Measurement of coronary artery calcium volume using ultra-high-resolution computed tomography: A preliminary phantom and cadaver study
OBJECTIVES: In this phantom- and cadaver study we investigated the differences of coronary artery calcium (CAC) volume on ultra-high-resolution computed tomography (U-HRCT) scans and conventional CT. METHODS: We scanned a coronary calcium phantom and the coronary arteries of five cadavers using U-HR...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7490539/ https://www.ncbi.nlm.nih.gov/pubmed/32964073 http://dx.doi.org/10.1016/j.ejro.2020.100253 |
Sumario: | OBJECTIVES: In this phantom- and cadaver study we investigated the differences of coronary artery calcium (CAC) volume on ultra-high-resolution computed tomography (U-HRCT) scans and conventional CT. METHODS: We scanned a coronary calcium phantom and the coronary arteries of five cadavers using U-HRCT in normal- and super-high resolution (NR, SHR) mode. The NR mode was similar to conventional CT; 896 detector channels, a matrix size of 512, and a slice thickness of 0.5 mm were applied. In SHR mode, we used 1792 detector channels, a matrix size of 1024, and a slice thickness of 0.25 mm. The CAC volume on NR- and SHR images were recorded. Differences in the physical- and the calculated CAC volume were defined as the error value and compared between NR- and SHR images of the phantom. Differences between the CAC volume on NR- and SHR scans of the cadavers were also recorded. RESULTS: The mean error value was lower on SHR- than NR images of the phantom (14.0 %, SD 11.1 vs 20.1 %, SD 15.2, p = 0.01). The mean CAC volume was significantly higher on SHR- than NR images of the cadavers (153.4 mm(3), SD 161.0 vs 144.7 mm(3), SD 164.8, p < 0.01). CONCLUSIONS: As small calcifications were more clearly visualized on U-HRCT images in SHR mode than on conventional (NR) CT scans, SHR imaging may facilitate the accurate quantification of the CAC. |
---|