Cargando…

Un-LINQed: Spontaneous extrusion of newer generation implantable loop recorders

BACKGROUND: Insertable cardiac monitors (ICMs) are often used for long-term monitoring of cardiac rhythm. The Medtronic’s LINQ Reveal ™ is a new generation wireless, automated, and patient responsive subcutaneous ECG monitoring device. Despite several advantages to its small size we have noted an un...

Descripción completa

Detalles Bibliográficos
Autores principales: Zakhar, Joseph, Blount, Thomas J., Gehi, Anil K., Ferns, Sunita J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7517543/
https://www.ncbi.nlm.nih.gov/pubmed/32298763
http://dx.doi.org/10.1016/j.ipej.2020.04.005
_version_ 1783587250162892800
author Zakhar, Joseph
Blount, Thomas J.
Gehi, Anil K.
Ferns, Sunita J.
author_facet Zakhar, Joseph
Blount, Thomas J.
Gehi, Anil K.
Ferns, Sunita J.
author_sort Zakhar, Joseph
collection PubMed
description BACKGROUND: Insertable cardiac monitors (ICMs) are often used for long-term monitoring of cardiac rhythm. The Medtronic’s LINQ Reveal ™ is a new generation wireless, automated, and patient responsive subcutaneous ECG monitoring device. Despite several advantages to its small size we have noted an unusually high incidence of extrusion at our center. METHODS: & Results: We conducted a retrospective case analysis to review Reveal LINQs implanted at our center. All devices were inserted using the provided insertion tools. Patients with extruded devices were identified and details regarding the site and technique of insertion, incision closure, use of peri-operative antibiotics, and follow-up details were collected. 81 patients underwent 85 Reveal LINQ implants at a tertiary care University Hospital referral center. The most common reason for implant was suspected arrhythmia with or without structural heart disease or unexplained syncope. There were 4 spontaneous extrusions occurring within 7–24 days after insertion with an incidence rate of 4.7%. One extruded device was anchored to subcutaneous tissue, and no pocket/device infections or hematomas were noted. CONCLUSIONS: Device migration and erosion through skin are important potential adverse events for the Reveal LINQ implantable loop recorder. This study reports an unexpectedly high rate of extrusion without infection. The authors suggest that the depth of the incision is the main factor impacting extrusions. Larger studies are recommended, however, and a proposed measure to avoid spontaneous extrusion is the design of a longer manufacturer’s blade in order to increase the depth of the incision and insertion.
format Online
Article
Text
id pubmed-7517543
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-75175432020-09-30 Un-LINQed: Spontaneous extrusion of newer generation implantable loop recorders Zakhar, Joseph Blount, Thomas J. Gehi, Anil K. Ferns, Sunita J. Indian Pacing Electrophysiol J Original Article BACKGROUND: Insertable cardiac monitors (ICMs) are often used for long-term monitoring of cardiac rhythm. The Medtronic’s LINQ Reveal ™ is a new generation wireless, automated, and patient responsive subcutaneous ECG monitoring device. Despite several advantages to its small size we have noted an unusually high incidence of extrusion at our center. METHODS: & Results: We conducted a retrospective case analysis to review Reveal LINQs implanted at our center. All devices were inserted using the provided insertion tools. Patients with extruded devices were identified and details regarding the site and technique of insertion, incision closure, use of peri-operative antibiotics, and follow-up details were collected. 81 patients underwent 85 Reveal LINQ implants at a tertiary care University Hospital referral center. The most common reason for implant was suspected arrhythmia with or without structural heart disease or unexplained syncope. There were 4 spontaneous extrusions occurring within 7–24 days after insertion with an incidence rate of 4.7%. One extruded device was anchored to subcutaneous tissue, and no pocket/device infections or hematomas were noted. CONCLUSIONS: Device migration and erosion through skin are important potential adverse events for the Reveal LINQ implantable loop recorder. This study reports an unexpectedly high rate of extrusion without infection. The authors suggest that the depth of the incision is the main factor impacting extrusions. Larger studies are recommended, however, and a proposed measure to avoid spontaneous extrusion is the design of a longer manufacturer’s blade in order to increase the depth of the incision and insertion. Elsevier 2020-04-13 /pmc/articles/PMC7517543/ /pubmed/32298763 http://dx.doi.org/10.1016/j.ipej.2020.04.005 Text en © 2020 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Article
Zakhar, Joseph
Blount, Thomas J.
Gehi, Anil K.
Ferns, Sunita J.
Un-LINQed: Spontaneous extrusion of newer generation implantable loop recorders
title Un-LINQed: Spontaneous extrusion of newer generation implantable loop recorders
title_full Un-LINQed: Spontaneous extrusion of newer generation implantable loop recorders
title_fullStr Un-LINQed: Spontaneous extrusion of newer generation implantable loop recorders
title_full_unstemmed Un-LINQed: Spontaneous extrusion of newer generation implantable loop recorders
title_short Un-LINQed: Spontaneous extrusion of newer generation implantable loop recorders
title_sort un-linqed: spontaneous extrusion of newer generation implantable loop recorders
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7517543/
https://www.ncbi.nlm.nih.gov/pubmed/32298763
http://dx.doi.org/10.1016/j.ipej.2020.04.005
work_keys_str_mv AT zakharjoseph unlinqedspontaneousextrusionofnewergenerationimplantablelooprecorders
AT blountthomasj unlinqedspontaneousextrusionofnewergenerationimplantablelooprecorders
AT gehianilk unlinqedspontaneousextrusionofnewergenerationimplantablelooprecorders
AT fernssunitaj unlinqedspontaneousextrusionofnewergenerationimplantablelooprecorders