Cargando…

Performance evaluation of five ELISA kits for detecting anti-SARS-COV-2 IgG antibodies

OBJECTIVES: To evaluate and compare the performances of five commercial ELISA assays (EDI, AnshLabs, Dia.Pro, NovaTec, and Lionex) for detecting anti-SARS-CoV-2 IgG. METHODS: Seventy negative control samples (collected before the COVID-19 pandemic) and samples from 101 RT-PCR-confirmed SARS-CoV-2 pa...

Descripción completa

Detalles Bibliográficos
Autores principales: Yassine, Hadi M., Al-Jighefee, Hadeel, Al-Sadeq, Duaa W., Dargham, Soha R., Younes, Salma N., Shurrab, Farah, Marei, Reham M., Hssain, Ali Ait., Taleb, Sara, Alhussain, Hashim, Al-Nesf, Maryam A., Al-Khal, Abdullatif, Qotba, Hamda, Althani, Asmaa A., Tang, Patrick, Abu-Raddad, Laith J., Nasrallah, Gheyath K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Author(s). Published by Elsevier Ltd on behalf of International Society for Infectious Diseases. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7590641/
https://www.ncbi.nlm.nih.gov/pubmed/33127504
http://dx.doi.org/10.1016/j.ijid.2020.10.042
Descripción
Sumario:OBJECTIVES: To evaluate and compare the performances of five commercial ELISA assays (EDI, AnshLabs, Dia.Pro, NovaTec, and Lionex) for detecting anti-SARS-CoV-2 IgG. METHODS: Seventy negative control samples (collected before the COVID-19 pandemic) and samples from 101 RT-PCR-confirmed SARS-CoV-2 patients (collected at different time points from symptom onset: ≤7, 8–14 and >14 days) were used to compare the sensitivity, specificity, agreement, and positive and negative predictive values of each assay with RT-PCR. A concordance assessment between the five assays was also conducted. Cross-reactivity with other HCoV, non-HCoV respiratory viruses, non-respiratory viruses, and nuclear antigens was investigated. RESULTS: Lionex showed the highest specificity (98.6%; 95% CI 92.3–99.8), followed by EDI and Dia.Pro (97.1%; 95% CI 90.2–99.2), NovaTec (85.7%; 95% CI 75.7–92.1), then AnshLabs (75.7%; 95% CI 64.5–84.2). All ELISA kits cross-reacted with one anti-MERS IgG-positive sample, except Lionex. The sensitivity was low during the early stages of the disease but improved over time. After 14 days from symptom onset, Lionex and NovaTec showed the highest sensitivity at 87.9% (95% CI 72.7–95.2) and 86.4% (95% CI 78.5–91.7), respectively. The agreement with RT-PCR results based on Cohen’s kappa was as follows: Lionex (0.89) > NovaTec (0.70) > Dia.Pro (0.69) > AnshLabs (0.63) > EDI (0.55). CONCLUSION: The Lionex and NovaLisa IgG ELISA kits, demonstrated the best overall performance.