Multi-site benchmarking of clinical (13)C RF coils at 3T

A quality assurance protocol for RF coils is proposed, which can be used for volume (Tx/Rx) and surface (Rx) coils. Following this protocol, a benchmarking of seven coils (from three different MR sites) dedicated to (13)C MRI at 3T is reported. Coil performance is particularly important for 3T MRI a...

Descripción completa

Detalles Bibliográficos
Autores principales: Sánchez-Heredia, Juan Diego, Olin, Rie B., McLean, Mary A., Laustsen, Christoffer, Hansen, Adam E., Hanson, Lars G., Ardenkjær-Larsen, Jan Henrik
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7501735/
https://www.ncbi.nlm.nih.gov/pubmed/32755748
http://dx.doi.org/10.1016/j.jmr.2020.106798
Descripción
Sumario:A quality assurance protocol for RF coils is proposed, which can be used for volume (Tx/Rx) and surface (Rx) coils. Following this protocol, a benchmarking of seven coils (from three different MR sites) dedicated to (13)C MRI at 3T is reported. Coil performance is particularly important for 3T MRI at the (13)C frequency, since the coil-to-sample noise ratio is typically high. The coils are evaluated experimentally using the proposed protocol based on MR spectroscopic imaging performed with two different phantoms: one head-shaped, and one with cylindrical shape and nearly twice the volume of the first one. To achieve an unbiased SNR comparison of volume and array coils, coil combination was done using sensitivity profiles extracted from the data. SNR, noise correlation matrices and example g-factor maps are reported. For globally calibrated, equal excitation angles, the measured SNR shows large differences for the volume coils of up to 115% at the phantom center for a head phantom. The arrays show lower differences in superficial SNR. The sample surface depth at which the volume coils outperform the arrays is estimated to 7 cm, and SNR furthest away from the coil surface is 28% lower for the best array compared to the best volume coil. A broad set of coils for (13)C at 3T have been benchmarked. The results reported, and the method used to benchmark them, should guide the (13)C community to choose the most suitable coil for a given experiment.