Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems
OBJECTIVES: The aim of this study was to determine the optimum combinations of target and filter materials for various X-ray tube voltage settings, as well as their effects on image quality and radiation dose. This was done using different digital mammography (DM) imaging systems with a breast equiv...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Scientific Scholar
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7533093/ https://www.ncbi.nlm.nih.gov/pubmed/33024611 http://dx.doi.org/10.25259/JCIS_30_2020 |
_version_ | 1783590063251128320 |
---|---|
author | Alkhalifah, Khaled Asbeutah, Akram Brindhaban, Ajit |
author_facet | Alkhalifah, Khaled Asbeutah, Akram Brindhaban, Ajit |
author_sort | Alkhalifah, Khaled |
collection | PubMed |
description | OBJECTIVES: The aim of this study was to determine the optimum combinations of target and filter materials for various X-ray tube voltage settings, as well as their effects on image quality and radiation dose. This was done using different digital mammography (DM) imaging systems with a breast equivalent phantom. MATERIAL AND METHODS: Two DM units with a tungsten (W) target, silver (Ag), and rhodium (Rh) filters and dual molybdenum (Mo) and Rh targets/filters were used. A tissue-equivalent mammography phantom of 6 cm thickness equivalent to a fibrofatty breast was exposed 20 times to different target/filter material combinations (W/Rh, W/Ag, Rh/Rh, Mo/Rh, and Mo/Mo) and various kV settings (28–34 kV). The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated for each exposure. RESULTS: The W/Ag combination resulted in the lowest entrance skin dose and mean glandular dose (MGD). The MGD for the W/Rh combination was 60% less than that of the W/Ag combination at 34 kV (P < 0.05). There was a direct relationship in the SNR with the Mo/Mo, Mo/Rh, and Rh/Rh combinations and an inverse relationship with the CNR in the 34 kV range. There were statistically significant differences between all five target/filter combinations, and the best SNR and CNR were observed for the W/Rh combination with a reduced radiation dose in the range of 28–30 kV (P < 0.05). CONCLUSION: For a breast thicknesses of 6 cm with a fibrofatty nature, the W/Rh combination delivers high performance in terms of image quality at a lower dose. |
format | Online Article Text |
id | pubmed-7533093 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Scientific Scholar |
record_format | MEDLINE/PubMed |
spelling | pubmed-75330932020-10-05 Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems Alkhalifah, Khaled Asbeutah, Akram Brindhaban, Ajit J Clin Imaging Sci Original Research OBJECTIVES: The aim of this study was to determine the optimum combinations of target and filter materials for various X-ray tube voltage settings, as well as their effects on image quality and radiation dose. This was done using different digital mammography (DM) imaging systems with a breast equivalent phantom. MATERIAL AND METHODS: Two DM units with a tungsten (W) target, silver (Ag), and rhodium (Rh) filters and dual molybdenum (Mo) and Rh targets/filters were used. A tissue-equivalent mammography phantom of 6 cm thickness equivalent to a fibrofatty breast was exposed 20 times to different target/filter material combinations (W/Rh, W/Ag, Rh/Rh, Mo/Rh, and Mo/Mo) and various kV settings (28–34 kV). The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated for each exposure. RESULTS: The W/Ag combination resulted in the lowest entrance skin dose and mean glandular dose (MGD). The MGD for the W/Rh combination was 60% less than that of the W/Ag combination at 34 kV (P < 0.05). There was a direct relationship in the SNR with the Mo/Mo, Mo/Rh, and Rh/Rh combinations and an inverse relationship with the CNR in the 34 kV range. There were statistically significant differences between all five target/filter combinations, and the best SNR and CNR were observed for the W/Rh combination with a reduced radiation dose in the range of 28–30 kV (P < 0.05). CONCLUSION: For a breast thicknesses of 6 cm with a fibrofatty nature, the W/Rh combination delivers high performance in terms of image quality at a lower dose. Scientific Scholar 2020-09-07 /pmc/articles/PMC7533093/ /pubmed/33024611 http://dx.doi.org/10.25259/JCIS_30_2020 Text en © 2020 Published by Scientific Scholar on behalf of Journal of Clinical Imaging Science https://creativecommons.org/licenses/by-nc-sa/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-Share Alike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Original Research Alkhalifah, Khaled Asbeutah, Akram Brindhaban, Ajit Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems |
title | Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems |
title_full | Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems |
title_fullStr | Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems |
title_full_unstemmed | Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems |
title_short | Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems |
title_sort | image quality and radiation dose for fibrofatty breast using target/filter combinations in two digital mammography systems |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7533093/ https://www.ncbi.nlm.nih.gov/pubmed/33024611 http://dx.doi.org/10.25259/JCIS_30_2020 |
work_keys_str_mv | AT alkhalifahkhaled imagequalityandradiationdoseforfibrofattybreastusingtargetfiltercombinationsintwodigitalmammographysystems AT asbeutahakram imagequalityandradiationdoseforfibrofattybreastusingtargetfiltercombinationsintwodigitalmammographysystems AT brindhabanajit imagequalityandradiationdoseforfibrofattybreastusingtargetfiltercombinationsintwodigitalmammographysystems |