Patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D

PURPOSE: Single‐isocenter multiple brain metastasis stereotactic radiosurgery is an efficient treatment modality increasing in clinical practice. The need to provide accurate, patient‐specific quality assurance (QA) for these plans is met by several options. This study reviews some of these options...

Descripción completa

Detalles Bibliográficos
Autores principales: McCulloch, James, Pawlowski, Jamie, Kirby, Neil, Rasmussen, Karl, Shi, Zheng, Myers, Pamela, Stathakis, Sotirios, Papanikolaou, Niko, Saenz, Daniel L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7497916/
https://www.ncbi.nlm.nih.gov/pubmed/32681753
http://dx.doi.org/10.1002/acm2.12979
_version_ 1783583406244757504
author McCulloch, James
Pawlowski, Jamie
Kirby, Neil
Rasmussen, Karl
Shi, Zheng
Myers, Pamela
Stathakis, Sotirios
Papanikolaou, Niko
Saenz, Daniel L.
author_facet McCulloch, James
Pawlowski, Jamie
Kirby, Neil
Rasmussen, Karl
Shi, Zheng
Myers, Pamela
Stathakis, Sotirios
Papanikolaou, Niko
Saenz, Daniel L.
author_sort McCulloch, James
collection PubMed
description PURPOSE: Single‐isocenter multiple brain metastasis stereotactic radiosurgery is an efficient treatment modality increasing in clinical practice. The need to provide accurate, patient‐specific quality assurance (QA) for these plans is met by several options. This study reviews some of these options and explores the use of the Octavius 4D as a solution for patient‐specific plan quality assurance. METHODS: The Octavius 4D Modular Phantom (O4D) with the 1000 SRS array was evaluated in this study. The array consists of 977 liquid‐filled ion chambers. The center 5.5 cm × 5.5 cm area has a detector spacing of 2.5 mm. The ability of the O4D to reconstruct three‐dimensional (3D) dose was validated against a 3D gel dosimeter, ion chamber, and film measurements. After validation, 15 patients with 2–11 targets had their plans delivered to the phantom. The criteria used for the gamma calculation was 3%/1 mm. The portion of targets which were measurable by the phantom was countable. The accompanying software compiled the measured doses allowing each target to be counted from the measured dose distribution. RESULTS: Spatial resolution was sufficient to verify the high dose distributions characteristic of SRS. Amongst the 15 patients there were 74 targets. Of the 74 targets, 61 (82%) of them were visible on the measured dose distribution. The average gamma passing rate was 99.3% (with sample standard deviation of 0.68%). CONCLUSIONS: The high resolution provided by the O4D with 1000 SRS board insert allows for very high‐resolution measurement. This high resolution in turn can allow for high gamma passing rates. The O4D with the 1000 SRS array is an acceptable method of performing quality assurance for single‐isocenter multiple brain metastasis SRS.
format Online
Article
Text
id pubmed-7497916
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-74979162020-09-25 Patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D McCulloch, James Pawlowski, Jamie Kirby, Neil Rasmussen, Karl Shi, Zheng Myers, Pamela Stathakis, Sotirios Papanikolaou, Niko Saenz, Daniel L. J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: Single‐isocenter multiple brain metastasis stereotactic radiosurgery is an efficient treatment modality increasing in clinical practice. The need to provide accurate, patient‐specific quality assurance (QA) for these plans is met by several options. This study reviews some of these options and explores the use of the Octavius 4D as a solution for patient‐specific plan quality assurance. METHODS: The Octavius 4D Modular Phantom (O4D) with the 1000 SRS array was evaluated in this study. The array consists of 977 liquid‐filled ion chambers. The center 5.5 cm × 5.5 cm area has a detector spacing of 2.5 mm. The ability of the O4D to reconstruct three‐dimensional (3D) dose was validated against a 3D gel dosimeter, ion chamber, and film measurements. After validation, 15 patients with 2–11 targets had their plans delivered to the phantom. The criteria used for the gamma calculation was 3%/1 mm. The portion of targets which were measurable by the phantom was countable. The accompanying software compiled the measured doses allowing each target to be counted from the measured dose distribution. RESULTS: Spatial resolution was sufficient to verify the high dose distributions characteristic of SRS. Amongst the 15 patients there were 74 targets. Of the 74 targets, 61 (82%) of them were visible on the measured dose distribution. The average gamma passing rate was 99.3% (with sample standard deviation of 0.68%). CONCLUSIONS: The high resolution provided by the O4D with 1000 SRS board insert allows for very high‐resolution measurement. This high resolution in turn can allow for high gamma passing rates. The O4D with the 1000 SRS array is an acceptable method of performing quality assurance for single‐isocenter multiple brain metastasis SRS. John Wiley and Sons Inc. 2020-07-18 /pmc/articles/PMC7497916/ /pubmed/32681753 http://dx.doi.org/10.1002/acm2.12979 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
McCulloch, James
Pawlowski, Jamie
Kirby, Neil
Rasmussen, Karl
Shi, Zheng
Myers, Pamela
Stathakis, Sotirios
Papanikolaou, Niko
Saenz, Daniel L.
Patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D
title Patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D
title_full Patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D
title_fullStr Patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D
title_full_unstemmed Patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D
title_short Patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D
title_sort patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using ptw octavius 4d
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7497916/
https://www.ncbi.nlm.nih.gov/pubmed/32681753
http://dx.doi.org/10.1002/acm2.12979
work_keys_str_mv AT mccullochjames patientspecificdosequalityassuranceofsingleisocentermultiplebrainmetastasisstereotacticradiosurgeryusingptwoctavius4d
AT pawlowskijamie patientspecificdosequalityassuranceofsingleisocentermultiplebrainmetastasisstereotacticradiosurgeryusingptwoctavius4d
AT kirbyneil patientspecificdosequalityassuranceofsingleisocentermultiplebrainmetastasisstereotacticradiosurgeryusingptwoctavius4d
AT rasmussenkarl patientspecificdosequalityassuranceofsingleisocentermultiplebrainmetastasisstereotacticradiosurgeryusingptwoctavius4d
AT shizheng patientspecificdosequalityassuranceofsingleisocentermultiplebrainmetastasisstereotacticradiosurgeryusingptwoctavius4d
AT myerspamela patientspecificdosequalityassuranceofsingleisocentermultiplebrainmetastasisstereotacticradiosurgeryusingptwoctavius4d
AT stathakissotirios patientspecificdosequalityassuranceofsingleisocentermultiplebrainmetastasisstereotacticradiosurgeryusingptwoctavius4d
AT papanikolaouniko patientspecificdosequalityassuranceofsingleisocentermultiplebrainmetastasisstereotacticradiosurgeryusingptwoctavius4d
AT saenzdaniell patientspecificdosequalityassuranceofsingleisocentermultiplebrainmetastasisstereotacticradiosurgeryusingptwoctavius4d