Cargando…
Ceramic Liner Fracture Caused by an Impingement between the Stem Shoulder and the Ceramic Liner
Mechanisms of ceramic liner fractures have not yet been fully elucidated. Impingement between a stem and ceramic liner is a proposed cause of ceramic liner fractures. We experienced a case of ceramic liner fracture caused by direct impingement between the stem shoulder and the ceramic liner. This ty...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Hip Society
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7476783/ https://www.ncbi.nlm.nih.gov/pubmed/32953708 http://dx.doi.org/10.5371/hp.2020.32.3.156 |
Sumario: | Mechanisms of ceramic liner fractures have not yet been fully elucidated. Impingement between a stem and ceramic liner is a proposed cause of ceramic liner fractures. We experienced a case of ceramic liner fracture caused by direct impingement between the stem shoulder and the ceramic liner. This type of impingement, unlike impingements with a stem neck, has not been previously reported. While we assume that certain characteristics of the stem contributed to the impingement, we report this case to note that caution may be needed when using certain stem designs. |
---|