The Mechanical Impedance of the Human Skull via Direct Bone Conduction Implants
PURPOSE: The mechanical skull impedance is used in the design of direct bone drive hearing systems. This impedance is also important for the design of skull simulators used in manufacturing, service, and fitting procedures of such devices. PATIENTS AND METHODS: The skull impedance was measured in 45...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7522424/ https://www.ncbi.nlm.nih.gov/pubmed/33061679 http://dx.doi.org/10.2147/MDER.S260732 |
_version_ | 1783588178043600896 |
---|---|
author | Håkansson, Bo Woelflin, Fausto Tjellström, Anders Hodgetts, William |
author_facet | Håkansson, Bo Woelflin, Fausto Tjellström, Anders Hodgetts, William |
author_sort | Håkansson, Bo |
collection | PubMed |
description | PURPOSE: The mechanical skull impedance is used in the design of direct bone drive hearing systems. This impedance is also important for the design of skull simulators used in manufacturing, service, and fitting procedures of such devices. PATIENTS AND METHODS: The skull impedance was measured in 45 patients (25 female and 20 male) who were using percutaneous bone conduction implants (Ponto system or Baha system). Patients were recruited as a consecutive prospective case series and having an average age of 55.4 years (range 18–80 years). Seven patients were treated in Gothenburg, Sweden, and 38 patients in Edmonton, Canada. An impedance head (B&K 8001), driven by an excitation transducer with emphasized low-frequency response, was used to measure the mechanical point impedance with a swept sine from 100 to 10k Hz. RESULTS AND DISCUSSION: The skull impedance was found to have an anti-resonance of approximately 150 Hz, with a median maximum magnitude of 4500 mechanical ohms. Below this anti-resonance, the mechanical impedance was mainly mass-controlled corresponding to an effective skull mass of 2.5 kg at 100 Hz with substantial damping from neck and shoulder. Above the anti-resonance and up to 4 kHz, the impedance was stiffness-controlled, with a total compliance of approximately 450n m/N with a small amount of damping. At frequencies above 4 kHz, the skull impedance becomes gradually mass-controlled originating from the mass of the osseointegrated implant and adjacent bone. No significant differences related to gender or skull abnormalities were seen, just a slight dependence on age and major ear surgeries. The variability of the mechanical impedance among patients was not found to have any clinical importance. CONCLUSION: The mechanical skull impedance of percutaneous implants was found to confirm previous studies and can be used for optimizing the design and test procedures of direct bone drive hearing implants. |
format | Online Article Text |
id | pubmed-7522424 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Dove |
record_format | MEDLINE/PubMed |
spelling | pubmed-75224242020-10-14 The Mechanical Impedance of the Human Skull via Direct Bone Conduction Implants Håkansson, Bo Woelflin, Fausto Tjellström, Anders Hodgetts, William Med Devices (Auckl) Original Research PURPOSE: The mechanical skull impedance is used in the design of direct bone drive hearing systems. This impedance is also important for the design of skull simulators used in manufacturing, service, and fitting procedures of such devices. PATIENTS AND METHODS: The skull impedance was measured in 45 patients (25 female and 20 male) who were using percutaneous bone conduction implants (Ponto system or Baha system). Patients were recruited as a consecutive prospective case series and having an average age of 55.4 years (range 18–80 years). Seven patients were treated in Gothenburg, Sweden, and 38 patients in Edmonton, Canada. An impedance head (B&K 8001), driven by an excitation transducer with emphasized low-frequency response, was used to measure the mechanical point impedance with a swept sine from 100 to 10k Hz. RESULTS AND DISCUSSION: The skull impedance was found to have an anti-resonance of approximately 150 Hz, with a median maximum magnitude of 4500 mechanical ohms. Below this anti-resonance, the mechanical impedance was mainly mass-controlled corresponding to an effective skull mass of 2.5 kg at 100 Hz with substantial damping from neck and shoulder. Above the anti-resonance and up to 4 kHz, the impedance was stiffness-controlled, with a total compliance of approximately 450n m/N with a small amount of damping. At frequencies above 4 kHz, the skull impedance becomes gradually mass-controlled originating from the mass of the osseointegrated implant and adjacent bone. No significant differences related to gender or skull abnormalities were seen, just a slight dependence on age and major ear surgeries. The variability of the mechanical impedance among patients was not found to have any clinical importance. CONCLUSION: The mechanical skull impedance of percutaneous implants was found to confirm previous studies and can be used for optimizing the design and test procedures of direct bone drive hearing implants. Dove 2020-09-24 /pmc/articles/PMC7522424/ /pubmed/33061679 http://dx.doi.org/10.2147/MDER.S260732 Text en © 2020 Håkansson et al. http://creativecommons.org/licenses/by-nc/3.0/ This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php). |
spellingShingle | Original Research Håkansson, Bo Woelflin, Fausto Tjellström, Anders Hodgetts, William The Mechanical Impedance of the Human Skull via Direct Bone Conduction Implants |
title | The Mechanical Impedance of the Human Skull via Direct Bone Conduction Implants |
title_full | The Mechanical Impedance of the Human Skull via Direct Bone Conduction Implants |
title_fullStr | The Mechanical Impedance of the Human Skull via Direct Bone Conduction Implants |
title_full_unstemmed | The Mechanical Impedance of the Human Skull via Direct Bone Conduction Implants |
title_short | The Mechanical Impedance of the Human Skull via Direct Bone Conduction Implants |
title_sort | mechanical impedance of the human skull via direct bone conduction implants |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7522424/ https://www.ncbi.nlm.nih.gov/pubmed/33061679 http://dx.doi.org/10.2147/MDER.S260732 |
work_keys_str_mv | AT hakanssonbo themechanicalimpedanceofthehumanskullviadirectboneconductionimplants AT woelflinfausto themechanicalimpedanceofthehumanskullviadirectboneconductionimplants AT tjellstromanders themechanicalimpedanceofthehumanskullviadirectboneconductionimplants AT hodgettswilliam themechanicalimpedanceofthehumanskullviadirectboneconductionimplants AT hakanssonbo mechanicalimpedanceofthehumanskullviadirectboneconductionimplants AT woelflinfausto mechanicalimpedanceofthehumanskullviadirectboneconductionimplants AT tjellstromanders mechanicalimpedanceofthehumanskullviadirectboneconductionimplants AT hodgettswilliam mechanicalimpedanceofthehumanskullviadirectboneconductionimplants |