Eliminating computed tomography imaging artifacts through 3D printed radiotherapy head supports
PURPOSE: The geometry of an immobilization device such as a headrest can cause undesired computed tomography (CT) artifacts that may affect both volume definition and dosimetry in radiotherapy of the brain. The purpose of this work was to reduce CT artifacts caused by a standard hard plastic hollow...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7592983/ https://www.ncbi.nlm.nih.gov/pubmed/32989881 http://dx.doi.org/10.1002/acm2.13027 |
_version_ | 1783601281738211328 |
---|---|
author | Bedwani, Stéphane Blais, Danis Roberge, David DeBlois, François |
author_facet | Bedwani, Stéphane Blais, Danis Roberge, David DeBlois, François |
author_sort | Bedwani, Stéphane |
collection | PubMed |
description | PURPOSE: The geometry of an immobilization device such as a headrest can cause undesired computed tomography (CT) artifacts that may affect both volume definition and dosimetry in radiotherapy of the brain. The purpose of this work was to reduce CT artifacts caused by a standard hard plastic hollow radiotherapy headrest. This was to be achieved through design and prototyping of a custom‐made head support. METHODS: A series of CT scans were acquired of both a water phantom and an anthropomorphic head phantom which were resting on custom‐made three‐dimensional (3D) printed supports. All custom‐made supports were made of polylactic acid (PLA) plastic filament and printed by fused deposition modeling (FDM) 3D printing technology. Initial designs were studied with a water phantom using a simplified support with straight and curved shapes both at the edges and as infill patterns. Imaging of a 3D printed clinical prototype was then compared to our standard headrest using an anthropomorphic head phantom. RESULTS: The presence of dark streaks inside both phantoms was seen on the CT images for headrests involving supports with straight shapes at the edges or as infill patterns. Such artifacts were ascribed to the exponential edge‐gradient effect (EEGE). No such artifact was observed when the support was designed with a combination of curved edges and infill patterns. CONCLUSION: When developing immobilization accessories for use in CT scanners, more attention could be paid to artifact attenuating design elements. This work illustrates the usefulness of 3D printing in prototyping radiotherapy accessories and solving concrete clinical problems. |
format | Online Article Text |
id | pubmed-7592983 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-75929832020-11-02 Eliminating computed tomography imaging artifacts through 3D printed radiotherapy head supports Bedwani, Stéphane Blais, Danis Roberge, David DeBlois, François J Appl Clin Med Phys Technical Notes PURPOSE: The geometry of an immobilization device such as a headrest can cause undesired computed tomography (CT) artifacts that may affect both volume definition and dosimetry in radiotherapy of the brain. The purpose of this work was to reduce CT artifacts caused by a standard hard plastic hollow radiotherapy headrest. This was to be achieved through design and prototyping of a custom‐made head support. METHODS: A series of CT scans were acquired of both a water phantom and an anthropomorphic head phantom which were resting on custom‐made three‐dimensional (3D) printed supports. All custom‐made supports were made of polylactic acid (PLA) plastic filament and printed by fused deposition modeling (FDM) 3D printing technology. Initial designs were studied with a water phantom using a simplified support with straight and curved shapes both at the edges and as infill patterns. Imaging of a 3D printed clinical prototype was then compared to our standard headrest using an anthropomorphic head phantom. RESULTS: The presence of dark streaks inside both phantoms was seen on the CT images for headrests involving supports with straight shapes at the edges or as infill patterns. Such artifacts were ascribed to the exponential edge‐gradient effect (EEGE). No such artifact was observed when the support was designed with a combination of curved edges and infill patterns. CONCLUSION: When developing immobilization accessories for use in CT scanners, more attention could be paid to artifact attenuating design elements. This work illustrates the usefulness of 3D printing in prototyping radiotherapy accessories and solving concrete clinical problems. John Wiley and Sons Inc. 2020-09-29 /pmc/articles/PMC7592983/ /pubmed/32989881 http://dx.doi.org/10.1002/acm2.13027 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Notes Bedwani, Stéphane Blais, Danis Roberge, David DeBlois, François Eliminating computed tomography imaging artifacts through 3D printed radiotherapy head supports |
title | Eliminating computed tomography imaging artifacts through 3D printed radiotherapy head supports |
title_full | Eliminating computed tomography imaging artifacts through 3D printed radiotherapy head supports |
title_fullStr | Eliminating computed tomography imaging artifacts through 3D printed radiotherapy head supports |
title_full_unstemmed | Eliminating computed tomography imaging artifacts through 3D printed radiotherapy head supports |
title_short | Eliminating computed tomography imaging artifacts through 3D printed radiotherapy head supports |
title_sort | eliminating computed tomography imaging artifacts through 3d printed radiotherapy head supports |
topic | Technical Notes |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7592983/ https://www.ncbi.nlm.nih.gov/pubmed/32989881 http://dx.doi.org/10.1002/acm2.13027 |
work_keys_str_mv | AT bedwanistephane eliminatingcomputedtomographyimagingartifactsthrough3dprintedradiotherapyheadsupports AT blaisdanis eliminatingcomputedtomographyimagingartifactsthrough3dprintedradiotherapyheadsupports AT robergedavid eliminatingcomputedtomographyimagingartifactsthrough3dprintedradiotherapyheadsupports AT debloisfrancois eliminatingcomputedtomographyimagingartifactsthrough3dprintedradiotherapyheadsupports |