Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography

PURPOSE: To compare central corneal thickness (CCT), aqueous depth (AQD), and anterior chamber depth (ACD) measurements using the swept-source (CASIA SS-1000, Tomey, Japan) and time-domain (Visante, Carl Zeiss Meditec, USA) anterior segment optical coherence tomographers (OCT) in normal eyes. METHOD...

Descripción completa

Detalles Bibliográficos
Autores principales: Chen, Sisi, Gao, Rongrong, McAlinden, Colm, Ye, Junming, Wang, Yiran, Chen, Min, Huang, Jinhai, Sun, Yong, Yu, A-Yong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7487088/
https://www.ncbi.nlm.nih.gov/pubmed/32953169
http://dx.doi.org/10.1155/2020/9739878
_version_ 1783581420076138496
author Chen, Sisi
Gao, Rongrong
McAlinden, Colm
Ye, Junming
Wang, Yiran
Chen, Min
Huang, Jinhai
Sun, Yong
Yu, A-Yong
author_facet Chen, Sisi
Gao, Rongrong
McAlinden, Colm
Ye, Junming
Wang, Yiran
Chen, Min
Huang, Jinhai
Sun, Yong
Yu, A-Yong
author_sort Chen, Sisi
collection PubMed
description PURPOSE: To compare central corneal thickness (CCT), aqueous depth (AQD), and anterior chamber depth (ACD) measurements using the swept-source (CASIA SS-1000, Tomey, Japan) and time-domain (Visante, Carl Zeiss Meditec, USA) anterior segment optical coherence tomographers (OCT) in normal eyes. METHODS: Sixty-eight eyes of 68 subjects were included. Three consecutive scans of each subject were obtained using both devices in a random order by one experienced operator. Standard deviation (S(w)), coefficient of repeatability (CoR), coefficients of variation (CoV), and intraclass correlation coefficients (ICC) were used to evaluate the intraoperator repeatability. Agreement was assessed using the Bland–Altman plots and 95% limits of agreement (LoA). RESULTS: All measurements of the swept-source OCT (SS-OCT) and time-domain OCT (TD-OCT) showed high repeatability with low CoR (CCT: 2.34 μm and 6.16 μm; AQD: 0.05 mm and 0.09 mm; ACD: 0.06 mm and 0.09 mm), low CoV (CCT: 0.16% and 0.42%; AQD: 0.61% and 0.97%; ACD: 0.53% and 0.83%), and high ICC (>0.98). The mean CCT with SS-OCT was slightly thicker than the results with TD-OCT (difference = 4.55 ± 2.62 μm, P < 0.001). There was no statistically significant difference in AQD or ACD measurements between the two devices (0.01 ± 0.05 mm, P=0.111; 0.02 ± 0.05 mm, P=0.022, respectively). The 95% LoA between the SS-OCT and TD-OCT were −0.59 to 9.69 μm for CCT, −0.10 to 0.12 mm for AQD, and −0.09 to 0.12 mm for ACD. CONCLUSIONS: High levels of repeatability and agreement were found between the two devices for all three parameters, suggesting interchangeability. SS-OCT demonstrated superior repeatability compared with TD-OCT.
format Online
Article
Text
id pubmed-7487088
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-74870882020-09-17 Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography Chen, Sisi Gao, Rongrong McAlinden, Colm Ye, Junming Wang, Yiran Chen, Min Huang, Jinhai Sun, Yong Yu, A-Yong J Ophthalmol Research Article PURPOSE: To compare central corneal thickness (CCT), aqueous depth (AQD), and anterior chamber depth (ACD) measurements using the swept-source (CASIA SS-1000, Tomey, Japan) and time-domain (Visante, Carl Zeiss Meditec, USA) anterior segment optical coherence tomographers (OCT) in normal eyes. METHODS: Sixty-eight eyes of 68 subjects were included. Three consecutive scans of each subject were obtained using both devices in a random order by one experienced operator. Standard deviation (S(w)), coefficient of repeatability (CoR), coefficients of variation (CoV), and intraclass correlation coefficients (ICC) were used to evaluate the intraoperator repeatability. Agreement was assessed using the Bland–Altman plots and 95% limits of agreement (LoA). RESULTS: All measurements of the swept-source OCT (SS-OCT) and time-domain OCT (TD-OCT) showed high repeatability with low CoR (CCT: 2.34 μm and 6.16 μm; AQD: 0.05 mm and 0.09 mm; ACD: 0.06 mm and 0.09 mm), low CoV (CCT: 0.16% and 0.42%; AQD: 0.61% and 0.97%; ACD: 0.53% and 0.83%), and high ICC (>0.98). The mean CCT with SS-OCT was slightly thicker than the results with TD-OCT (difference = 4.55 ± 2.62 μm, P < 0.001). There was no statistically significant difference in AQD or ACD measurements between the two devices (0.01 ± 0.05 mm, P=0.111; 0.02 ± 0.05 mm, P=0.022, respectively). The 95% LoA between the SS-OCT and TD-OCT were −0.59 to 9.69 μm for CCT, −0.10 to 0.12 mm for AQD, and −0.09 to 0.12 mm for ACD. CONCLUSIONS: High levels of repeatability and agreement were found between the two devices for all three parameters, suggesting interchangeability. SS-OCT demonstrated superior repeatability compared with TD-OCT. Hindawi 2020-09-04 /pmc/articles/PMC7487088/ /pubmed/32953169 http://dx.doi.org/10.1155/2020/9739878 Text en Copyright © 2020 Sisi Chen et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Chen, Sisi
Gao, Rongrong
McAlinden, Colm
Ye, Junming
Wang, Yiran
Chen, Min
Huang, Jinhai
Sun, Yong
Yu, A-Yong
Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography
title Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography
title_full Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography
title_fullStr Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography
title_full_unstemmed Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography
title_short Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography
title_sort comparison of anterior ocular biometric measurements using swept-source and time-domain optical coherence tomography
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7487088/
https://www.ncbi.nlm.nih.gov/pubmed/32953169
http://dx.doi.org/10.1155/2020/9739878
work_keys_str_mv AT chensisi comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography
AT gaorongrong comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography
AT mcalindencolm comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography
AT yejunming comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography
AT wangyiran comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography
AT chenmin comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography
AT huangjinhai comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography
AT sunyong comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography
AT yuayong comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography