Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens

The i-STAT1 clinical analyzer has become an increasingly popular tool in clinical production animal medicine as it can provide pen-side results in a cost effective and timely manner when compared to standard benchtop serum biochemistry blood gas and chemistry analyses. This study compares the result...

Descripción completa

Detalles Bibliográficos
Autores principales: Sauer, Z.C., Taylor, K., Wolc, A., Viall, A., Fulton, J.E., Settar, P., Rubinoff, I., Schaal, T., Sato, Y.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7597810/
https://www.ncbi.nlm.nih.gov/pubmed/32616243
http://dx.doi.org/10.1016/j.psj.2020.03.059
_version_ 1783602446202830848
author Sauer, Z.C.
Taylor, K.
Wolc, A.
Viall, A.
Fulton, J.E.
Settar, P.
Rubinoff, I.
Schaal, T.
Sato, Y.
author_facet Sauer, Z.C.
Taylor, K.
Wolc, A.
Viall, A.
Fulton, J.E.
Settar, P.
Rubinoff, I.
Schaal, T.
Sato, Y.
author_sort Sauer, Z.C.
collection PubMed
description The i-STAT1 clinical analyzer has become an increasingly popular tool in clinical production animal medicine as it can provide pen-side results in a cost effective and timely manner when compared to standard benchtop serum biochemistry blood gas and chemistry analyses. This study compares the results of the portable Abbott i-STAT1 analyzer and the Abaxis VetScan VS2 for glucose (Glu, mg/dL), ionized Ca (mmol/L), Na (mmol/L), and K (mmol/L) values. Three genetically distinct commercial varieties (CV) of Hy-Line white-egg laying hens are used in this study (Hy-Line W-36, Hy-Line W-80, and Hy-Line W-80+). Thirty blood samples (n = 10 per CV) were obtained in the production house from the brachial vein and concurrently analyzed by the i-STAT1 portable device. Serum from 22 of these same samples was analyzed via VetScan VS2, a benchtop serum clinical biochemistry analyzer, using VetScan Avian/Reptilian Profile Plus reagent rotors. A paired T-test was used to test for statistical differences in means between the 2 instruments for each of the parameters. Parameters with significant mean differences were then subject to correlation and regression analysis to further evaluate relationships between the results from the 2 methods. Significant differences between means were found for Glu, Na, and K levels. Ca levels were found to be not directly comparable by the 2 analysis instruments. This comparison elucidates the importance of clinical analyzer validations when applying different strategies of diagnostic medicine in poultry.
format Online
Article
Text
id pubmed-7597810
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-75978102020-11-03 Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens Sauer, Z.C. Taylor, K. Wolc, A. Viall, A. Fulton, J.E. Settar, P. Rubinoff, I. Schaal, T. Sato, Y. Poult Sci Immunology, Health and Disease The i-STAT1 clinical analyzer has become an increasingly popular tool in clinical production animal medicine as it can provide pen-side results in a cost effective and timely manner when compared to standard benchtop serum biochemistry blood gas and chemistry analyses. This study compares the results of the portable Abbott i-STAT1 analyzer and the Abaxis VetScan VS2 for glucose (Glu, mg/dL), ionized Ca (mmol/L), Na (mmol/L), and K (mmol/L) values. Three genetically distinct commercial varieties (CV) of Hy-Line white-egg laying hens are used in this study (Hy-Line W-36, Hy-Line W-80, and Hy-Line W-80+). Thirty blood samples (n = 10 per CV) were obtained in the production house from the brachial vein and concurrently analyzed by the i-STAT1 portable device. Serum from 22 of these same samples was analyzed via VetScan VS2, a benchtop serum clinical biochemistry analyzer, using VetScan Avian/Reptilian Profile Plus reagent rotors. A paired T-test was used to test for statistical differences in means between the 2 instruments for each of the parameters. Parameters with significant mean differences were then subject to correlation and regression analysis to further evaluate relationships between the results from the 2 methods. Significant differences between means were found for Glu, Na, and K levels. Ca levels were found to be not directly comparable by the 2 analysis instruments. This comparison elucidates the importance of clinical analyzer validations when applying different strategies of diagnostic medicine in poultry. Elsevier 2020-05-15 /pmc/articles/PMC7597810/ /pubmed/32616243 http://dx.doi.org/10.1016/j.psj.2020.03.059 Text en © 2020 Published by Elsevier Inc. on behalf of Poultry Science Association Inc. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Immunology, Health and Disease
Sauer, Z.C.
Taylor, K.
Wolc, A.
Viall, A.
Fulton, J.E.
Settar, P.
Rubinoff, I.
Schaal, T.
Sato, Y.
Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens
title Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens
title_full Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens
title_fullStr Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens
title_full_unstemmed Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens
title_short Research Note: Comparison of chicken blood chemistry and electrolyte parameters between the portable i-STAT1 clinical analyzer and VetScan VS2 serum biochemistry panel using Hy-Line commercial white-egg laying hens
title_sort research note: comparison of chicken blood chemistry and electrolyte parameters between the portable i-stat1 clinical analyzer and vetscan vs2 serum biochemistry panel using hy-line commercial white-egg laying hens
topic Immunology, Health and Disease
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7597810/
https://www.ncbi.nlm.nih.gov/pubmed/32616243
http://dx.doi.org/10.1016/j.psj.2020.03.059
work_keys_str_mv AT sauerzc researchnotecomparisonofchickenbloodchemistryandelectrolyteparametersbetweentheportableistat1clinicalanalyzerandvetscanvs2serumbiochemistrypanelusinghylinecommercialwhiteegglayinghens
AT taylork researchnotecomparisonofchickenbloodchemistryandelectrolyteparametersbetweentheportableistat1clinicalanalyzerandvetscanvs2serumbiochemistrypanelusinghylinecommercialwhiteegglayinghens
AT wolca researchnotecomparisonofchickenbloodchemistryandelectrolyteparametersbetweentheportableistat1clinicalanalyzerandvetscanvs2serumbiochemistrypanelusinghylinecommercialwhiteegglayinghens
AT vialla researchnotecomparisonofchickenbloodchemistryandelectrolyteparametersbetweentheportableistat1clinicalanalyzerandvetscanvs2serumbiochemistrypanelusinghylinecommercialwhiteegglayinghens
AT fultonje researchnotecomparisonofchickenbloodchemistryandelectrolyteparametersbetweentheportableistat1clinicalanalyzerandvetscanvs2serumbiochemistrypanelusinghylinecommercialwhiteegglayinghens
AT settarp researchnotecomparisonofchickenbloodchemistryandelectrolyteparametersbetweentheportableistat1clinicalanalyzerandvetscanvs2serumbiochemistrypanelusinghylinecommercialwhiteegglayinghens
AT rubinoffi researchnotecomparisonofchickenbloodchemistryandelectrolyteparametersbetweentheportableistat1clinicalanalyzerandvetscanvs2serumbiochemistrypanelusinghylinecommercialwhiteegglayinghens
AT schaalt researchnotecomparisonofchickenbloodchemistryandelectrolyteparametersbetweentheportableistat1clinicalanalyzerandvetscanvs2serumbiochemistrypanelusinghylinecommercialwhiteegglayinghens
AT satoy researchnotecomparisonofchickenbloodchemistryandelectrolyteparametersbetweentheportableistat1clinicalanalyzerandvetscanvs2serumbiochemistrypanelusinghylinecommercialwhiteegglayinghens