Evaluation of the automated cartridge-based ARIES SARS-CoV-2 Assay (RUO) against automated Cepheid Xpert Xpress SARS-CoV-2 PCR as gold standard

INTRODUCTION: To evaluate the automated cartridge-based PCR approach ARIES SARS-CoV-2 Assay targeting the ORF-sequence and the N-gene of SARS-CoV-2. METHODS: In line with the suggestions by Rabenau and colleagues, the automated ARIES SARS-CoV-2 Assay was challenged with strongly positive samples, we...

Descripción completa

Detalles Bibliográficos
Autores principales: TANIDA, KONSTANTIN, KOSTE, LARS, KOENIG, CHRISTIAN, WENZEL, WERNER, FRITSCH, ANDREAS, FRICKMANN, HAGEN
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Akadémiai Kiadó 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7592518/
https://www.ncbi.nlm.nih.gov/pubmed/32804668
http://dx.doi.org/10.1556/1886.2020.00017
Descripción
Sumario:INTRODUCTION: To evaluate the automated cartridge-based PCR approach ARIES SARS-CoV-2 Assay targeting the ORF-sequence and the N-gene of SARS-CoV-2. METHODS: In line with the suggestions by Rabenau and colleagues, the automated ARIES SARS-CoV-2 Assay was challenged with strongly positive samples, weakly positive samples and negative samples. Further, intra-assay and inter-assay precision as well as the limit-of-detection (lod) were defined with quantified target RNA and DNA. The Cepheid Xpert Xpress SARS-Cov-2 Assay was used as gold standard. RESULTS: Concordance between the ARIES assay and the Cepheid assay was 100% for strongly positive samples and for negative samples, respectively. For weakly positive samples as confirmed applying the Cepheid assay, a relevant minority of 4 out of 15 samples (26.7%) went undetected by the ARIES assay. Intra- and inter-assay precision were satisfactory, while the lod was in the 10(3) DNA copies/reaction-range, in the 10(3) virus copies/reaction-range, or in the 10(3)-10(4) free RNA copies/reaction-range in our hands. CONCLUSIONS: The automated ARIES assay shows comparable test characteristics as the Cepheid assay focusing on strongly positive and negative samples but a slightly reduced sensitivity with weakly positive samples. Decisions on diagnostic use should include considerations on the lod.