Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS)

BACKGROUND: Blood pressure (BP) measurement accuracy is critical to the diagnosis and management of hypertension. The aim of the present study was to validate the Omron HBP-1100-E professional blood pressure measuring device in accordance with the American Association for the Advancement of Medical...

Descripción completa

Detalles Bibliográficos
Autores principales: Joukar, Farahnaz, Yeganeh, Sara, Naghipour, Mohammadreza, Hassanipour, Soheil, Nikbakht, Hossein-Ali, Mansour-Ghanaei, Fariborz
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7478364/
https://www.ncbi.nlm.nih.gov/pubmed/32943949
http://dx.doi.org/10.2147/MDER.S253638
_version_ 1783580033980301312
author Joukar, Farahnaz
Yeganeh, Sara
Naghipour, Mohammadreza
Hassanipour, Soheil
Nikbakht, Hossein-Ali
Mansour-Ghanaei, Fariborz
author_facet Joukar, Farahnaz
Yeganeh, Sara
Naghipour, Mohammadreza
Hassanipour, Soheil
Nikbakht, Hossein-Ali
Mansour-Ghanaei, Fariborz
author_sort Joukar, Farahnaz
collection PubMed
description BACKGROUND: Blood pressure (BP) measurement accuracy is critical to the diagnosis and management of hypertension. The aim of the present study was to validate the Omron HBP-1100-E professional blood pressure measuring device in accordance with the American Association for the Advancement of Medical Instrumentation in Iranian adults. MATERIALS AND METHODS: Simultaneous blood pressure auscultator measurements were obtained by two observers using mercury sphygmomanometers as a reference, sequentially with a measurement by using the Omron HBP-1100-E device. Absolute device-reference blood pressure differences were categorized into three error categories (within 5, 10, and 15 mmHg), and mean device-reference blood pressure difference (standard deviation) was calculated and evaluated using the American Association for the Advancement of Medical Instrumentation criteria. RESULTS: A total of 85 participants (250 paired readings) were enrolled to the study. 26.8%, 55.6%, and 79.6% of the device-reference blood pressure differences agreed to within 5, 10 and 15 mmHg, respectively, for systolic blood pressure, and 39.6%, 69.2%, and 81.6% of measurements for diastolic blood pressure, respectively, and failed to pass the protocol criteria. The mean device-reference blood pressure difference was 8.0 ± 13.1 mmHg for systolic BP and 2.2 ± 11.3 mmHg for diastolic BP, and was >5.0 ± 8.0 mmHg (required criteria). CONCLUSION: Omron HBP-1100-E professional blood pressure monitor is not desirable for measuring the BP for Iranian adults as it overestimates blood pressure in this population.
format Online
Article
Text
id pubmed-7478364
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Dove
record_format MEDLINE/PubMed
spelling pubmed-74783642020-09-16 Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS) Joukar, Farahnaz Yeganeh, Sara Naghipour, Mohammadreza Hassanipour, Soheil Nikbakht, Hossein-Ali Mansour-Ghanaei, Fariborz Med Devices (Auckl) Original Research BACKGROUND: Blood pressure (BP) measurement accuracy is critical to the diagnosis and management of hypertension. The aim of the present study was to validate the Omron HBP-1100-E professional blood pressure measuring device in accordance with the American Association for the Advancement of Medical Instrumentation in Iranian adults. MATERIALS AND METHODS: Simultaneous blood pressure auscultator measurements were obtained by two observers using mercury sphygmomanometers as a reference, sequentially with a measurement by using the Omron HBP-1100-E device. Absolute device-reference blood pressure differences were categorized into three error categories (within 5, 10, and 15 mmHg), and mean device-reference blood pressure difference (standard deviation) was calculated and evaluated using the American Association for the Advancement of Medical Instrumentation criteria. RESULTS: A total of 85 participants (250 paired readings) were enrolled to the study. 26.8%, 55.6%, and 79.6% of the device-reference blood pressure differences agreed to within 5, 10 and 15 mmHg, respectively, for systolic blood pressure, and 39.6%, 69.2%, and 81.6% of measurements for diastolic blood pressure, respectively, and failed to pass the protocol criteria. The mean device-reference blood pressure difference was 8.0 ± 13.1 mmHg for systolic BP and 2.2 ± 11.3 mmHg for diastolic BP, and was >5.0 ± 8.0 mmHg (required criteria). CONCLUSION: Omron HBP-1100-E professional blood pressure monitor is not desirable for measuring the BP for Iranian adults as it overestimates blood pressure in this population. Dove 2020-09-03 /pmc/articles/PMC7478364/ /pubmed/32943949 http://dx.doi.org/10.2147/MDER.S253638 Text en © 2020 Joukar et al. http://creativecommons.org/licenses/by-nc/3.0/ This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
spellingShingle Original Research
Joukar, Farahnaz
Yeganeh, Sara
Naghipour, Mohammadreza
Hassanipour, Soheil
Nikbakht, Hossein-Ali
Mansour-Ghanaei, Fariborz
Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS)
title Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS)
title_full Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS)
title_fullStr Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS)
title_full_unstemmed Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS)
title_short Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS)
title_sort validation of omron hbp-1100-e professional blood pressure measuring device according to the american association for the advancement of medical instrumentation protocol: the persian guilan cohort study (pgcs)
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7478364/
https://www.ncbi.nlm.nih.gov/pubmed/32943949
http://dx.doi.org/10.2147/MDER.S253638
work_keys_str_mv AT joukarfarahnaz validationofomronhbp1100eprofessionalbloodpressuremeasuringdeviceaccordingtotheamericanassociationfortheadvancementofmedicalinstrumentationprotocolthepersianguilancohortstudypgcs
AT yeganehsara validationofomronhbp1100eprofessionalbloodpressuremeasuringdeviceaccordingtotheamericanassociationfortheadvancementofmedicalinstrumentationprotocolthepersianguilancohortstudypgcs
AT naghipourmohammadreza validationofomronhbp1100eprofessionalbloodpressuremeasuringdeviceaccordingtotheamericanassociationfortheadvancementofmedicalinstrumentationprotocolthepersianguilancohortstudypgcs
AT hassanipoursoheil validationofomronhbp1100eprofessionalbloodpressuremeasuringdeviceaccordingtotheamericanassociationfortheadvancementofmedicalinstrumentationprotocolthepersianguilancohortstudypgcs
AT nikbakhthosseinali validationofomronhbp1100eprofessionalbloodpressuremeasuringdeviceaccordingtotheamericanassociationfortheadvancementofmedicalinstrumentationprotocolthepersianguilancohortstudypgcs
AT mansourghanaeifariborz validationofomronhbp1100eprofessionalbloodpressuremeasuringdeviceaccordingtotheamericanassociationfortheadvancementofmedicalinstrumentationprotocolthepersianguilancohortstudypgcs