BACKGROUND: Emergency carotid artery stenting (CAS) is a frequent endovascular procedure, especially in combination with intracranial thrombectomy. Balloon guide catheters are frequently used in these procedures. Our aim was to determine if mechanical aspiration through the working lumen of a balloo...

Descripción completa

Detalles Bibliográficos
Autores principales: Schubert, Tilman, Rivera-Rivera, Leonardo, Roldan-Alzate, Alejandro, Consigny, Daniel, Leitner, Lorenz, Strother, Charles, Aagaard-Kienitz, Beverly
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7483693/
https://www.ncbi.nlm.nih.gov/pubmed/32910271
http://dx.doi.org/10.1186/s42155-020-00134-1
_version_ 1783580951624810496
author Schubert, Tilman
Rivera-Rivera, Leonardo
Roldan-Alzate, Alejandro
Consigny, Daniel
Leitner, Lorenz
Strother, Charles
Aagaard-Kienitz, Beverly
author_facet Schubert, Tilman
Rivera-Rivera, Leonardo
Roldan-Alzate, Alejandro
Consigny, Daniel
Leitner, Lorenz
Strother, Charles
Aagaard-Kienitz, Beverly
author_sort Schubert, Tilman
collection PubMed
description BACKGROUND: Emergency carotid artery stenting (CAS) is a frequent endovascular procedure, especially in combination with intracranial thrombectomy. Balloon guide catheters are frequently used in these procedures. Our aim was to determine if mechanical aspiration through the working lumen of a balloon occlusion catheter during the steps of a carotid stenting procedure achieve flow rates that may lead to internal carotid artery (ICA) flow reversal which consecutively may prevent distal embolism. METHODS: Aspiration experiments were conducted using a commercially available aspiration pump. Aspiration flow rates/min with 6 different types of carotid stents inserted into a balloon guide catheter were measured. Measurements were repeated three times with increasing pressure in the phantom. To determine if the achieved aspiration flow rates were similar to physiologic values, flow rates in the ICA and external carotid artery (ECA) in 10 healthy volunteers were measured using 4D-flow MRI. RESULTS: Aspiration flow rates ranged from 25 to 82 mL/min depending on the stent model. The pressure in the phantom had a significant influence on the aspiration volume. Mean blood flow volumes in volunteers were 210 mL/min in the ICA and 101 mL/min in the ECA. CONCLUSIONS: Based on the results of this study, flow reversal in the ICA during common carotid artery occlusion is most likely achieved with the smallest diameter stent sheath and the stent model with the shortest outer stent sheath maximum diameter. This implies that embolic protection during emergency CAS through aspiration is most effective with these models.
format Online
Article
Text
id pubmed-7483693
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-74836932020-09-21 Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting Schubert, Tilman Rivera-Rivera, Leonardo Roldan-Alzate, Alejandro Consigny, Daniel Leitner, Lorenz Strother, Charles Aagaard-Kienitz, Beverly CVIR Endovasc Original Article BACKGROUND: Emergency carotid artery stenting (CAS) is a frequent endovascular procedure, especially in combination with intracranial thrombectomy. Balloon guide catheters are frequently used in these procedures. Our aim was to determine if mechanical aspiration through the working lumen of a balloon occlusion catheter during the steps of a carotid stenting procedure achieve flow rates that may lead to internal carotid artery (ICA) flow reversal which consecutively may prevent distal embolism. METHODS: Aspiration experiments were conducted using a commercially available aspiration pump. Aspiration flow rates/min with 6 different types of carotid stents inserted into a balloon guide catheter were measured. Measurements were repeated three times with increasing pressure in the phantom. To determine if the achieved aspiration flow rates were similar to physiologic values, flow rates in the ICA and external carotid artery (ECA) in 10 healthy volunteers were measured using 4D-flow MRI. RESULTS: Aspiration flow rates ranged from 25 to 82 mL/min depending on the stent model. The pressure in the phantom had a significant influence on the aspiration volume. Mean blood flow volumes in volunteers were 210 mL/min in the ICA and 101 mL/min in the ECA. CONCLUSIONS: Based on the results of this study, flow reversal in the ICA during common carotid artery occlusion is most likely achieved with the smallest diameter stent sheath and the stent model with the shortest outer stent sheath maximum diameter. This implies that embolic protection during emergency CAS through aspiration is most effective with these models. Springer International Publishing 2020-09-10 /pmc/articles/PMC7483693/ /pubmed/32910271 http://dx.doi.org/10.1186/s42155-020-00134-1 Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Original Article
Schubert, Tilman
Rivera-Rivera, Leonardo
Roldan-Alzate, Alejandro
Consigny, Daniel
Leitner, Lorenz
Strother, Charles
Aagaard-Kienitz, Beverly
Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting
title Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting
title_full Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting
title_fullStr Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting
title_full_unstemmed Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting
title_short Achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting
title_sort achievable aspiration flow rates with large balloon guide catheters during carotid artery stenting
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7483693/
https://www.ncbi.nlm.nih.gov/pubmed/32910271
http://dx.doi.org/10.1186/s42155-020-00134-1
work_keys_str_mv AT schuberttilman achievableaspirationflowrateswithlargeballoonguidecathetersduringcarotidarterystenting
AT riverariveraleonardo achievableaspirationflowrateswithlargeballoonguidecathetersduringcarotidarterystenting
AT roldanalzatealejandro achievableaspirationflowrateswithlargeballoonguidecathetersduringcarotidarterystenting
AT consignydaniel achievableaspirationflowrateswithlargeballoonguidecathetersduringcarotidarterystenting
AT leitnerlorenz achievableaspirationflowrateswithlargeballoonguidecathetersduringcarotidarterystenting
AT strothercharles achievableaspirationflowrateswithlargeballoonguidecathetersduringcarotidarterystenting
AT aagaardkienitzbeverly achievableaspirationflowrateswithlargeballoonguidecathetersduringcarotidarterystenting