Total skin electron beam therapy rationalization and utility of in vivo dosimetry in a high-volume centre

OBJECTIVE: This paper reports on the rationalization of a substantial pool of in vivo dosimetry (IVD) data from patients treated with total skin electron beam therapy (TSEBT) and the application of this to verify the accurate delivery of TSEBT when changing linac manufacturer. METHODS: Thermolumines...

Descripción completa

Detalles Bibliográficos
Autores principales: Misson-Yates, Sarah, Naeem, Marium, Palmer, Isabel, Holden, Eleanor, Hedley, Owen, McGovern, Mark, Morris, Stephen, Greener, Anthony G
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Institute of Radiology. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7592482/
https://www.ncbi.nlm.nih.gov/pubmed/33178940
http://dx.doi.org/10.1259/bjro.20190008
Descripción
Sumario:OBJECTIVE: This paper reports on the rationalization of a substantial pool of in vivo dosimetry (IVD) data from patients treated with total skin electron beam therapy (TSEBT) and the application of this to verify the accurate delivery of TSEBT when changing linac manufacturer. METHODS: Thermoluminescent dosimeter IVD data from 149 patients were analyzed comparing the population mean and standard deviation for each site. The number of sites required to confirm the prescribed dose were reviewed considering both dosimetric and clinical relevance. The reduced sites were then used to assess the continued dosimetric accuracy on new equipment and the results were compared statistically using the Mann–Witney test. RESULTS: The trunk dose measurement points were reduced from nine to six and five extra trunk sites were identified and reviewed clinically prior to removal. Following change in manufacturer the trunk dose points showed no statistically significant change and confirmed that patients had received within 1.3% of the intended mean trunk dose using both delivery methods. A statistically significant change in 4 out of the 13 extra trunk sites was seen following the move to the new centre. However, all but one site showed a change of less than 1 standard deviation. CONCLUSION: The total number of measurement points per patient were reduced from 27 to 19 which constituted a 25% saving in preparation and read out. Accurate delivery of prescribed dose was confirmed following measurement point reduction for treatments delivered on linacs from two different manufacturers. ADVANCES IN KNOWLEDGE: Proven methodology for rationalization of IVD measurements for TSEBT