Measurement Device for Ambient Carbonyl Sulfide by Means of Catalytic Reduction Followed by Wet Scrubbing/Fluorescence Detection
[Image: see text] A portable chemical analysis system for monitoring ambient carbonyl sulfide (COS) was investigated for the first time. COS is paid attention to from the perspectives of photosynthesis tracer, breath diagnosis marker, and new process-use in the manufacture of semiconductors. Recentl...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
American
Chemical Society
2020
|
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7557214/ https://www.ncbi.nlm.nih.gov/pubmed/33073096 http://dx.doi.org/10.1021/acsomega.0c02985 |
Sumario: | [Image: see text] A portable chemical analysis system for monitoring ambient carbonyl sulfide (COS) was investigated for the first time. COS is paid attention to from the perspectives of photosynthesis tracer, breath diagnosis marker, and new process-use in the manufacture of semiconductors. Recently, the threshold level value of COS was settled at 5 ppm in volume ratio (ppmv) for workplace safety management. In this work, COS was converted to H(2)S by a small column packed with alumina catalyzer at 65 °C. Then, the H(2)S produced was collected in a small channel scrubber to react with fluorescein mercuric acetate (FMA), and the resulting fluorescence quenching was monitored using an LED/photodiode-based miniature detector. The miniature channel scrubber was re-examined to determine its robustness and easy fabrication, and conditions of the catalyzer were optimized. When the FMA concentration used was 1 μM, the limit of detection and dynamic range, which were both proportional to the FMA concentration, were 0.07 and 25 ppbv, respectively. Ambient COS in the background level and even contaminated COS in the nitrogen gas cylinder could be detected. If necessary, H(2)S was removed selectively by reproducible adsorbent columns. COS concentrations of engine exhaust were measured by the proposed method and by cryo-trap-gas chromatography-flame photometric detection, and the results obtained (0.5–5.9 ppbv) by the two methods agreed well (R(2) = 0.945, n = 19). COS in ambient air and exhaust gases was successfully measured without any batchwise pretreatment. |
---|