Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap

PURPOSE: To evaluate repeatability, reproducibility, and agreement of intraocular pressure measurement with Tono-Pen using Ocufilm and polyethylene wrap tip cover in human eyes. METHODS: This is a cross-sectional, experimental study. A gas-sterilized, polyethylene wrap was used as an alternative for...

Descripción completa

Detalles Bibliográficos
Autores principales: Suvannachart, Pukkapol, Asawaphureekorn, Somkiat, Chansangpetch, Sunee, Inobhas, Abhibol, Pongpirul, Krit
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7521710/
https://www.ncbi.nlm.nih.gov/pubmed/32986771
http://dx.doi.org/10.1371/journal.pone.0239875
_version_ 1783588028469477376
author Suvannachart, Pukkapol
Asawaphureekorn, Somkiat
Chansangpetch, Sunee
Inobhas, Abhibol
Pongpirul, Krit
author_facet Suvannachart, Pukkapol
Asawaphureekorn, Somkiat
Chansangpetch, Sunee
Inobhas, Abhibol
Pongpirul, Krit
author_sort Suvannachart, Pukkapol
collection PubMed
description PURPOSE: To evaluate repeatability, reproducibility, and agreement of intraocular pressure measurement with Tono-Pen using Ocufilm and polyethylene wrap tip cover in human eyes. METHODS: This is a cross-sectional, experimental study. A gas-sterilized, polyethylene wrap was used as an alternative for Tono-Pen tip cover. For the right eye, 4 measurements using polyethylene wrap tip cover were done by two examiners (A and B) in random order to assess intra-observer repeatability and inter-observer reproducibility. For the left eye, 4 measurements were done by examiner A using both polyethylene wrap tip cover and Ocufilm in random order to assess intra-observer repeatability and agreement. Bland-Altman plot and intraclass correlation coefficient (ICC) were used in all analyses. Cost minimization analysis was evaluated. RESULTS: For examiner A, the repeatability of polyethylene wrap tip cover was -0.34, 95% limits of agreement (LOA) were -3.04 to 2.36, and ICC was 0.93 in the right eyes. As for the left eyes, the repeatability of polyethylene wrap tip cover was -0.33, 95% LOA were -3.01 to 2.36, and ICC was 0.93. For examiner B, the repeatability of polyethylene wrap tip cover was -0.02, 95% LOA were -2.88 to 2.83, and ICC was 0.92. The inter-observer reproducibility of polyethylene wrap tip cover was 0.36, 95% LOA were -3.34 to 4.07, and ICC was 0.90. The repeatability of Ocufilm was -0.42, 95% LOA were -2.75 to 1.91, and ICC was 0.95. The agreement of polyethylene wrap tip cover and Ocufilm was -0.71, 95% LOA were -5.18 to 3.76, and ICC was 0.83. There were no allergic reactions or serious complications. From the cost minimization analysis, the local cost for polyethylene tip cover was approximately 8 times lower compared to Ocufilm. CONCLUSIONS: Tono-pen with Ocufilm and polyethylene wrap tip cover were used to measure the intraocular pressure. The polyethylene wrap tip cover demonstrated acceptable repeatability, reproducibility, and agreement with Ocufilm in normotensive eyes, and had a good safety profile.
format Online
Article
Text
id pubmed-7521710
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-75217102020-10-06 Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap Suvannachart, Pukkapol Asawaphureekorn, Somkiat Chansangpetch, Sunee Inobhas, Abhibol Pongpirul, Krit PLoS One Research Article PURPOSE: To evaluate repeatability, reproducibility, and agreement of intraocular pressure measurement with Tono-Pen using Ocufilm and polyethylene wrap tip cover in human eyes. METHODS: This is a cross-sectional, experimental study. A gas-sterilized, polyethylene wrap was used as an alternative for Tono-Pen tip cover. For the right eye, 4 measurements using polyethylene wrap tip cover were done by two examiners (A and B) in random order to assess intra-observer repeatability and inter-observer reproducibility. For the left eye, 4 measurements were done by examiner A using both polyethylene wrap tip cover and Ocufilm in random order to assess intra-observer repeatability and agreement. Bland-Altman plot and intraclass correlation coefficient (ICC) were used in all analyses. Cost minimization analysis was evaluated. RESULTS: For examiner A, the repeatability of polyethylene wrap tip cover was -0.34, 95% limits of agreement (LOA) were -3.04 to 2.36, and ICC was 0.93 in the right eyes. As for the left eyes, the repeatability of polyethylene wrap tip cover was -0.33, 95% LOA were -3.01 to 2.36, and ICC was 0.93. For examiner B, the repeatability of polyethylene wrap tip cover was -0.02, 95% LOA were -2.88 to 2.83, and ICC was 0.92. The inter-observer reproducibility of polyethylene wrap tip cover was 0.36, 95% LOA were -3.34 to 4.07, and ICC was 0.90. The repeatability of Ocufilm was -0.42, 95% LOA were -2.75 to 1.91, and ICC was 0.95. The agreement of polyethylene wrap tip cover and Ocufilm was -0.71, 95% LOA were -5.18 to 3.76, and ICC was 0.83. There were no allergic reactions or serious complications. From the cost minimization analysis, the local cost for polyethylene tip cover was approximately 8 times lower compared to Ocufilm. CONCLUSIONS: Tono-pen with Ocufilm and polyethylene wrap tip cover were used to measure the intraocular pressure. The polyethylene wrap tip cover demonstrated acceptable repeatability, reproducibility, and agreement with Ocufilm in normotensive eyes, and had a good safety profile. Public Library of Science 2020-09-28 /pmc/articles/PMC7521710/ /pubmed/32986771 http://dx.doi.org/10.1371/journal.pone.0239875 Text en © 2020 Suvannachart et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Suvannachart, Pukkapol
Asawaphureekorn, Somkiat
Chansangpetch, Sunee
Inobhas, Abhibol
Pongpirul, Krit
Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap
title Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap
title_full Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap
title_fullStr Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap
title_full_unstemmed Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap
title_short Repeatability, reproducibility, agreement, and safety of Tono-Pen tip cover for intraocular measurement using latex and polyethylene wrap
title_sort repeatability, reproducibility, agreement, and safety of tono-pen tip cover for intraocular measurement using latex and polyethylene wrap
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7521710/
https://www.ncbi.nlm.nih.gov/pubmed/32986771
http://dx.doi.org/10.1371/journal.pone.0239875
work_keys_str_mv AT suvannachartpukkapol repeatabilityreproducibilityagreementandsafetyoftonopentipcoverforintraocularmeasurementusinglatexandpolyethylenewrap
AT asawaphureekornsomkiat repeatabilityreproducibilityagreementandsafetyoftonopentipcoverforintraocularmeasurementusinglatexandpolyethylenewrap
AT chansangpetchsunee repeatabilityreproducibilityagreementandsafetyoftonopentipcoverforintraocularmeasurementusinglatexandpolyethylenewrap
AT inobhasabhibol repeatabilityreproducibilityagreementandsafetyoftonopentipcoverforintraocularmeasurementusinglatexandpolyethylenewrap
AT pongpirulkrit repeatabilityreproducibilityagreementandsafetyoftonopentipcoverforintraocularmeasurementusinglatexandpolyethylenewrap