Automated and robust beam data validation of a preconfigured ring gantry linear accelerator using a 1D tank with synchronized beam delivery and couch motions
PURPOSE: To develop an efficient and automated methodology for beam data validation for a preconfigured ring gantry linear accelerator using scripting and a one‐dimensional (1D) tank with automated couch motions. MATERIALS AND METHODS: Using an application programming interface, a program was develo...
Autores principales: | Knutson, Nels C., Schmidt, Matthew C., Reynoso, Francisco J., Hao, Yao, Mazur, Thomas R., Laugeman, Eric, Hugo, Geoffrey, Mutic, Sasa, Li, H. Harold, Ngwa, Wilfred, Cai, Bin, Sajo, Erno |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7484825/ https://www.ncbi.nlm.nih.gov/pubmed/32614511 http://dx.doi.org/10.1002/acm2.12946 |
Ejemplares similares
-
Equivalency of beam scan data collection using a 1D tank and automated couch movements to traditional 3D tank measurements
por: Knutson, Nels C., et al.
Publicado: (2018) -
Commissioning and initial experience with the first clinical gantry‐mounted proton therapy system
por: Zhao, Tianyu, et al.
Publicado: (2016) -
Intracranial Stereotactic Radiation Therapy With a Jawless Ring Gantry Linear Accelerator Equipped With New Dual Layer Multileaf Collimator
por: Knutson, Nels C., et al.
Publicado: (2020) -
Collision indicator charts for gantry‐couch position combinations for Varian linacs
por: Becker, Stewart J.
Publicado: (2011) -
Validation of the preconfigured Varian Ethos Acuros XB Beam Model for treatment planning dose calculations: A dosimetric study
por: Hu, Yunfei, et al.
Publicado: (2020)