Cargando…

Comparative investigation of different telemetric methods for measuring intracranial pressure: a prospective pilot study

OBJECTIVES: Measurement of intracranial pressure (ICP) plays an important role in long-term monitoring and neuro-intensive treatment of patients with a cerebral shunt. Currently, only two complete telemetric implants with different technical features are available worldwide. This prospective pilot s...

Descripción completa

Detalles Bibliográficos
Autores principales: Rot, S., Dweek, M., Gutowski, P., Goelz, L., Meier, U., Lemcke, J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7568395/
https://www.ncbi.nlm.nih.gov/pubmed/33069242
http://dx.doi.org/10.1186/s12987-020-00225-0
_version_ 1783596510985846784
author Rot, S.
Dweek, M.
Gutowski, P.
Goelz, L.
Meier, U.
Lemcke, J.
author_facet Rot, S.
Dweek, M.
Gutowski, P.
Goelz, L.
Meier, U.
Lemcke, J.
author_sort Rot, S.
collection PubMed
description OBJECTIVES: Measurement of intracranial pressure (ICP) plays an important role in long-term monitoring and neuro-intensive treatment of patients with a cerebral shunt. Currently, only two complete telemetric implants with different technical features are available worldwide. This prospective pilot study aims to examine patients who had both probes implanted at overlapping times for clinical reasons and represents the first in vivo comparison of both measurement methods. MATERIALS AND METHODS: Patients with a primary subarachnoid hemorrhage or a spontaneous intracerebral hemorrhage with ventricular hemorrhage who had received a telemetric ICP probe (Raumedic(®) NEUROVENT(®)-P-tel) were included in the study. Conventional external ventricular drainages (EVD) and ventriculoperitoneal shunts with a telemetric ICP probe (Miethke Sensor Reservoir) were implanted in patients with hydrocephalus who required CSF (cerebrospinal fluid) drainage. Absolute ICP values from all systems were obtained. Due to the overlapping implantation time, parallel ICP measurements were performed via two devices simultaneously. ICP measurements via the sensor reservoir were repeated after 3 and 9 months. Differences between the absolute ICP values measured via the NEUROVENT(®)-P-tel probe, the Miethke sensor reservoir(®), and the EVD were analyzed. RESULTS: Seventeen patients were included in the present study between 2016 and 2018. 63% of all patients were male. In 11 patients the ICP measurements were followed up with both devices for 3 months. ICP measurements of the sensor reservoir showed corresponding trends in 9 cases compared to ICP measurement via the telemetry probe or EVD. Difference in absolute ICP values ranged between 14.5 mmHg and 0.0 mmHg. The average difference of the absolute ICP values in 8 cases was ≤ 3.5 mmHg. CONCLUSION: ICP measurements with both systems continuously showed synchronous absolute ICP values, however absolute values of ICP measurement with the different systems did not match.
format Online
Article
Text
id pubmed-7568395
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-75683952020-10-20 Comparative investigation of different telemetric methods for measuring intracranial pressure: a prospective pilot study Rot, S. Dweek, M. Gutowski, P. Goelz, L. Meier, U. Lemcke, J. Fluids Barriers CNS Research OBJECTIVES: Measurement of intracranial pressure (ICP) plays an important role in long-term monitoring and neuro-intensive treatment of patients with a cerebral shunt. Currently, only two complete telemetric implants with different technical features are available worldwide. This prospective pilot study aims to examine patients who had both probes implanted at overlapping times for clinical reasons and represents the first in vivo comparison of both measurement methods. MATERIALS AND METHODS: Patients with a primary subarachnoid hemorrhage or a spontaneous intracerebral hemorrhage with ventricular hemorrhage who had received a telemetric ICP probe (Raumedic(®) NEUROVENT(®)-P-tel) were included in the study. Conventional external ventricular drainages (EVD) and ventriculoperitoneal shunts with a telemetric ICP probe (Miethke Sensor Reservoir) were implanted in patients with hydrocephalus who required CSF (cerebrospinal fluid) drainage. Absolute ICP values from all systems were obtained. Due to the overlapping implantation time, parallel ICP measurements were performed via two devices simultaneously. ICP measurements via the sensor reservoir were repeated after 3 and 9 months. Differences between the absolute ICP values measured via the NEUROVENT(®)-P-tel probe, the Miethke sensor reservoir(®), and the EVD were analyzed. RESULTS: Seventeen patients were included in the present study between 2016 and 2018. 63% of all patients were male. In 11 patients the ICP measurements were followed up with both devices for 3 months. ICP measurements of the sensor reservoir showed corresponding trends in 9 cases compared to ICP measurement via the telemetry probe or EVD. Difference in absolute ICP values ranged between 14.5 mmHg and 0.0 mmHg. The average difference of the absolute ICP values in 8 cases was ≤ 3.5 mmHg. CONCLUSION: ICP measurements with both systems continuously showed synchronous absolute ICP values, however absolute values of ICP measurement with the different systems did not match. BioMed Central 2020-10-17 /pmc/articles/PMC7568395/ /pubmed/33069242 http://dx.doi.org/10.1186/s12987-020-00225-0 Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Rot, S.
Dweek, M.
Gutowski, P.
Goelz, L.
Meier, U.
Lemcke, J.
Comparative investigation of different telemetric methods for measuring intracranial pressure: a prospective pilot study
title Comparative investigation of different telemetric methods for measuring intracranial pressure: a prospective pilot study
title_full Comparative investigation of different telemetric methods for measuring intracranial pressure: a prospective pilot study
title_fullStr Comparative investigation of different telemetric methods for measuring intracranial pressure: a prospective pilot study
title_full_unstemmed Comparative investigation of different telemetric methods for measuring intracranial pressure: a prospective pilot study
title_short Comparative investigation of different telemetric methods for measuring intracranial pressure: a prospective pilot study
title_sort comparative investigation of different telemetric methods for measuring intracranial pressure: a prospective pilot study
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7568395/
https://www.ncbi.nlm.nih.gov/pubmed/33069242
http://dx.doi.org/10.1186/s12987-020-00225-0
work_keys_str_mv AT rots comparativeinvestigationofdifferenttelemetricmethodsformeasuringintracranialpressureaprospectivepilotstudy
AT dweekm comparativeinvestigationofdifferenttelemetricmethodsformeasuringintracranialpressureaprospectivepilotstudy
AT gutowskip comparativeinvestigationofdifferenttelemetricmethodsformeasuringintracranialpressureaprospectivepilotstudy
AT goelzl comparativeinvestigationofdifferenttelemetricmethodsformeasuringintracranialpressureaprospectivepilotstudy
AT meieru comparativeinvestigationofdifferenttelemetricmethodsformeasuringintracranialpressureaprospectivepilotstudy
AT lemckej comparativeinvestigationofdifferenttelemetricmethodsformeasuringintracranialpressureaprospectivepilotstudy