Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems

OBJECTIVES: The aim of this study was to determine the optimum combinations of target and filter materials for various X-ray tube voltage settings, as well as their effects on image quality and radiation dose. This was done using different digital mammography (DM) imaging systems with a breast equiv...

Descripción completa

Detalles Bibliográficos
Autores principales: Alkhalifah, Khaled, Asbeutah, Akram, Brindhaban, Ajit
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Scientific Scholar 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7533093/
https://www.ncbi.nlm.nih.gov/pubmed/33024611
http://dx.doi.org/10.25259/JCIS_30_2020
_version_ 1783590063251128320
author Alkhalifah, Khaled
Asbeutah, Akram
Brindhaban, Ajit
author_facet Alkhalifah, Khaled
Asbeutah, Akram
Brindhaban, Ajit
author_sort Alkhalifah, Khaled
collection PubMed
description OBJECTIVES: The aim of this study was to determine the optimum combinations of target and filter materials for various X-ray tube voltage settings, as well as their effects on image quality and radiation dose. This was done using different digital mammography (DM) imaging systems with a breast equivalent phantom. MATERIAL AND METHODS: Two DM units with a tungsten (W) target, silver (Ag), and rhodium (Rh) filters and dual molybdenum (Mo) and Rh targets/filters were used. A tissue-equivalent mammography phantom of 6 cm thickness equivalent to a fibrofatty breast was exposed 20 times to different target/filter material combinations (W/Rh, W/Ag, Rh/Rh, Mo/Rh, and Mo/Mo) and various kV settings (28–34 kV). The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated for each exposure. RESULTS: The W/Ag combination resulted in the lowest entrance skin dose and mean glandular dose (MGD). The MGD for the W/Rh combination was 60% less than that of the W/Ag combination at 34 kV (P < 0.05). There was a direct relationship in the SNR with the Mo/Mo, Mo/Rh, and Rh/Rh combinations and an inverse relationship with the CNR in the 34 kV range. There were statistically significant differences between all five target/filter combinations, and the best SNR and CNR were observed for the W/Rh combination with a reduced radiation dose in the range of 28–30 kV (P < 0.05). CONCLUSION: For a breast thicknesses of 6 cm with a fibrofatty nature, the W/Rh combination delivers high performance in terms of image quality at a lower dose.
format Online
Article
Text
id pubmed-7533093
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Scientific Scholar
record_format MEDLINE/PubMed
spelling pubmed-75330932020-10-05 Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems Alkhalifah, Khaled Asbeutah, Akram Brindhaban, Ajit J Clin Imaging Sci Original Research OBJECTIVES: The aim of this study was to determine the optimum combinations of target and filter materials for various X-ray tube voltage settings, as well as their effects on image quality and radiation dose. This was done using different digital mammography (DM) imaging systems with a breast equivalent phantom. MATERIAL AND METHODS: Two DM units with a tungsten (W) target, silver (Ag), and rhodium (Rh) filters and dual molybdenum (Mo) and Rh targets/filters were used. A tissue-equivalent mammography phantom of 6 cm thickness equivalent to a fibrofatty breast was exposed 20 times to different target/filter material combinations (W/Rh, W/Ag, Rh/Rh, Mo/Rh, and Mo/Mo) and various kV settings (28–34 kV). The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated for each exposure. RESULTS: The W/Ag combination resulted in the lowest entrance skin dose and mean glandular dose (MGD). The MGD for the W/Rh combination was 60% less than that of the W/Ag combination at 34 kV (P < 0.05). There was a direct relationship in the SNR with the Mo/Mo, Mo/Rh, and Rh/Rh combinations and an inverse relationship with the CNR in the 34 kV range. There were statistically significant differences between all five target/filter combinations, and the best SNR and CNR were observed for the W/Rh combination with a reduced radiation dose in the range of 28–30 kV (P < 0.05). CONCLUSION: For a breast thicknesses of 6 cm with a fibrofatty nature, the W/Rh combination delivers high performance in terms of image quality at a lower dose. Scientific Scholar 2020-09-07 /pmc/articles/PMC7533093/ /pubmed/33024611 http://dx.doi.org/10.25259/JCIS_30_2020 Text en © 2020 Published by Scientific Scholar on behalf of Journal of Clinical Imaging Science https://creativecommons.org/licenses/by-nc-sa/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-Share Alike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Original Research
Alkhalifah, Khaled
Asbeutah, Akram
Brindhaban, Ajit
Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems
title Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems
title_full Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems
title_fullStr Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems
title_full_unstemmed Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems
title_short Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems
title_sort image quality and radiation dose for fibrofatty breast using target/filter combinations in two digital mammography systems
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7533093/
https://www.ncbi.nlm.nih.gov/pubmed/33024611
http://dx.doi.org/10.25259/JCIS_30_2020
work_keys_str_mv AT alkhalifahkhaled imagequalityandradiationdoseforfibrofattybreastusingtargetfiltercombinationsintwodigitalmammographysystems
AT asbeutahakram imagequalityandradiationdoseforfibrofattybreastusingtargetfiltercombinationsintwodigitalmammographysystems
AT brindhabanajit imagequalityandradiationdoseforfibrofattybreastusingtargetfiltercombinationsintwodigitalmammographysystems