Dorsal Root Ganglion Stimulation Lead Fracture Within the Superficial Fascial Layers in 4 Cases
We present 4 cases of dorsal root ganglion stimulation lead fracture. In these cases, the surgical technique involved (1) traversing fascial layers for placement of leads via a Tuohy needle in the upper low back, (2) subcutaneous tunneling from the implantable pulse generator site to the lead punctu...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7523575/ https://www.ncbi.nlm.nih.gov/pubmed/32935950 http://dx.doi.org/10.1213/XAA.0000000000001307 |
_version_ | 1783588404691206144 |
---|---|
author | Chapman, Kenneth B. Patel, Kiran V. van Helmond, Noud Chien, George C. Chang |
author_facet | Chapman, Kenneth B. Patel, Kiran V. van Helmond, Noud Chien, George C. Chang |
author_sort | Chapman, Kenneth B. |
collection | PubMed |
description | We present 4 cases of dorsal root ganglion stimulation lead fracture. In these cases, the surgical technique involved (1) traversing fascial layers for placement of leads via a Tuohy needle in the upper low back, (2) subcutaneous tunneling from the implantable pulse generator site to the lead puncture site without dissecting below the superficial fascial plane at the puncture site, and (3) connection of the lead/extension with the generator. All fractures occurred adjacent to the original lead puncture site. These cases suggest lead entrapment within the membranous fascial plane, with tension on a thin lead, is a mechanism underlying lead fracture. |
format | Online Article Text |
id | pubmed-7523575 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Lippincott Williams & Wilkins |
record_format | MEDLINE/PubMed |
spelling | pubmed-75235752020-10-14 Dorsal Root Ganglion Stimulation Lead Fracture Within the Superficial Fascial Layers in 4 Cases Chapman, Kenneth B. Patel, Kiran V. van Helmond, Noud Chien, George C. Chang A A Pract Case Series We present 4 cases of dorsal root ganglion stimulation lead fracture. In these cases, the surgical technique involved (1) traversing fascial layers for placement of leads via a Tuohy needle in the upper low back, (2) subcutaneous tunneling from the implantable pulse generator site to the lead puncture site without dissecting below the superficial fascial plane at the puncture site, and (3) connection of the lead/extension with the generator. All fractures occurred adjacent to the original lead puncture site. These cases suggest lead entrapment within the membranous fascial plane, with tension on a thin lead, is a mechanism underlying lead fracture. Lippincott Williams & Wilkins 2020-09-11 /pmc/articles/PMC7523575/ /pubmed/32935950 http://dx.doi.org/10.1213/XAA.0000000000001307 Text en Copyright © 2020 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of the International Anesthesia Research Society. This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND) (http://creativecommons.org/licenses/by-nc-nd/4.0/) , where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal. |
spellingShingle | Case Series Chapman, Kenneth B. Patel, Kiran V. van Helmond, Noud Chien, George C. Chang Dorsal Root Ganglion Stimulation Lead Fracture Within the Superficial Fascial Layers in 4 Cases |
title | Dorsal Root Ganglion Stimulation Lead Fracture Within the Superficial Fascial Layers in 4 Cases |
title_full | Dorsal Root Ganglion Stimulation Lead Fracture Within the Superficial Fascial Layers in 4 Cases |
title_fullStr | Dorsal Root Ganglion Stimulation Lead Fracture Within the Superficial Fascial Layers in 4 Cases |
title_full_unstemmed | Dorsal Root Ganglion Stimulation Lead Fracture Within the Superficial Fascial Layers in 4 Cases |
title_short | Dorsal Root Ganglion Stimulation Lead Fracture Within the Superficial Fascial Layers in 4 Cases |
title_sort | dorsal root ganglion stimulation lead fracture within the superficial fascial layers in 4 cases |
topic | Case Series |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7523575/ https://www.ncbi.nlm.nih.gov/pubmed/32935950 http://dx.doi.org/10.1213/XAA.0000000000001307 |
work_keys_str_mv | AT chapmankennethb dorsalrootganglionstimulationleadfracturewithinthesuperficialfasciallayersin4cases AT patelkiranv dorsalrootganglionstimulationleadfracturewithinthesuperficialfasciallayersin4cases AT vanhelmondnoud dorsalrootganglionstimulationleadfracturewithinthesuperficialfasciallayersin4cases AT chiengeorgecchang dorsalrootganglionstimulationleadfracturewithinthesuperficialfasciallayersin4cases |