Performance Evaluation of VITEK MS for the Identification of a Wide Spectrum of Clinically Relevant Filamentous Fungi Using a Korean Collection
The correct identification of filamentous fungi is challenging. We evaluated the performance of the VITEK MS v3.0 system (bioMérieux, Marcy-l’Étoile, France) for the identification of a wide spectrum of clinically relevant filamentous fungi using a Korean collection. Strains that were added to the u...
Autores principales: | Shin, Ju Hyeon, Kim, Soo Hyun, Lee, Dain, Lee, Seung Yeob, Chun, Sejong, Lee, Jun Hyung, Won, Eun Jeong, Choi, Hyun Jung, Choi, Hyun Woo, Kee, Seung Jung, Shin, Myung Geun, Shin, Jong Hee |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Society for Laboratory Medicine
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7591280/ https://www.ncbi.nlm.nih.gov/pubmed/33063683 http://dx.doi.org/10.3343/alm.2021.41.2.214 |
Ejemplares similares
-
Evaluation of Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry-Based VITEK MS System for the Identification of Acinetobacter Species from Blood Cultures: Comparison with VITEK 2 and MicroScan Systems
por: Lee, Seung Yeob, et al.
Publicado: (2015) -
Comparison of the Bruker Biotyper and VITEK MS Matrix-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry Systems Using a Formic Acid Extraction Method to Identify Common and Uncommon Yeast Isolates
por: Lee, Hyun-Seung, et al.
Publicado: (2017) -
Performance of Microflex LT Biotyper and VITEK MS for Routine Identification of Yeasts
por: Byun, Jung-Hyun, et al.
Publicado: (2018) -
Evaluation of a Fully Automated Antinuclear Antibody Indirect Immunofluorescence Assay in Routine Use
por: Choi, Hyun-Woo, et al.
Publicado: (2020) -
The First Korean Case of NUP98-NSD1 and a Novel SNRK-ETV6 Fusion in a Pediatric Therapy-related Acute Myeloid Leukemia Patient Detected by Targeted RNA Sequencing
por: Lim, Ha Jin, et al.
Publicado: (2021)