Repeatability and Reproducibility of Axial and Lateral Measurements on Handheld Optical Coherence Tomography Systems Compared with Tabletop System
PURPOSE: To compare the repeatability and reproducibility of axial and lateral retinal measurements using handheld optical coherence tomography (OCT) systems and a tabletop OCT system. METHODS: Graders measured central foveal thickness (CFT), optic nerve-to-fovea distance (OFD), and retinal nerve fi...
Autores principales: | , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Association for Research in Vision and Ophthalmology
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7585396/ https://www.ncbi.nlm.nih.gov/pubmed/33150050 http://dx.doi.org/10.1167/tvst.9.11.25 |
_version_ | 1783599781116903424 |
---|---|
author | Chen, Xi Tai, Vincent McGeehan, Brendan Ying, Gui-Shuang Viehland, Christian Imperio, Ryan Winter, Katrina P. Raynor, William Tran-Viet, Du Mangalesh, Shwetha Maguire, Maureen G. Toth, Cynthia A. |
author_facet | Chen, Xi Tai, Vincent McGeehan, Brendan Ying, Gui-Shuang Viehland, Christian Imperio, Ryan Winter, Katrina P. Raynor, William Tran-Viet, Du Mangalesh, Shwetha Maguire, Maureen G. Toth, Cynthia A. |
author_sort | Chen, Xi |
collection | PubMed |
description | PURPOSE: To compare the repeatability and reproducibility of axial and lateral retinal measurements using handheld optical coherence tomography (OCT) systems and a tabletop OCT system. METHODS: Graders measured central foveal thickness (CFT), optic nerve-to-fovea distance (OFD), and retinal nerve fiber layer (RNFL) thickness on OCT scans of the right eye of 10 healthy adults. Three OCT systems were used: handheld Leica Envisu, investigational handheld swept-source OCT (UC3), and Heidelberg Spectralis tabletop system. All eyes were imaged five times with each OCT system by each of two imagers. A components of variance analysis provided estimates of repeatability (variation due to random error) and reproducibility (variation due to imager, grader, and random error) expressed as standard deviation and (coefficient of variation %). RESULTS: Repeatability of CFT (µm) for Envisu, UC3, and Spectralis was 5.9 (2.6%), 6.9 (2.9%), and 4.7 (2.1%), and the reproducibility was 6.1 (2.7%), 7.3 (3.1%), and 4.7 (2.1%), respectively. The repeatability of OFD (mm) was 0.13 (2.9%), 0.10 (2.3%), and 0.07 (1.6%), and the reproducibility was 0.13 (3.0%), 0.10 (2.3%), and 0.07 (1.6%,) respectively. The repeatability for RNFL thickness (µm) for Envisu, UC3, and Spectralis was 4.3 (7.8%), 2.7 (5.4%), and 2.9 (4.9%), and the reproducibility was 4.5 (8.3%), 2.9 (5.8%), and 2.9 (4.9%), respectively. CONCLUSIONS: All three OCT systems had good repeatability and reproducibility with coefficients of variation of less than 3.5% for CFT and OFD measurements, and less than 8.5% for RNFL thickness. TRANSLATIONAL RELEVANCE: Our findings inform the repeatability and reproducibility of retinal axial and lateral measurements on handheld OCT and are useful for both clinical research and patient care. |
format | Online Article Text |
id | pubmed-7585396 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | The Association for Research in Vision and Ophthalmology |
record_format | MEDLINE/PubMed |
spelling | pubmed-75853962020-11-03 Repeatability and Reproducibility of Axial and Lateral Measurements on Handheld Optical Coherence Tomography Systems Compared with Tabletop System Chen, Xi Tai, Vincent McGeehan, Brendan Ying, Gui-Shuang Viehland, Christian Imperio, Ryan Winter, Katrina P. Raynor, William Tran-Viet, Du Mangalesh, Shwetha Maguire, Maureen G. Toth, Cynthia A. Transl Vis Sci Technol Article PURPOSE: To compare the repeatability and reproducibility of axial and lateral retinal measurements using handheld optical coherence tomography (OCT) systems and a tabletop OCT system. METHODS: Graders measured central foveal thickness (CFT), optic nerve-to-fovea distance (OFD), and retinal nerve fiber layer (RNFL) thickness on OCT scans of the right eye of 10 healthy adults. Three OCT systems were used: handheld Leica Envisu, investigational handheld swept-source OCT (UC3), and Heidelberg Spectralis tabletop system. All eyes were imaged five times with each OCT system by each of two imagers. A components of variance analysis provided estimates of repeatability (variation due to random error) and reproducibility (variation due to imager, grader, and random error) expressed as standard deviation and (coefficient of variation %). RESULTS: Repeatability of CFT (µm) for Envisu, UC3, and Spectralis was 5.9 (2.6%), 6.9 (2.9%), and 4.7 (2.1%), and the reproducibility was 6.1 (2.7%), 7.3 (3.1%), and 4.7 (2.1%), respectively. The repeatability of OFD (mm) was 0.13 (2.9%), 0.10 (2.3%), and 0.07 (1.6%), and the reproducibility was 0.13 (3.0%), 0.10 (2.3%), and 0.07 (1.6%,) respectively. The repeatability for RNFL thickness (µm) for Envisu, UC3, and Spectralis was 4.3 (7.8%), 2.7 (5.4%), and 2.9 (4.9%), and the reproducibility was 4.5 (8.3%), 2.9 (5.8%), and 2.9 (4.9%), respectively. CONCLUSIONS: All three OCT systems had good repeatability and reproducibility with coefficients of variation of less than 3.5% for CFT and OFD measurements, and less than 8.5% for RNFL thickness. TRANSLATIONAL RELEVANCE: Our findings inform the repeatability and reproducibility of retinal axial and lateral measurements on handheld OCT and are useful for both clinical research and patient care. The Association for Research in Vision and Ophthalmology 2020-10-21 /pmc/articles/PMC7585396/ /pubmed/33150050 http://dx.doi.org/10.1167/tvst.9.11.25 Text en Copyright 2020 The Authors http://creativecommons.org/licenses/by/4.0/ This work is licensed under a Creative Commons Attribution 4.0 International License. |
spellingShingle | Article Chen, Xi Tai, Vincent McGeehan, Brendan Ying, Gui-Shuang Viehland, Christian Imperio, Ryan Winter, Katrina P. Raynor, William Tran-Viet, Du Mangalesh, Shwetha Maguire, Maureen G. Toth, Cynthia A. Repeatability and Reproducibility of Axial and Lateral Measurements on Handheld Optical Coherence Tomography Systems Compared with Tabletop System |
title | Repeatability and Reproducibility of Axial and Lateral Measurements on Handheld Optical Coherence Tomography Systems Compared with Tabletop System |
title_full | Repeatability and Reproducibility of Axial and Lateral Measurements on Handheld Optical Coherence Tomography Systems Compared with Tabletop System |
title_fullStr | Repeatability and Reproducibility of Axial and Lateral Measurements on Handheld Optical Coherence Tomography Systems Compared with Tabletop System |
title_full_unstemmed | Repeatability and Reproducibility of Axial and Lateral Measurements on Handheld Optical Coherence Tomography Systems Compared with Tabletop System |
title_short | Repeatability and Reproducibility of Axial and Lateral Measurements on Handheld Optical Coherence Tomography Systems Compared with Tabletop System |
title_sort | repeatability and reproducibility of axial and lateral measurements on handheld optical coherence tomography systems compared with tabletop system |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7585396/ https://www.ncbi.nlm.nih.gov/pubmed/33150050 http://dx.doi.org/10.1167/tvst.9.11.25 |
work_keys_str_mv | AT chenxi repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT taivincent repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT mcgeehanbrendan repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT yingguishuang repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT viehlandchristian repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT imperioryan repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT winterkatrinap repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT raynorwilliam repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT tranvietdu repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT mangaleshshwetha repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT maguiremaureeng repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT tothcynthiaa repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem AT repeatabilityandreproducibilityofaxialandlateralmeasurementsonhandheldopticalcoherencetomographysystemscomparedwithtabletopsystem |