Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube?
INTRODUCTION: The aim was to evaluate the BD Barricor tubes by comparison with the BD Rapid Serum Tubes (RST) through measuring 25 analytes and monitoring sample stability after 24 hours and 7 days. MATERIALS AND METHODS: Samples of 52 patients from different hospital departments were examined. Bloo...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7528640/ https://www.ncbi.nlm.nih.gov/pubmed/33071555 http://dx.doi.org/10.11613/BM.2020.030704 |
_version_ | 1783589300566228992 |
---|---|
author | Mandić, Sanja Mandić, Dario Lukić, Iva Rolić, Tara Horvat, Vesna Lukić, Maja Osvald, Silvija Šerić, Vatroslav |
author_facet | Mandić, Sanja Mandić, Dario Lukić, Iva Rolić, Tara Horvat, Vesna Lukić, Maja Osvald, Silvija Šerić, Vatroslav |
author_sort | Mandić, Sanja |
collection | PubMed |
description | INTRODUCTION: The aim was to evaluate the BD Barricor tubes by comparison with the BD Rapid Serum Tubes (RST) through measuring 25 analytes and monitoring sample stability after 24 hours and 7 days. MATERIALS AND METHODS: Samples of 52 patients from different hospital departments were examined. Blood was collected in BD RST and BD Barricor tubes (Becton, Dickinson and Company, Franklin Lakes, USA). Analytes were measured by Beckman Coulter AU 480 (Beckman Coulter, Brea, USA), Dimension EXL (Siemens Healthcare Diagnostics, Newark, USA) and ARCHITECT i2000SR (Abbott Diagnostics, Lake Forest, USA). Between-tube comparison for each analyte was performed, along with testing analyte stability after storing samples at 4 °C. RESULTS: BD Barricor tubes showed unacceptable bias compared to BD RST tubes for potassium (K) (- 4.5%) and total protein (4.4%). Analyte stability after 24 hours was acceptable in both tested tubes for most of analytes, except for glucose, aspartate aminotransferase (AST) and lactate dehydrogenase (LD) in BD Barricor and free triiodothyronine in BD RST sample tubes. Analyte stability after 7 days was unacceptable for sodium, K, calcium, creatine kinase isoenzyme MB, AST, LD and troponin I in both samples; additionally for glucose, alkaline phosphatase and albumin in BD Barricor. CONCLUSION: All analytes, except K and total protein, can be measured interchangeably in BD RST and BD Barricor tubes, applying the same reference intervals. For most of the analytes, sample re-analysis can be performed in both tubes after 24 hours and 7 days, although BD RST tubes show better 7-day analytes stability over BD Barricor tubes. |
format | Online Article Text |
id | pubmed-7528640 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-75286402020-10-15 Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube? Mandić, Sanja Mandić, Dario Lukić, Iva Rolić, Tara Horvat, Vesna Lukić, Maja Osvald, Silvija Šerić, Vatroslav Biochem Med (Zagreb) Original Articles INTRODUCTION: The aim was to evaluate the BD Barricor tubes by comparison with the BD Rapid Serum Tubes (RST) through measuring 25 analytes and monitoring sample stability after 24 hours and 7 days. MATERIALS AND METHODS: Samples of 52 patients from different hospital departments were examined. Blood was collected in BD RST and BD Barricor tubes (Becton, Dickinson and Company, Franklin Lakes, USA). Analytes were measured by Beckman Coulter AU 480 (Beckman Coulter, Brea, USA), Dimension EXL (Siemens Healthcare Diagnostics, Newark, USA) and ARCHITECT i2000SR (Abbott Diagnostics, Lake Forest, USA). Between-tube comparison for each analyte was performed, along with testing analyte stability after storing samples at 4 °C. RESULTS: BD Barricor tubes showed unacceptable bias compared to BD RST tubes for potassium (K) (- 4.5%) and total protein (4.4%). Analyte stability after 24 hours was acceptable in both tested tubes for most of analytes, except for glucose, aspartate aminotransferase (AST) and lactate dehydrogenase (LD) in BD Barricor and free triiodothyronine in BD RST sample tubes. Analyte stability after 7 days was unacceptable for sodium, K, calcium, creatine kinase isoenzyme MB, AST, LD and troponin I in both samples; additionally for glucose, alkaline phosphatase and albumin in BD Barricor. CONCLUSION: All analytes, except K and total protein, can be measured interchangeably in BD RST and BD Barricor tubes, applying the same reference intervals. For most of the analytes, sample re-analysis can be performed in both tubes after 24 hours and 7 days, although BD RST tubes show better 7-day analytes stability over BD Barricor tubes. Croatian Society of Medical Biochemistry and Laboratory Medicine 2020-10-15 2020-10-15 /pmc/articles/PMC7528640/ /pubmed/33071555 http://dx.doi.org/10.11613/BM.2020.030704 Text en Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Articles Mandić, Sanja Mandić, Dario Lukić, Iva Rolić, Tara Horvat, Vesna Lukić, Maja Osvald, Silvija Šerić, Vatroslav Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube? |
title | Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube? |
title_full | Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube? |
title_fullStr | Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube? |
title_full_unstemmed | Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube? |
title_short | Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube? |
title_sort | test results comparison and sample stability study: is the bd barricor tube a suitable replacement for the bd rst tube? |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7528640/ https://www.ncbi.nlm.nih.gov/pubmed/33071555 http://dx.doi.org/10.11613/BM.2020.030704 |
work_keys_str_mv | AT mandicsanja testresultscomparisonandsamplestabilitystudyisthebdbarricortubeasuitablereplacementforthebdrsttube AT mandicdario testresultscomparisonandsamplestabilitystudyisthebdbarricortubeasuitablereplacementforthebdrsttube AT lukiciva testresultscomparisonandsamplestabilitystudyisthebdbarricortubeasuitablereplacementforthebdrsttube AT rolictara testresultscomparisonandsamplestabilitystudyisthebdbarricortubeasuitablereplacementforthebdrsttube AT horvatvesna testresultscomparisonandsamplestabilitystudyisthebdbarricortubeasuitablereplacementforthebdrsttube AT lukicmaja testresultscomparisonandsamplestabilitystudyisthebdbarricortubeasuitablereplacementforthebdrsttube AT osvaldsilvija testresultscomparisonandsamplestabilitystudyisthebdbarricortubeasuitablereplacementforthebdrsttube AT sericvatroslav testresultscomparisonandsamplestabilitystudyisthebdbarricortubeasuitablereplacementforthebdrsttube |