Machine Learning Model Based on Transthoracic Bioimpedance and Heart Rate Variability for Lung Fluid Accumulation Detection: Prospective Clinical Study

BACKGROUND: Accumulation of excess body fluid and autonomic dysregulation are clinically important characteristics of acute decompensated heart failure. We hypothesized that transthoracic bioimpedance, a noninvasive, simple method for measuring fluid retention in lungs, and heart rate variability, a...

Descripción completa

Detalles Bibliográficos
Autores principales: Reljin, Natasa, Posada-Quintero, Hugo F, Eaton-Robb, Caitlin, Binici, Sophia, Ensom, Emily, Ding, Eric, Hayes, Anna, Riistama, Jarno, Darling, Chad, McManus, David, Chon, Ki H
Formato: Online Artículo Texto
Lenguaje:English
Publicado: JMIR Publications 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7484776/
https://www.ncbi.nlm.nih.gov/pubmed/32852277
http://dx.doi.org/10.2196/18715
_version_ 1783581041296932864
author Reljin, Natasa
Posada-Quintero, Hugo F
Eaton-Robb, Caitlin
Binici, Sophia
Ensom, Emily
Ding, Eric
Hayes, Anna
Riistama, Jarno
Darling, Chad
McManus, David
Chon, Ki H
author_facet Reljin, Natasa
Posada-Quintero, Hugo F
Eaton-Robb, Caitlin
Binici, Sophia
Ensom, Emily
Ding, Eric
Hayes, Anna
Riistama, Jarno
Darling, Chad
McManus, David
Chon, Ki H
author_sort Reljin, Natasa
collection PubMed
description BACKGROUND: Accumulation of excess body fluid and autonomic dysregulation are clinically important characteristics of acute decompensated heart failure. We hypothesized that transthoracic bioimpedance, a noninvasive, simple method for measuring fluid retention in lungs, and heart rate variability, an assessment of autonomic function, can be used for detection of fluid accumulation in patients with acute decompensated heart failure. OBJECTIVE: We aimed to evaluate the performance of transthoracic bioimpedance and heart rate variability parameters obtained using a fluid accumulation vest with carbon black–polydimethylsiloxane dry electrodes in a prospective clinical study (System for Heart Failure Identification Using an External Lung Fluid Device; SHIELD). METHODS: We computed 15 parameters: 8 were calculated from the model to fit Cole-Cole plots from transthoracic bioimpedance measurements (extracellular, intracellular, intracellular-extracellular difference, and intracellular-extracellular parallel circuit resistances as well as fitting error, resonance frequency, tissue heterogeneity, and cellular membrane capacitance), and 7 were based on linear (mean heart rate, low-frequency components of heart rate variability, high-frequency components of heart rate variability, normalized low-frequency components of heart rate variability, normalized high-frequency components of heart rate variability) and nonlinear (principal dynamic mode index of sympathetic function, and principal dynamic mode index of parasympathetic function) analysis of heart rate variability. We compared the values of these parameters between 3 participant data sets: control (n=32, patients who did not have heart failure), baseline (n=23, patients with acute decompensated heart failure taken at the time of admittance to the hospital), and discharge (n=17, patients with acute decompensated heart failure taken at the time of discharge from hospital). We used several machine learning approaches to classify participants with fluid accumulation (baseline) and without fluid accumulation (control and discharge), termed with fluid and without fluid groups, respectively. RESULTS: Among the 15 parameters, 3 transthoracic bioimpedance (extracellular resistance, R(0); difference in extracellular-intracellular resistance, R(0) – R(∞), and tissue heterogeneity, α) and 3 heart rate variability (high-frequency, normalized low-frequency, and normalized high-frequency components) parameters were found to be the most discriminatory between groups (patients with and patients without heart failure). R(0) and R(0) – R(∞) had significantly lower values for patients with heart failure than for those without heart failure (R(0): P=.006; R(0) – R(∞): P=.001), indicating that a higher volume of fluids accumulated in the lungs of patients with heart failure. A cubic support vector machine model using the 5 parameters achieved an accuracy of 92% for with fluid and without fluid group classification. The transthoracic bioimpedance parameters were related to intra- and extracellular fluid, whereas the heart rate variability parameters were mostly related to sympathetic activation. CONCLUSIONS: This is useful, for instance, for an in-home diagnostic wearable to detect fluid accumulation. Results suggest that fluid accumulation, and subsequently acute decompensated heart failure detection, could be performed using transthoracic bioimpedance and heart rate variability measurements acquired with a wearable vest.
format Online
Article
Text
id pubmed-7484776
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher JMIR Publications
record_format MEDLINE/PubMed
spelling pubmed-74847762020-09-21 Machine Learning Model Based on Transthoracic Bioimpedance and Heart Rate Variability for Lung Fluid Accumulation Detection: Prospective Clinical Study Reljin, Natasa Posada-Quintero, Hugo F Eaton-Robb, Caitlin Binici, Sophia Ensom, Emily Ding, Eric Hayes, Anna Riistama, Jarno Darling, Chad McManus, David Chon, Ki H JMIR Med Inform Original Paper BACKGROUND: Accumulation of excess body fluid and autonomic dysregulation are clinically important characteristics of acute decompensated heart failure. We hypothesized that transthoracic bioimpedance, a noninvasive, simple method for measuring fluid retention in lungs, and heart rate variability, an assessment of autonomic function, can be used for detection of fluid accumulation in patients with acute decompensated heart failure. OBJECTIVE: We aimed to evaluate the performance of transthoracic bioimpedance and heart rate variability parameters obtained using a fluid accumulation vest with carbon black–polydimethylsiloxane dry electrodes in a prospective clinical study (System for Heart Failure Identification Using an External Lung Fluid Device; SHIELD). METHODS: We computed 15 parameters: 8 were calculated from the model to fit Cole-Cole plots from transthoracic bioimpedance measurements (extracellular, intracellular, intracellular-extracellular difference, and intracellular-extracellular parallel circuit resistances as well as fitting error, resonance frequency, tissue heterogeneity, and cellular membrane capacitance), and 7 were based on linear (mean heart rate, low-frequency components of heart rate variability, high-frequency components of heart rate variability, normalized low-frequency components of heart rate variability, normalized high-frequency components of heart rate variability) and nonlinear (principal dynamic mode index of sympathetic function, and principal dynamic mode index of parasympathetic function) analysis of heart rate variability. We compared the values of these parameters between 3 participant data sets: control (n=32, patients who did not have heart failure), baseline (n=23, patients with acute decompensated heart failure taken at the time of admittance to the hospital), and discharge (n=17, patients with acute decompensated heart failure taken at the time of discharge from hospital). We used several machine learning approaches to classify participants with fluid accumulation (baseline) and without fluid accumulation (control and discharge), termed with fluid and without fluid groups, respectively. RESULTS: Among the 15 parameters, 3 transthoracic bioimpedance (extracellular resistance, R(0); difference in extracellular-intracellular resistance, R(0) – R(∞), and tissue heterogeneity, α) and 3 heart rate variability (high-frequency, normalized low-frequency, and normalized high-frequency components) parameters were found to be the most discriminatory between groups (patients with and patients without heart failure). R(0) and R(0) – R(∞) had significantly lower values for patients with heart failure than for those without heart failure (R(0): P=.006; R(0) – R(∞): P=.001), indicating that a higher volume of fluids accumulated in the lungs of patients with heart failure. A cubic support vector machine model using the 5 parameters achieved an accuracy of 92% for with fluid and without fluid group classification. The transthoracic bioimpedance parameters were related to intra- and extracellular fluid, whereas the heart rate variability parameters were mostly related to sympathetic activation. CONCLUSIONS: This is useful, for instance, for an in-home diagnostic wearable to detect fluid accumulation. Results suggest that fluid accumulation, and subsequently acute decompensated heart failure detection, could be performed using transthoracic bioimpedance and heart rate variability measurements acquired with a wearable vest. JMIR Publications 2020-08-27 /pmc/articles/PMC7484776/ /pubmed/32852277 http://dx.doi.org/10.2196/18715 Text en ©Natasa Reljin, Hugo F. Posada-Quintero, Caitlin Eaton-Robb, Sophia Binici, Emily Ensom, Eric Ding, Anna Hayes, Jarno Riistama, Chad Darling, David McManus, Ki H. Chon. Originally published in JMIR Medical Informatics (http://medinform.jmir.org), 27.08.2020. https://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Medical Informatics, is properly cited. The complete bibliographic information, a link to the original publication on http://medinform.jmir.org/, as well as this copyright and license information must be included.
spellingShingle Original Paper
Reljin, Natasa
Posada-Quintero, Hugo F
Eaton-Robb, Caitlin
Binici, Sophia
Ensom, Emily
Ding, Eric
Hayes, Anna
Riistama, Jarno
Darling, Chad
McManus, David
Chon, Ki H
Machine Learning Model Based on Transthoracic Bioimpedance and Heart Rate Variability for Lung Fluid Accumulation Detection: Prospective Clinical Study
title Machine Learning Model Based on Transthoracic Bioimpedance and Heart Rate Variability for Lung Fluid Accumulation Detection: Prospective Clinical Study
title_full Machine Learning Model Based on Transthoracic Bioimpedance and Heart Rate Variability for Lung Fluid Accumulation Detection: Prospective Clinical Study
title_fullStr Machine Learning Model Based on Transthoracic Bioimpedance and Heart Rate Variability for Lung Fluid Accumulation Detection: Prospective Clinical Study
title_full_unstemmed Machine Learning Model Based on Transthoracic Bioimpedance and Heart Rate Variability for Lung Fluid Accumulation Detection: Prospective Clinical Study
title_short Machine Learning Model Based on Transthoracic Bioimpedance and Heart Rate Variability for Lung Fluid Accumulation Detection: Prospective Clinical Study
title_sort machine learning model based on transthoracic bioimpedance and heart rate variability for lung fluid accumulation detection: prospective clinical study
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7484776/
https://www.ncbi.nlm.nih.gov/pubmed/32852277
http://dx.doi.org/10.2196/18715
work_keys_str_mv AT reljinnatasa machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT posadaquinterohugof machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT eatonrobbcaitlin machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT binicisophia machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT ensomemily machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT dingeric machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT hayesanna machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT riistamajarno machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT darlingchad machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT mcmanusdavid machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy
AT chonkih machinelearningmodelbasedontransthoracicbioimpedanceandheartratevariabilityforlungfluidaccumulationdetectionprospectiveclinicalstudy