Comparison of Dilution on Eastern Box Turtle (Terrapene carolina carolina) and Marine Toad (Rhinella marinus) Blood Parameters as Measured on a Portable Chemistry Analyzer

Biochemical testing is an important clinical tool in evaluating the physiology of reptiles and amphibians. Suitable point of care analyzers can allow for rapid delivery of results, but small patient size can inhibit sufficient sample collection. This study evaluated the utility of sample dilution wi...

Descripción completa

Detalles Bibliográficos
Autores principales: Griffioen, John A., Stowe, Devorah M., Trosclair, Macy, Minter, Larry J., Vanetten, Chelsey, Harrison, Tara M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7474372/
https://www.ncbi.nlm.nih.gov/pubmed/32908664
http://dx.doi.org/10.1155/2020/8843058
Descripción
Sumario:Biochemical testing is an important clinical tool in evaluating the physiology of reptiles and amphibians. Suitable point of care analyzers can allow for rapid delivery of results, but small patient size can inhibit sufficient sample collection. This study evaluated the utility of sample dilution with sterile distilled water as a means of biochemical evaluation when sample volume is limited. Blood was collected from 12 eastern box turtles (Terrapene carolina carolina) and 12 marine toads (Rhinella marinus) and analyzed via i-STAT CHEM8+ cartridges. Two undiluted samples and two samples diluted 1 : 1 with sterile water were evaluated immediately following collection for each animal in the study. Analytes reported in the diluted samples were limited to glucose, ionized calcium, and total carbon dioxide. The expected dilution ratio value of diluted to undiluted samples was 0.5, of which glucose in both turtles and toads was nearest. Dilution ratio values for ionized calcium, however, were higher than expected in both turtles and toads. Sample dilution is not recommended for most analytes included on the CHEM8+ cartridge due to values occurring outside the limits of detection for the analyzer. Glucose and ionized calcium values obtained on diluted samples should be interpreted with caution but may provide clinical utility in reptile and amphibian patients where sample volume is limited.