograph vision 450
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7471223/
https://www.ncbi.nlm.nih.gov/pubmed/32880792
http://dx.doi.org/10.1186/s40658-020-00323-w
work_keys_str_mv AT carlierthomas fromapmtbasedtoasipmbasedpetsystemastudytodefinematchedacquisitionreconstructionparametersandnemaperformanceofthebiographvision450
AT ferrerludovic fromapmtbasedtoasipmbasedpetsystemastudytodefinematchedacquisitionreconstructionparametersandnemaperformanceofthebiographvision450
AT contimaurizio fromapmtbasedtoasipmbasedpetsystemastudytodefinematchedacquisitionreconstructionparametersandnemaperformanceofthebiographvision450
AT bodetmilincaroline fromapmtbasedtoasipmbasedpetsystemastudytodefinematchedacquisitionreconstructionparametersandnemaperformanceofthebiographvision450
AT rousseaucaroline fromapmtbasedtoasipmbasedpetsystemastudytodefinematchedacquisitionreconstructionparametersandnemaperformanceofthebiographvision450
AT bercieryanic fromapmtbasedtoasipmbasedpetsystemastudytodefinematchedacquisitionreconstructionparametersandnemaperformanceofthebiographvision450
AT bendriembernard fromapmtbasedtoasipmbasedpetsystemastudytodefinematchedacquisitionreconstructionparametersandnemaperformanceofthebiographvision450
AT kraeberboderefrancoise fromapmtbasedtoasipmbasedpetsystemastudytodefinematchedacquisitionreconstructionparametersandnemaperformanceofthebiographvision450