Introduction of a New Parameter for Evaluation of Digital Radiography System Performance

BACKGROUND: The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) phantom. METHODS: The image quality and radiation dose for seven different digital radiography systems were compared using the CDRAD...

Descripción completa

Detalles Bibliográficos
Autores principales: Choopani, Mohammad Reza, Chaparian, Ali
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7528983/
https://www.ncbi.nlm.nih.gov/pubmed/33062611
http://dx.doi.org/10.4103/jmss.JMSS_53_19
_version_ 1783589356088328192
author Choopani, Mohammad Reza
Chaparian, Ali
author_facet Choopani, Mohammad Reza
Chaparian, Ali
author_sort Choopani, Mohammad Reza
collection PubMed
description BACKGROUND: The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) phantom. METHODS: The image quality and radiation dose for seven different digital radiography systems were compared using the CDRAD phantom. Incident air kerma (IAK) values were measured for certain exposure settings in all digital radiography systems. The images from the CDRAD phantom were evaluated by three observers. The results were displayed in the form of a contrast-detail (CD) curve. In addition, the inverse image quality figure (IQFinv)-to-IAK ratios were used for quantitative comparison of different digital radiography system performance. RESULTS: Results of this study showed that the CD curves cannot be suitable criterion for determining the performance of digital radiography systems. For this reason, IQFinv-to-radiation dose (IAK) ratios in a fixed radiation condition were used. The highest performance in terms of producing high-quality images and low radiation dose was related to X-ray unit 1 and the lowest performance was for X-ray unit 5. CONCLUSION: The ratio of IQFinv to IAK for performance evaluation of digital radiography systems is an innovation of this study. A digital radiography system with a higher IQFinv-to-IAK ratio is associated with lower patient dose and better image quality. Therefore, it is recommended to equip the new imaging centers with the systems that have higher IQFinv-to-IAK ratios.
format Online
Article
Text
id pubmed-7528983
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-75289832020-10-13 Introduction of a New Parameter for Evaluation of Digital Radiography System Performance Choopani, Mohammad Reza Chaparian, Ali J Med Signals Sens Original Article BACKGROUND: The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) phantom. METHODS: The image quality and radiation dose for seven different digital radiography systems were compared using the CDRAD phantom. Incident air kerma (IAK) values were measured for certain exposure settings in all digital radiography systems. The images from the CDRAD phantom were evaluated by three observers. The results were displayed in the form of a contrast-detail (CD) curve. In addition, the inverse image quality figure (IQFinv)-to-IAK ratios were used for quantitative comparison of different digital radiography system performance. RESULTS: Results of this study showed that the CD curves cannot be suitable criterion for determining the performance of digital radiography systems. For this reason, IQFinv-to-radiation dose (IAK) ratios in a fixed radiation condition were used. The highest performance in terms of producing high-quality images and low radiation dose was related to X-ray unit 1 and the lowest performance was for X-ray unit 5. CONCLUSION: The ratio of IQFinv to IAK for performance evaluation of digital radiography systems is an innovation of this study. A digital radiography system with a higher IQFinv-to-IAK ratio is associated with lower patient dose and better image quality. Therefore, it is recommended to equip the new imaging centers with the systems that have higher IQFinv-to-IAK ratios. Wolters Kluwer - Medknow 2020-07-03 /pmc/articles/PMC7528983/ /pubmed/33062611 http://dx.doi.org/10.4103/jmss.JMSS_53_19 Text en Copyright: © 2020 Journal of Medical Signals & Sensors http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Choopani, Mohammad Reza
Chaparian, Ali
Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
title Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
title_full Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
title_fullStr Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
title_full_unstemmed Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
title_short Introduction of a New Parameter for Evaluation of Digital Radiography System Performance
title_sort introduction of a new parameter for evaluation of digital radiography system performance
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7528983/
https://www.ncbi.nlm.nih.gov/pubmed/33062611
http://dx.doi.org/10.4103/jmss.JMSS_53_19
work_keys_str_mv AT choopanimohammadreza introductionofanewparameterforevaluationofdigitalradiographysystemperformance
AT chaparianali introductionofanewparameterforevaluationofdigitalradiographysystemperformance