Cargando…

Late Intraocular Lens Dislocation Resulting From Haptic Breakage Following Cardiopulmonary Resuscitation

We report a case of an asymptomatic 65-year-old male who on routine eye examination had anterior dislocation of an intraocular lens (IOL) implant placed 23 years prior. Ten months prior to presentation, the patient had cardiac surgery complicated by cardiac arrest requiring chest compressions. Dislo...

Descripción completa

Detalles Bibliográficos
Autores principales: Prabakaran, Samantha, Moureiden, Mouwafak, Qureshi, Saima, Sharma, Aman, Tran, Trisha, Shaikh, Saad
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7557322/
https://www.ncbi.nlm.nih.gov/pubmed/33072464
http://dx.doi.org/10.7759/cureus.10456
Descripción
Sumario:We report a case of an asymptomatic 65-year-old male who on routine eye examination had anterior dislocation of an intraocular lens (IOL) implant placed 23 years prior. Ten months prior to presentation, the patient had cardiac surgery complicated by cardiac arrest requiring chest compressions. Dislocation of an intraocular lens is a rare complication of cataract surgery. One of the causative factors for haptic breakage in our case was the polyimide haptic material. Polyimide has been shown to become brittle over time in warm and moist environments such as the human eye. This case demonstrates a case of late IOL dislocation chest compressions and, to the best of our knowledge, the first such case reported in the literature.