Cargando…
Multiple use applicator for vaginal tablets/vaginal inserts: compliance verification and suitability studies
BACKGROUND: A newly developed multiple use applicator for vaginal tablets/vaginal inserts* was evaluated for its general suitability. There are no standard procedures described in guidelines or general accepted publications as to how this kind of product should be tested for suitability of purpose....
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7558499/ https://www.ncbi.nlm.nih.gov/pubmed/33059630 http://dx.doi.org/10.1186/s12905-020-01099-y |
_version_ | 1783594652034662400 |
---|---|
author | Brunner, Heiko Theodor, Rudolf A. |
author_facet | Brunner, Heiko Theodor, Rudolf A. |
author_sort | Brunner, Heiko |
collection | PubMed |
description | BACKGROUND: A newly developed multiple use applicator for vaginal tablets/vaginal inserts* was evaluated for its general suitability. There are no standard procedures described in guidelines or general accepted publications as to how this kind of product should be tested for suitability of purpose. METHODS: Due to the lack of existing standard procedures, three separate tests were designed and successful executed: (a) First, a patient acceptability evaluation was carried out as part of a phase III trial (registered in EudraCT on 9 Jan 2017, number 2017-000142-22 2. https://www.clinicaltrialsregister.eu/ctr-search/search?query=eudract_number%3A2017-000142-22). (b) Secondly, a cleaning procedure for the applicator after simulated multiple use was developed and verified in order to prove a minimized risk of microbiological contamination of the device. A newly developed vaginal fluid to simulate multiple applications was applied for this trial. (c) Lastly, a third trial evaluated the mechanical stability and proper functionality of the applicator after multiple simulated uses. Even potential abrasion of material of the device was checked. RESULTS: Acceptable patient compliance of the new multiple use applicator was verified after 2 weeks of daily use. Furthermore, diary data assessments of patients participating in the Pharmacodynamic part of the trial were evaluated as well. Overall, patient acceptability of the new applicator was proven. The easy-to-use cleaning procedure for the applicator, which can even be carried out in a domestic environment, was developed and successfully verified for effectiveness, meeting all microbiological acceptance criteria for vaginal products of the European Pharmacopeia. The mechanical stability and proper functionality of the applicator after 50 simulated uses was also evaluated. All tested applicator batches (fresh and aged) passed the final evaluation, with no limitations in functionality. CONCLUSION: The tests developed and executed consider various compliance aspects of the newly developed applicator. Results of these individual tests met the expectations and/or the predefined acceptance criteria. All included trials performed produced results justifying and qualifying the applicator for the intended multiple-use. The procedures outlined may also be a guide as to how this kind of medical device can be tested for suitability. Trial registration Registered in EudraCT, number 2017-000142-22, start date 24 May 2017. |
format | Online Article Text |
id | pubmed-7558499 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-75584992020-10-15 Multiple use applicator for vaginal tablets/vaginal inserts: compliance verification and suitability studies Brunner, Heiko Theodor, Rudolf A. BMC Womens Health Technical Advance BACKGROUND: A newly developed multiple use applicator for vaginal tablets/vaginal inserts* was evaluated for its general suitability. There are no standard procedures described in guidelines or general accepted publications as to how this kind of product should be tested for suitability of purpose. METHODS: Due to the lack of existing standard procedures, three separate tests were designed and successful executed: (a) First, a patient acceptability evaluation was carried out as part of a phase III trial (registered in EudraCT on 9 Jan 2017, number 2017-000142-22 2. https://www.clinicaltrialsregister.eu/ctr-search/search?query=eudract_number%3A2017-000142-22). (b) Secondly, a cleaning procedure for the applicator after simulated multiple use was developed and verified in order to prove a minimized risk of microbiological contamination of the device. A newly developed vaginal fluid to simulate multiple applications was applied for this trial. (c) Lastly, a third trial evaluated the mechanical stability and proper functionality of the applicator after multiple simulated uses. Even potential abrasion of material of the device was checked. RESULTS: Acceptable patient compliance of the new multiple use applicator was verified after 2 weeks of daily use. Furthermore, diary data assessments of patients participating in the Pharmacodynamic part of the trial were evaluated as well. Overall, patient acceptability of the new applicator was proven. The easy-to-use cleaning procedure for the applicator, which can even be carried out in a domestic environment, was developed and successfully verified for effectiveness, meeting all microbiological acceptance criteria for vaginal products of the European Pharmacopeia. The mechanical stability and proper functionality of the applicator after 50 simulated uses was also evaluated. All tested applicator batches (fresh and aged) passed the final evaluation, with no limitations in functionality. CONCLUSION: The tests developed and executed consider various compliance aspects of the newly developed applicator. Results of these individual tests met the expectations and/or the predefined acceptance criteria. All included trials performed produced results justifying and qualifying the applicator for the intended multiple-use. The procedures outlined may also be a guide as to how this kind of medical device can be tested for suitability. Trial registration Registered in EudraCT, number 2017-000142-22, start date 24 May 2017. BioMed Central 2020-10-15 /pmc/articles/PMC7558499/ /pubmed/33059630 http://dx.doi.org/10.1186/s12905-020-01099-y Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Technical Advance Brunner, Heiko Theodor, Rudolf A. Multiple use applicator for vaginal tablets/vaginal inserts: compliance verification and suitability studies |
title | Multiple use applicator for vaginal tablets/vaginal inserts: compliance verification and suitability studies |
title_full | Multiple use applicator for vaginal tablets/vaginal inserts: compliance verification and suitability studies |
title_fullStr | Multiple use applicator for vaginal tablets/vaginal inserts: compliance verification and suitability studies |
title_full_unstemmed | Multiple use applicator for vaginal tablets/vaginal inserts: compliance verification and suitability studies |
title_short | Multiple use applicator for vaginal tablets/vaginal inserts: compliance verification and suitability studies |
title_sort | multiple use applicator for vaginal tablets/vaginal inserts: compliance verification and suitability studies |
topic | Technical Advance |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7558499/ https://www.ncbi.nlm.nih.gov/pubmed/33059630 http://dx.doi.org/10.1186/s12905-020-01099-y |
work_keys_str_mv | AT brunnerheiko multipleuseapplicatorforvaginaltabletsvaginalinsertscomplianceverificationandsuitabilitystudies AT theodorrudolfa multipleuseapplicatorforvaginaltabletsvaginalinsertscomplianceverificationandsuitabilitystudies |