Cargando…

Visual Acuity, Defocus Curve, Reading Speed and Patient Satisfaction with a Combined Extended Depth of Focus Intraocular Lens and Multifocal Intraocular Lens Modality

PURPOSE: To evaluate the vision, defocus curve, reading speed and patient satisfaction after implantation of an extended depth of focus (EDOF) IOL in one eye and a diffractive multifocal in the fellow eye. SETTING: One clinical practice in the USA. DESIGN: Prospective unmasked non-randomized clinica...

Descripción completa

Detalles Bibliográficos
Autores principales: Sandoval, Helga P, Potvin, Richard, Solomon, Kerry D
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7501984/
https://www.ncbi.nlm.nih.gov/pubmed/32982159
http://dx.doi.org/10.2147/OPTH.S276120
Descripción
Sumario:PURPOSE: To evaluate the vision, defocus curve, reading speed and patient satisfaction after implantation of an extended depth of focus (EDOF) IOL in one eye and a diffractive multifocal in the fellow eye. SETTING: One clinical practice in the USA. DESIGN: Prospective unmasked non-randomized clinical trial. METHODS: Subjects presenting for routine cataract surgery interested in reducing their dependence on spectacles were enrolled. Study endpoints included uncorrected and distance-corrected binocular distance (4 m), intermediate (66 cm) and near (40 cm) visual acuity at 3 months. Additional endpoints included the residual refraction, spectacle independence, overall satisfaction with vision, visual symptoms, reading speed and defocus curve. RESULTS: With a best distance correction, 77% (30/39) of subjects had 20/25 or better VA at distance, intermediate and near and nearly all subjects had 20/32 or better VA at all three distances. Defocus curve results showed mean continuous vision of 20/25 or better from plano to −2.50 D. Nearly 80% (31/39) of subjects had 20/25 visual acuity from 0.00 D to −2.50 D. The critical print size was between 0.3 and 0.4 logMAR (20/40 to 20/50 Snellen Equivalent). Spectacle independence was 100% at distance, 95% at intermediate and approximately 70% at near. The percentage of subjects who were “not at all” or “slightly” bothered by visual disturbances ranged from 64% (16/25) for Halos to 88% (22/25) for Starbursts. CONCLUSIONS: EDOF/bifocal IOL blended implantation results in at least 20/25 mean visual acuity from distance to near with good spectacle independence and low reports of severe visual disturbances.