Systematic evaluation of particle loss during handling in the percutaneous transluminal angioplasty for eight different drug-coated balloons

Paclitaxel drug coated balloons (DCBs) should provide optimal drug transfer exclusively to the target tissue. The aim of this study was to evaluate the particle loss by handling during angioplasty. A robotic arm was developed for systematic and reproducible drug abrasion experiments. The contact for...

Descripción completa

Detalles Bibliográficos
Autores principales: Heinrich, Andreas, Engler, Martin S., Güttler, Felix V., Matthäus, Christian, Popp, Jürgen, Teichgräber, Ulf K.-M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7560869/
https://www.ncbi.nlm.nih.gov/pubmed/33057070
http://dx.doi.org/10.1038/s41598-020-74227-1
_version_ 1783595168820101120
author Heinrich, Andreas
Engler, Martin S.
Güttler, Felix V.
Matthäus, Christian
Popp, Jürgen
Teichgräber, Ulf K.-M.
author_facet Heinrich, Andreas
Engler, Martin S.
Güttler, Felix V.
Matthäus, Christian
Popp, Jürgen
Teichgräber, Ulf K.-M.
author_sort Heinrich, Andreas
collection PubMed
description Paclitaxel drug coated balloons (DCBs) should provide optimal drug transfer exclusively to the target tissue. The aim of this study was to evaluate the particle loss by handling during angioplasty. A robotic arm was developed for systematic and reproducible drug abrasion experiments. The contact force on eight different commercially available DCB types was gradually increased, and high-resolution microscopic images of the deflated and inflated balloons were recorded. Three types of DCBs were classified: no abrasion of the drug in both statuses (deflated and inflated), significant abrasion only in the inflated status, and significant abrasion in both statuses. Quantitative measurements via image processing confirmed the qualitative classification and showed changes of the drug area between 2.25 and 45.73% (13.28 ± 14.29%) in the deflated status, and between 1.66 and 40.41% (21.43 ± 16.48%) in the inflated status. The structures and compositions of the DCBs are different, some are significantly more susceptible to drug loss. Particle loss by handling during angioplasty leads to different paclitaxel doses in the target regions for same DCB types. Susceptibility to involuntary drug loss may cause side effects, such as varying effective paclitaxel doses, which may explain variations in studies regarding the therapeutic outcome.
format Online
Article
Text
id pubmed-7560869
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Nature Publishing Group UK
record_format MEDLINE/PubMed
spelling pubmed-75608692020-10-19 Systematic evaluation of particle loss during handling in the percutaneous transluminal angioplasty for eight different drug-coated balloons Heinrich, Andreas Engler, Martin S. Güttler, Felix V. Matthäus, Christian Popp, Jürgen Teichgräber, Ulf K.-M. Sci Rep Article Paclitaxel drug coated balloons (DCBs) should provide optimal drug transfer exclusively to the target tissue. The aim of this study was to evaluate the particle loss by handling during angioplasty. A robotic arm was developed for systematic and reproducible drug abrasion experiments. The contact force on eight different commercially available DCB types was gradually increased, and high-resolution microscopic images of the deflated and inflated balloons were recorded. Three types of DCBs were classified: no abrasion of the drug in both statuses (deflated and inflated), significant abrasion only in the inflated status, and significant abrasion in both statuses. Quantitative measurements via image processing confirmed the qualitative classification and showed changes of the drug area between 2.25 and 45.73% (13.28 ± 14.29%) in the deflated status, and between 1.66 and 40.41% (21.43 ± 16.48%) in the inflated status. The structures and compositions of the DCBs are different, some are significantly more susceptible to drug loss. Particle loss by handling during angioplasty leads to different paclitaxel doses in the target regions for same DCB types. Susceptibility to involuntary drug loss may cause side effects, such as varying effective paclitaxel doses, which may explain variations in studies regarding the therapeutic outcome. Nature Publishing Group UK 2020-10-14 /pmc/articles/PMC7560869/ /pubmed/33057070 http://dx.doi.org/10.1038/s41598-020-74227-1 Text en © The Author(s) 2020 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Article
Heinrich, Andreas
Engler, Martin S.
Güttler, Felix V.
Matthäus, Christian
Popp, Jürgen
Teichgräber, Ulf K.-M.
Systematic evaluation of particle loss during handling in the percutaneous transluminal angioplasty for eight different drug-coated balloons
title Systematic evaluation of particle loss during handling in the percutaneous transluminal angioplasty for eight different drug-coated balloons
title_full Systematic evaluation of particle loss during handling in the percutaneous transluminal angioplasty for eight different drug-coated balloons
title_fullStr Systematic evaluation of particle loss during handling in the percutaneous transluminal angioplasty for eight different drug-coated balloons
title_full_unstemmed Systematic evaluation of particle loss during handling in the percutaneous transluminal angioplasty for eight different drug-coated balloons
title_short Systematic evaluation of particle loss during handling in the percutaneous transluminal angioplasty for eight different drug-coated balloons
title_sort systematic evaluation of particle loss during handling in the percutaneous transluminal angioplasty for eight different drug-coated balloons
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7560869/
https://www.ncbi.nlm.nih.gov/pubmed/33057070
http://dx.doi.org/10.1038/s41598-020-74227-1
work_keys_str_mv AT heinrichandreas systematicevaluationofparticlelossduringhandlinginthepercutaneoustransluminalangioplastyforeightdifferentdrugcoatedballoons
AT englermartins systematicevaluationofparticlelossduringhandlinginthepercutaneoustransluminalangioplastyforeightdifferentdrugcoatedballoons
AT guttlerfelixv systematicevaluationofparticlelossduringhandlinginthepercutaneoustransluminalangioplastyforeightdifferentdrugcoatedballoons
AT matthauschristian systematicevaluationofparticlelossduringhandlinginthepercutaneoustransluminalangioplastyforeightdifferentdrugcoatedballoons
AT poppjurgen systematicevaluationofparticlelossduringhandlinginthepercutaneoustransluminalangioplastyforeightdifferentdrugcoatedballoons
AT teichgraberulfkm systematicevaluationofparticlelossduringhandlinginthepercutaneoustransluminalangioplastyforeightdifferentdrugcoatedballoons