Cargando…

Reproducibility of respiratory mechanics measurements in patients on invasive mechanical ventilation

OBJECTIVE: To evaluate the intra- and interexaminer reproducibility of measurements of the resistance and static and dynamic compliance of the respiratory system in patients on mechanical ventilation. METHODS: This was an analytical study conducted with individuals aged ≥ 18 years who were on invasi...

Descripción completa

Detalles Bibliográficos
Autores principales: Menezes Júnior, José da Natividade, Silva, Ludmilla Mota, Santos, Leonardo José Morais, Correia, Helena França, Lopes, Wende, Silva, Virgínia Eugênia Pinheiro e, dos Anjos, Jorge Luis Motta, Martinez, Bruno Prata
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Associação de Medicina Intensiva Brasileira - AMIB 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7595717/
https://www.ncbi.nlm.nih.gov/pubmed/33053029
http://dx.doi.org/10.5935/0103-507X.20200068
Descripción
Sumario:OBJECTIVE: To evaluate the intra- and interexaminer reproducibility of measurements of the resistance and static and dynamic compliance of the respiratory system in patients on mechanical ventilation. METHODS: This was an analytical study conducted with individuals aged ≥ 18 years who were on invasive mechanical ventilation and had no clinical diagnosis of respiratory system disease and/or chest abnormality. Three measurements of respiratory mechanics were performed with a 1-minute interval between them. The first and third measurements were performed by examiner A, the second by examiner B. The values for the resistance and static and dynamic compliance of the respiratory system were compared using the intraclass correlation coefficient. RESULTS: A total of 198 measurements of respiratory mechanics were performed for 66 patients on mechanical ventilation. The patients had a mean age of 52.6 ± 18.6 years and a mean body mass index of 21.6 ± 2.1kg/m(2); a surgical profile (61.5%) and female sex (53.8%) were predominant. Mean values were obtained for the three measurements of respiratory system resistance (A1: 15.7 ± 6.8cmH(2)O/L/s; B1: 15.7 ± 6.4cmH(2)O/L/s and A2: 15.9 ± 6.2cmH(2)O/L/s), respiratory system static compliance (A1: 42.1 ± 13.7mL/cmH(2)O; B1: 42.4 ± 14.6mL/cmH(2)O and A2: 42.2 ± 14.5mL/cmH(2)O) and respiratory system dynamic compliance (A1: 21.3 ± 7.3mL/cmH(2)O; B1: 21.4 ± 7.5mL/cmH(2)O and A2: 21.3 ± 6.2mL/cmH(2)O). The intraclass correlation coefficient was also calculated for respiratory system resistance (R = 0.882 and p = 0.001; R = 0.949 and p = 0.001 - interexaminer A1 versus B and B versus A2, respectively; R = 0.932 and p = 0.001 - intraexaminer); respiratory system static compliance (R = 0.951 and p = 0.001; R = 0.958 and p = 0.001 - interexaminer A1 versus B and B versus A2, respectively; R = 0.965 and p = 0.001 - intraexaminer) and respiratory system dynamic compliance (R = 0.957 and p = 0.001; R = 0.946 and p = 0.001 - interexaminer A1 versus B and B versus A2, respectively; R = 0.926 and p = 0.001 - intraexaminer). CONCLUSION: The measurements of resistance and static and dynamic compliance of the respiratory system show good intra- and interexaminer reproducibility for ventilated patients.