Retrospective analysis of the impact of respiratory motion in treatment margins for frameless lung SBRT based on respiratory‐correlated CBCT data‐sets
PURPOSE: To investigate the impact of respiratory motion in the treatment margins for lung SBRT frameless treatments and to validate our treatment margins using 4D CBCT data analysis. METHODS: Two hundred and twenty nine fractions with early stage NSCLC were retrospectively analyzed. All patients we...
Autores principales: | Thengumpallil, Sheeba, Racine, Damien, Germond, Jean‐François, Péguret, Nicolas, Bourhis, Jean, Bochud, François, Moeckli, Raphaël |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7592980/ https://www.ncbi.nlm.nih.gov/pubmed/32996669 http://dx.doi.org/10.1002/acm2.13034 |
Ejemplares similares
-
Difference in performance between 3D and 4D CBCT for lung imaging: a dose and image quality analysis
por: Thengumpallil, Sheeba, et al.
Publicado: (2016) -
Implementation and validation of a beam‐current transformer on a medical pulsed electron beam LINAC for FLASH‐RT beam monitoring
por: Oesterle, Roxane, et al.
Publicado: (2021) -
Validation of Monte Carlo dose calculation algorithm for CyberKnife multileaf collimator
por: Gondré, Maude, et al.
Publicado: (2021) -
The minimal FLASH sparing effect needed to compensate the increase of radiobiological damage due to hypofractionation for late‐reacting tissues
por: Böhlen, Till Tobias, et al.
Publicado: (2022) -
Commissioning and validation of RayStation treatment planning system for CyberKnife M6
por: Gondré, Maude, et al.
Publicado: (2022)