Determination of machine‐specific tolerances using statistical process control analysis of long‐term uniform scanning proton machine QA results

PURPOSE: The purpose of this study was twofold: (a) report the long‐term monthly quality assurance (QA) dosimetry results of the uniform scanning beam delivery system, and (b) derive the machine‐specific tolerances based on the statistic process control (SPC) methodology and compare them against the...

Descripción completa

Detalles Bibliográficos
Autores principales: Rana, Suresh, Eckert, Colton, Singh, Hardev, Zheng, Yuanshui, Chacko, Michael, Storey, Mark, Chang, John
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7497929/
https://www.ncbi.nlm.nih.gov/pubmed/32741135
http://dx.doi.org/10.1002/acm2.12990
_version_ 1783583409284579328
author Rana, Suresh
Eckert, Colton
Singh, Hardev
Zheng, Yuanshui
Chacko, Michael
Storey, Mark
Chang, John
author_facet Rana, Suresh
Eckert, Colton
Singh, Hardev
Zheng, Yuanshui
Chacko, Michael
Storey, Mark
Chang, John
author_sort Rana, Suresh
collection PubMed
description PURPOSE: The purpose of this study was twofold: (a) report the long‐term monthly quality assurance (QA) dosimetry results of the uniform scanning beam delivery system, and (b) derive the machine‐specific tolerances based on the statistic process control (SPC) methodology and compare them against the AAPM TG224 recommended tolerances. METHODS: The Oklahoma Proton Center has four treatment rooms (TR1, TR2, TR3, and TR4) with a cyclotron and a universal nozzle. Monthly QA dosimetry results of four treatment rooms over a period of 6 yr (Feb 2014–Jan 2020) were retrieved from the QA database. The dosimetry parameters included dose output, range, flatness, and symmetry. The monthly QA results were analyzed using the SPC method, which included individuals and moving range (I‐MR) chart. The upper control limit (UCL) and lower control limit (LCL) were set at 3σ above and below the mean value, respectively. RESULTS: The mean difference in dose output was −0.3% (2σ = ±0.9% and 3σ = ±1.3%) in TR1, 0% (2σ = ±1.4% and 3σ = ±2.1%) in TR2, −0.2% (2σ = ±1.0% and 3σ = ±1.6%) in TR3, and −0.5% (2σ = ±0.9% and 3σ = ±1.3%) in TR4. The mean flatness and symmetry differences of all beams among the four treatment rooms were within ±1.0%. The 3σ for the flatness difference ranged from ±0.5% to ±1.2%. The 3σ for the symmetry difference ranged from ±0.4% to ±1.4%. The SPC analysis showed that the 3σ for range 10 cm (R10), R16, and R22 were within ±1 mm, whereas the 3σ for R28 exceeded ±1 mm in two rooms (3σ = ±1.9 mm in TR2 and 3σ = ±1.3 mm in TR3). CONCLUSION: The 3σ of the dose output, flatness, and symmetry differences in all four rooms were comparable to the TG224 tolerance (±2%). For the uniform scanning system, if the measured range is compared against the requested range, it may not always be possible to achieve the range difference within ±1 mm (TG224) for all the ranges.
format Online
Article
Text
id pubmed-7497929
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-74979292020-09-25 Determination of machine‐specific tolerances using statistical process control analysis of long‐term uniform scanning proton machine QA results Rana, Suresh Eckert, Colton Singh, Hardev Zheng, Yuanshui Chacko, Michael Storey, Mark Chang, John J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: The purpose of this study was twofold: (a) report the long‐term monthly quality assurance (QA) dosimetry results of the uniform scanning beam delivery system, and (b) derive the machine‐specific tolerances based on the statistic process control (SPC) methodology and compare them against the AAPM TG224 recommended tolerances. METHODS: The Oklahoma Proton Center has four treatment rooms (TR1, TR2, TR3, and TR4) with a cyclotron and a universal nozzle. Monthly QA dosimetry results of four treatment rooms over a period of 6 yr (Feb 2014–Jan 2020) were retrieved from the QA database. The dosimetry parameters included dose output, range, flatness, and symmetry. The monthly QA results were analyzed using the SPC method, which included individuals and moving range (I‐MR) chart. The upper control limit (UCL) and lower control limit (LCL) were set at 3σ above and below the mean value, respectively. RESULTS: The mean difference in dose output was −0.3% (2σ = ±0.9% and 3σ = ±1.3%) in TR1, 0% (2σ = ±1.4% and 3σ = ±2.1%) in TR2, −0.2% (2σ = ±1.0% and 3σ = ±1.6%) in TR3, and −0.5% (2σ = ±0.9% and 3σ = ±1.3%) in TR4. The mean flatness and symmetry differences of all beams among the four treatment rooms were within ±1.0%. The 3σ for the flatness difference ranged from ±0.5% to ±1.2%. The 3σ for the symmetry difference ranged from ±0.4% to ±1.4%. The SPC analysis showed that the 3σ for range 10 cm (R10), R16, and R22 were within ±1 mm, whereas the 3σ for R28 exceeded ±1 mm in two rooms (3σ = ±1.9 mm in TR2 and 3σ = ±1.3 mm in TR3). CONCLUSION: The 3σ of the dose output, flatness, and symmetry differences in all four rooms were comparable to the TG224 tolerance (±2%). For the uniform scanning system, if the measured range is compared against the requested range, it may not always be possible to achieve the range difference within ±1 mm (TG224) for all the ranges. John Wiley and Sons Inc. 2020-08-01 /pmc/articles/PMC7497929/ /pubmed/32741135 http://dx.doi.org/10.1002/acm2.12990 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Rana, Suresh
Eckert, Colton
Singh, Hardev
Zheng, Yuanshui
Chacko, Michael
Storey, Mark
Chang, John
Determination of machine‐specific tolerances using statistical process control analysis of long‐term uniform scanning proton machine QA results
title Determination of machine‐specific tolerances using statistical process control analysis of long‐term uniform scanning proton machine QA results
title_full Determination of machine‐specific tolerances using statistical process control analysis of long‐term uniform scanning proton machine QA results
title_fullStr Determination of machine‐specific tolerances using statistical process control analysis of long‐term uniform scanning proton machine QA results
title_full_unstemmed Determination of machine‐specific tolerances using statistical process control analysis of long‐term uniform scanning proton machine QA results
title_short Determination of machine‐specific tolerances using statistical process control analysis of long‐term uniform scanning proton machine QA results
title_sort determination of machine‐specific tolerances using statistical process control analysis of long‐term uniform scanning proton machine qa results
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7497929/
https://www.ncbi.nlm.nih.gov/pubmed/32741135
http://dx.doi.org/10.1002/acm2.12990
work_keys_str_mv AT ranasuresh determinationofmachinespecifictolerancesusingstatisticalprocesscontrolanalysisoflongtermuniformscanningprotonmachineqaresults
AT eckertcolton determinationofmachinespecifictolerancesusingstatisticalprocesscontrolanalysisoflongtermuniformscanningprotonmachineqaresults
AT singhhardev determinationofmachinespecifictolerancesusingstatisticalprocesscontrolanalysisoflongtermuniformscanningprotonmachineqaresults
AT zhengyuanshui determinationofmachinespecifictolerancesusingstatisticalprocesscontrolanalysisoflongtermuniformscanningprotonmachineqaresults
AT chackomichael determinationofmachinespecifictolerancesusingstatisticalprocesscontrolanalysisoflongtermuniformscanningprotonmachineqaresults
AT storeymark determinationofmachinespecifictolerancesusingstatisticalprocesscontrolanalysisoflongtermuniformscanningprotonmachineqaresults
AT changjohn determinationofmachinespecifictolerancesusingstatisticalprocesscontrolanalysisoflongtermuniformscanningprotonmachineqaresults