Performance of an automated chemiluminescence SARS-CoV-2 IG-G assay

INTRODUCTION: We describe our evaluation of the Abbott SARS-CoV-2 IgG assay on the Architect immunoassay analyser. METHODS: We assessed assay precision, sensitivity, specificity, positive/negative predictive values (PPV/NPV), cross-reactivity (influenza/dengue/hepatitis B and C/rheumatoid factor/ant...

Descripción completa

Detalles Bibliográficos
Autores principales: Lau, C.S., Oh, H.M.L., Hoo, S.P., Liang, Y.L., Phua, S.K., Aw, T.C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier B.V. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7476884/
https://www.ncbi.nlm.nih.gov/pubmed/32910979
http://dx.doi.org/10.1016/j.cca.2020.09.005
Descripción
Sumario:INTRODUCTION: We describe our evaluation of the Abbott SARS-CoV-2 IgG assay on the Architect immunoassay analyser. METHODS: We assessed assay precision, sensitivity, specificity, positive/negative predictive values (PPV/NPV), cross-reactivity (influenza/dengue/hepatitis B and C/rheumatoid factor/anti-nuclear/double-stranded DNA/syphilis) and sample throughput in samples from real-time polymerase chain reaction (RT-PCR) positive patients/healthcare workers (HCWs)/pre-pandemic samples. We compared the cut-off indexes (COIs) between all control samples (HCWs and pre-pandemic) to generate an optimised COI limit for reactivity. RESULTS: The assay specificity was 99.8% (n = 980) and sensitivity was 45.9–96.7% (n = 279). When tested ≥ 14 days post-positive RT-PCR (POS), the PPV/NPV was 96.4%/99.8%. The difference between the COIs of HCWs/pre-pandemic samples was small (0.01, p < 0.0001). There was minimal cross-reactivity with other antibodies. A lower COI limit for reactivity (≥0.55, using the 99th percentile COI of our controls and ROC analysis) improved diagnostic sensitivity, especially at 0–6 days POS (45.9–55.8%), with a small decrease in specificity (98.9%). The assay throughput was 100 samples in 70 min. CONCLUSION: The Abbott SARS-CoV-2 IgG assay shows excellent performance in patients ≥ 14 days POS. The difference between the COIs of HCWs and pre-pandemic samples was numerically small. A lower COI limit improves assay sensitivity with a slight decrease in specificity.